S500 series GSM temperature humidity data logger is the new generation data logger , which has been added with the GSM alarm message function based on the original S500 series. Users can dial the data logger through any mobile phone to get the temperature and humidity values, besides, the temperature humidity data logger will send alarm message to defined users automatically once the setting limits being exceeded.
Operation manual for GSM temperature humidity data logger 1.3 Applications It has been widely applied in agricultural research industry, food industry, medical industry, electrical industry, environment protection and laboratory fields. 1.4 Installation and instructions for SIM card 1、 Suggest to buy GSM SIM card (SIM card without monthly rent), it must be GSM network SIM card, CDMA network is invalid.
Operation manual for GSM temperature humidity data logger MAX/MIN checking button Serial Number External Probe Port Reset button External Probe USB port SIM card slot DC Power Jack (9~12V) Wireless antenna 1.7 LCD display instructions Temperature and humidity values display. Symbol for backlight.
Operation manual for GSM temperature humidity data logger Symbol for temperature unit 1.8 Instructions for buttons Log function button, turn on/off the log function. Maximum and minimum button. Press to check the maximum and minimum value of all the recording data. Backlight on/off button.
Operation manual for GSM temperature humidity data logger Chapter 2. Software 2.1 The Requirement of the Computer Hardware OS: Windows XP/Vista/7/8 (32/64bit) CPU: 1.6GHz Physical Memory: 512MB Hard-drive Space: 4 GB USB Port: 1 2.2 Install Driver & Software Install Driver Notice: Windows 64bit please install 64bit-driver.
Page 8
Operation manual for GSM temperature humidity data logger Read Settings Sync Property Text Description Name English Letters or Numbers Name of the logger 10 Characters Must be the same as the one in the label of logger. Sampling Interval (s) Number from 1 to 240 Sampling frequency in LCD screen Logging Interval (s)
Page 9
Operation manual for GSM temperature humidity data logger 7. Choose 【GSM Setup】to setup cell phone number to receive alarming message. Input Cell# Click to confirm Click to Sync Settings...
Operation manual for GSM temperature humidity data logger 8. Setup the temperature and humidity upper and lower limits. Temp upper and lower limits Humid upper and lower limits 9. Then sync the setup with loggers after finish setup. 2.4 Data Analysis 1.
Page 11
Operation manual for GSM temperature humidity data logger Data List 3. Click on to save the data in the Excel format; or click on to save the data in a PDF file; or click on to save data in a picture. 4.
Operation manual for GSM temperature humidity data logger Chapter 3. Attention ⚫ S500 Series Data Logger are not waterproof. Do not connect loggers with computer when the logger is on water. ⚫ S500 Series Data Logger plastic shell is ABS, flame retardant, not resistant to acid and alkali.