MICRO TOUCH 1300
F13, User Manual , V3 09.08.2020
Biochemistry Analyser
MICRO TOUCH 1300
____________________________________________________________________________________________________________________________
Page 1 of 104

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MICRO TOUCH 1300 and is the answer not in the manual?

Questions and answers

Summary of Contents for GLOBAL DIAGNOSTICS MICRO TOUCH 1300

  • Page 1 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Biochemistry Analyser MICRO TOUCH 1300 ____________________________________________________________________________________________________________________________ Page 1 of 104...
  • Page 2: Table Of Contents

    SE OF THE NSTRUMENT ..........................3. INTRODUCTION........................11 3.1. D ESCRIPTION ..............................3.2. F EATURES ................................3.3. T MICRO TOUCH 1300 ……………………………………….13 ECHNICAL SPECIFICATION OF 4. PACKING, TRANSPORT AND STORAGE ................15 4.1. G ENERAL ARNINGS ............................ 4.2. P ACKING ................................
  • Page 3 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 7. PRECAUTIONS .......................... 27 8. MAIN MENU ..........................28 8.1. P ROGRAMMING DDING A EST ........................8.2. T ECHNICAL IAGNOSIS ..........................8.2.1. C LEAR EMORY ............................8.2.2. S ELECT ILTER ............................
  • Page 4 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 9.3. KINETIC ..........................69 9.4. END POINT ........................... 71 9.5. D IFFERENTIAL ..............................9.6. R ATIO .................................. 10. QC run in Biochemistry mode ....................79 11. Coagulation ..........................82 12. RFID ............................85 13.
  • Page 5: General Information

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 1. GENERAL INFORMATION 1.1. Warranty Information: Each Instrument is completely tested and guaranteed for twelve months from delivery. The warranty applies to all the mechanical and electrical parts. It is valid only for proper installation, use, and maintenance in compliance with the instructions given in this manual.
  • Page 6: Disposal Instruction

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 1.3. Disposal Instruction: In case of removal or disposal of instrument, following instructions need to be followed • Do not dispose in municipal waste; follow local regulations for instrument disposal. •...
  • Page 7: General Safety Warnings

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 2. GENERAL SAFETY WARNINGS 2.1. Danger - warnings symbols: The following symbols are used to inform the user of the safety rules. This symbol indicates generic danger. It means that, serious damage can occur to the operator if described precautions are not observed.
  • Page 8 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 General Symbols Symbol for “Manufacturer” Symbol for “ IN VITRO DIAGNOSTIC MEDICAL DEVICE” Page 8 of 104...
  • Page 9: Use Of The Instrument

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 2.2. Use of the Instrument: • The instrument has to be used for the designed purposes under specified conditions, following proper procedures and safety rules, by qualified personnel. THIS MANUAL CONTAINS INSTRUCTIONS FOR OPERATION BY QUALIFIED PERSONNEL •...
  • Page 10 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 • Warning: This equipment has been designed and tested to CISPER11 Class A. In a domestic environment it may cause radio interference, in which case, you may need to take measures to mitigate the interference."...
  • Page 11: Introduction

    3. INTRODUCTION 3.1. DESCRIPTION: MICRO TOUCH 1300 is a programmable Biochemistry Analyzer with a user-friendly touch screen. It measures the optical densities of samples and it uses algorithm to calculate results, which are used for biochemical investigation. It is a photometer operating in the visible range. The instrument is an open photometer suitable for absorbance (optical density) measures as well as sample concentration determination.
  • Page 12: Features

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 3.2. FEATURES: • Effective temperature regulation system with Peltier controlled cuvette / flow cell block. • Dual reading mode. • Robust system with built in stabilizer. • Latest technology with battery backup for 250 Biochem tests, 25 Coag & 25 RFID test with 30 results saved in each 3 levels of QC •...
  • Page 13: Technical Specification Of Micro Touch 1300

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 3.3. Technical specification of MICRO TOUCH 1300 : Human Machine Interface : TOUCHPANEL / KEYPAD Linear measurement range : 0.0000 to 3.0000 Absorbance Units (A) Photometric Accuracy : ± 2% or 0.007 whichever is higher, from 0 to 1.5A ±...
  • Page 14 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 : 5” Color TFT LCD module Display Printer : Built in Thermal Printer Memory : 512KB FRAM Storage Capacity 250 biochem, 25 coag & 25 RFID Open Tests with 30 QC...
  • Page 15: Packing, Transport And Storage

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 4. PACKING, TRANSPORT AND STORAGE 4.1. GENERAL WARNINGS: Instrument has to be decontaminated before packing for transportation. 4.2. PACKING: Packaging is needed whenever the instrument is to be transported or shipped by courier or other means.
  • Page 16: Instrument Description

    5. INSTRUMENT DESCRIPTION 5.1. Touch Sensitive screen: MICRO TOUCH 1300 provides a Touch sensitive TFT LCD panel and a KEYPAD for easy user interface. The Menus are displayed; the text of the parameter forms the TOUCH ZONE. Touch screen Layout...
  • Page 17: Perspective View

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 5.2. Perspective View Front View: Incubator Flow cell Printer Touch Screen Aspiration Keypad Switch Page 17 of 104...
  • Page 18 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Rear View: Cooling Fan ON / OFF Switch RS 232 Socket Waste Outlet Peristaltic Pump Module SMPS Page 18 of 104...
  • Page 19: Keypad

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 5.3. KEYPAD The FEED key is used to advance the printer paper by 1 line. PRINT key is used to take the print. WASH key is provided for aspiration of liquid into the Flowcell and for washing flow cell.
  • Page 20: Probe

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 5.4 Probe: The instrument is provided with a probe to aspirate the sample into the flow cell. The push button (Aspiration Switch) has to be pressed to activate the peristaltic pump.
  • Page 21 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Following is the diagram representing the fluid system of the instrument. The peristaltic pump is activated by the push button located on the probe. The sample is sucked into the flow cell for the photometric reading...
  • Page 22 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 PRINCIPLE OF WORKING: Below is the diagram representing the main functional elements of the instrument White light produced by the lamp is focused into a beam by lens and passes through the interference filter and monochromatic light beam falls on the sample.
  • Page 23: Printer

    09.08.2020 5.7 Printer A) Internal Printer (Thermal Printer) MICRO TOUCH 1300 comes with a built in 20 column Thermal Printer. User has to take proper care to handle this delicate instrument. TIPS FOR CAREFUL USAGE OF PRINTER Do not pull the paper when loaded •...
  • Page 24: Installation And Start-Up Instructions

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 6 INSTALLATION AND START-UP INSTRUCTIONS While installing and setting up the instrument, the safety warnings and general precautions described in section 7 must be observed. 6.1 Placing the Instrument: Place the instrument on a flat working surface or bench top capable of supporting the weight of the instrument.
  • Page 25: Start Up Instructions

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 6.4 Start up Instructions: Switch on the instrument. The instrument initializes all the parameters internally and carries out a power on self-test and then displays the following screen. MICRO TOUCH 1300 Please wait..
  • Page 26 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 • After the instrument completes the above steps, a TEST LIST/MENU SCREEN appears. • The instrument is now in IDLE mode, and ready for use. • Note: If the instrument is in Flow cell Mode and flow cell is missing, “Insert flow cell”...
  • Page 27: Precautions

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 7 PRECAUTIONS Keep the place dry and clean. Check all the grounding wires properly. Repeat the readings, if Absorbance is more than 2.0 A. Use original Packing for transportation. Use clean Cuvettes. Check the blank absorbance of the cuvette at regular intervals.
  • Page 28: Main Menu

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8 MAIN MENU 8.1 Programming/ADDING a NEW TEST When the unit is first turned on, Test List screen / Main Menu Screen appears on the display. GLUCOSE CREAT MENU ALBUMIN GAMMAGT...
  • Page 29 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 NAME MODE BLANK ABSORB FIL-P READ SAVE FIL-S FACT 1.0000 UNIT TEMP LIMIT No_Unit 1.0000 Now, select “MODE” touch zone. After selecting mode it will display following screen. This screen shows 6 different modes namely –...
  • Page 30 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 User can select Primary or Secondary filters. For Filter Selection, one needs to select “FIL-P” Or “FIL-S” touch zone present on the screen. It will display following screen where user can set Primary Or Secondary filters.
  • Page 31 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 User can select Aspiration or Sipping Volume. For volume Selection, one needs to select “VOL” touch zone presents on the screen. It will display following screen where user can set Aspiration volume.
  • Page 32 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Select “TEMP” touch zone presents on the screen. It will display following screen where user can select the required temperature needed to run the reagent. (NOTE: By Default the temperature is set to 37°C).
  • Page 33 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 After entering required parameter for Absorbance mode click on save, it will display following screen i.e. it will create Absorbance test in test list/ menu screen. GLUCOSE CREAT ABSORB MENU ALBUMIN...
  • Page 34: Technical Diagnosis

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.2 TECHNICAL DIAGNOSIS: TECHNICAL DIAGNOSIS CLEAR MEMORY LAMP AMPLITUDE Ref 8.2.4 Ref 8.2.1 SENSOR TEMPERATURE SELECT FILTER SETTING Ref 8.2.2 Ref 8.2.5 HOME FILTER WHEEL PRINTER SELF TEST Ref 8.2.3 Ref 8.2.6 8.2.1 Clear Memory:...
  • Page 35: Select Filter

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.2.2 Select Filter: When selected, brings the filter wheel TECHNICAL DIAGNOSIS TECHNICAL DIAGNOSIS to the selected filter position CLEAR MEMORY LAMP AMPLITUDE CLEAR MEMORY LAMP AMPLITUDE 8.2.3 Home Filter wheel: SELECT FILTER VAL:...
  • Page 36: Lamp Amplitude

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.2.4 Lamp Amplitude: To check lamp amplitude of the filters. TECHNICAL DIAGNOSIS MAIN MENU CLEAR MEMORY LAMP AMPLITUDE TECHNICAL QUALITY DIAGNOSIS CONTROL SELECT FILTER TEMPERATURE UTILITIES COMMUNICATION SETTING HOME FILTER PRINTER SELF...
  • Page 37: Temperature Setting

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.2.5 Temperature Setting: To verify the set temperature. 36.658 36.658 TECHNICAL DIAGNOSIS TEMPERATURE SETTING CLEAR MEMORY LAMP AMPLITUDE SET CUVETTE TEMP SELECT FILTER TEMPERATURE SET INCUB TEMP SETTING HOME FILTER PRINTER SELF...
  • Page 38: Printer Self Test

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.2.6 Printer Self Test: Routine test to check printer. TECHNICAL DIAGNOSIS MAIN MENU CLEAR MEMORY LAMP AMPLITUDE TECHNICAL QUALITY DIAGNOSIS CONTROL If printer is ON, and paper is properly inserted, the analyser prints following string.
  • Page 39: Utilities Screen

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3 UTILITIES SCREEN UTILITIES VER-GDM 3.2.7.013 DATE WASH LAMP PUMP CAL FLOW CELL PRINTER :OFF PROG FILTER CLINIC NAME DIST. NAME ABCD PASSWORD PRIORITISE TEST COMM : USB UPDATE FIRMWARE 8.3.1 To ENTER/CHANGE Date and Time UTILITIES VER-GDM 3.2.7.013...
  • Page 40: Password Utility

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.2 Password Utility: UTILITIES VER-GDM 3.2.7.013 DATE WASH LAMP MAIN MENU PUMP CAL FLOW CELL TECHNICAL QUALITY DIAGNOSIS CONTROL PRINTER: OFF PROG FILTER CLINIC NAME DIST. NAME UTILITIES COMMUNICATION ABCD PASSWORD...
  • Page 41 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 On selecting LOGIN, it will display following screen where one must enter accurate USER Name PASSWORD UTILITY PASSWORD UTILITY ENABLE/DISABLE ENABLE/DISABLE SET PASSWORD USER SET PASSWORD LOGIN LOGIN LOGOUT LOGOUT NON-RENTAL MODE NON-RENTAL MODE To Login/ enable user, enter the name.
  • Page 42 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 On selecting RENTAL MODE, it will save the card data and when minimum 10 days is remaining then it will pop up the message i.e. “Rental no. of days 10” PASSWORD UTILITY...
  • Page 43: Wash F.c

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.1. WASH: Wash is used to wash the flow cell. Selecting “Wash” will allow the aspiration of water or cleaning solution in the Flow cell. UTILITIES VER-GDM 3.2.7.013 MAIN MENU DATE...
  • Page 44 F13, User Manual , V3 09.08.2020 It is possible with MICRO TOUCH 1300 to use either the Flowcell or cuvettes to execute readings. The analyser is factory fitted with a Flowcell in the optical block. To operate the analyser with cuvettes select “UTILITIES”...
  • Page 45: Pump Cal

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.3. Pump Cal: The option of “Pump Cal” allows the operator to adjust the volume being aspirated through the Flowcell. The adjustment of the peristaltic pump has two functions: • To adjust the volume of liquid aspirated into the flow cell. The volume of liquid must be enough to rinse Flowcell sufficiently, but not more than the reaction mixture.
  • Page 46 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Step2: Select “Pump Cal” Step1: Go to Utilities UTILITIES VER-GDM 3.2.7.013 36.658 MAIN MENU DATE WASH LAMP TECHNICAL QUALITY PUMP CAL FLOW CELL DIAGNOSIS CONTROL PRINTER :ON PROG FILTER UTILITIES COMMUNICATION CLINIC NAME DIST.
  • Page 47 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Pump calibrator New Residual vol: 222 New Residual Cal Val: 1011 Pump calibration successful Page 47 of 104...
  • Page 48: Program Filter

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.6 Program Filter: The option of “PROG FILTER” allows the operator to enter the wavelength of the filters in 7 and 8 position. This option is to be used only in case of instruments where it is required to filters in 7...
  • Page 49: To Disable/ Enable Printer

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.7. To Disable/ Enable Printer: It is a toggle to disable or enable the printer. (Toggle: Any instruction that works first one way and then the other; it turns something on the first...
  • Page 50: T Oenter Clinic / Doctor Name

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.8 To Enter Clinic / Doctor Name: Step 1 : Step 2: Go to “Utilities” Touch “Enter Clinic Name” UTILITIES VER-GDM 3.2.7.013 MAIN MENU DATE WASH LAMP PUMP CAL FLOW CELL...
  • Page 51: Lamp: On

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.9 Lamp: ON It is a toggle to switch the lamp ON and OFF. (Toggle: any instruction that works first one way and then the other; it turns something on the first time it is used and then turns it off the next time) Step 1: Go to “Utilities”...
  • Page 52: Priorities Test

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.11. PRIORITISE Test: This particular option is used for setting the priority to the particular test. Step 1: Go to “Utilities” Step : Touch “Priorities Test” UTILITIES VER-GDM 3.2.7.013 MAIN MENU...
  • Page 53 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Step 6: Select “TEST” Step 7: Touh “UPDATE” PRIORITISE TEST GLUCOSE CREAT ALBUMIN GAMMAGT SWAP TEST: GLUCOSE UREA TRIG WITH HBA1C HDL CHOL NEW TEST : UPDATE CALCIUM SGOT GLUFAST RFID...
  • Page 54 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 This will interchange both the tests according to the priority given to them. After Updating the test the LIST screen will be displayed as shown below: TRIG CREAT MENU ALBUMIN GAMMAGT...
  • Page 55: Distributor Name

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.3.12. Distributor Name (DIST. NAME): This option is used for displaying distributor name up to 15 character. UTILITIES VER-GDM 3.2.7.013 MAIN MENU DATE WASH LAMP PUMP CAL FLOW CELL TECHNICAL QUALITY...
  • Page 56: Quality Control

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.4. QUALITY CONTROL Quality Control is a process that checks an instrument or testing site to make sure it is reporting accurate results on patients. The reproducibility of a result from a testing site or instrument should fall within a certain range.
  • Page 57 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 For example: Step 1: Go to “Quality Control” Step 2: Select “Saved test (e.g. TRIG)” 36.658 36.658 TRIG CREAT MAIN MENU ALBUMIN GAMMAGT TECHNICAL QUALITY DIAGNOSIS CONTROL UREA GLUCOSE HBA1C HDL CHOL...
  • Page 58 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Step 5: Select “Levy Jennings Step 6: Touch “PLOT” or STD Deviation” NAME NAME L1LOW 0.0000 L1LOW 0.0000 L1HIGH 2.0000 L1HIGH 2.0000 DATE RESULT DATE RESULT 17/07/18 1.2310 17/07/18 1.2310 PLOT...
  • Page 59: View Stored Results/ Print Patient Report

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.5. View Stored Results/ Print Patient Report Displays the latest 10 stored tests results. The system can store 2500 results in its database. The moment a sample is RUN, the result is stored in the system database.
  • Page 60: Communication

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 8.6. Communication: Both ports USB and SERIAL are available for communication. The instrument is equipped with an RS232 serial port for PC configuration (user- computer interface.). A cable is available to link the instrument to PC Communication will only start when both ends detects the presence of an active terminal or device.
  • Page 61 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Select “SEND RESULTS” to transfer the data (ie. Results) from instrument to PC. COMMUNICATION MAIN MENU SEND RESULTS TECHNICAL QUALITY DIAGNOSIS CONTROL SEND TEST RECOERDS UTILITIES COMMUNICATION RECV TEST RECORDS VIEW STORED...
  • Page 62 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Select “SEND TEST RECORDS” to transfer the data (ie. Saved Tests) from one instrument to another. COMMUNICATION COMMUNICATION SEND RESULTS SEND RESULTS SEND TEST RECOERDS SEND TEST RECOERDS RECV TEST RECORDS...
  • Page 63 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 And finally, select “RECEIVE TEST RECORDS” to receive the data (ie.Test Records) from other instrument. COMMUNICATION COMMUNICATION 36.658 SEND RESULTS SEND RESULTS SEND TEST RECOERDS SEND TEST RECOERDS RECV TEST RECORDS...
  • Page 64 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 COMMUNICATION COMMUNICATION SEND RESULTS SEND RESULTS SEND TEST RECOERDS SEND TEST RECOERDS RECV TEST RECORDS RECV TEST RECORDS ARE RXD AND TXD SHORTED ? CHECK SERIAL COMM CHECK SERIAL COMM TXD SHORTED? Short Pin no.
  • Page 65: Programming Modes

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 9 PROGRAMMING MODES A] User Defined Programming modes: - 9.1 ABSORBANCE The instrument measures the blank transmittance and the sample transmittance, subtracts them and calculates the sample absorbance (monochromatic mode).It is possible to repeat the measure with a different wavelength (Filter 2/ secondary filter) and consider the difference between the two measures.
  • Page 66 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Recalling /Running (Operating) a pre-programmed test. For example: 36.658 TRIG CREAT MENU ABSORBANCE Mode screen after recalling a saved test. ALBUMIN GAMMAGT To view the programmed test touch” VIEW” To run the programmed test, touch “RUN”...
  • Page 67: Fixed Time

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 9.2 FIXED TIME Change in absorbance of sample is taken at programmed time interval and concentration is calculated either from the factor fed by the user or using standard. PROGRAMMING / ADDING a NEW TEST...
  • Page 68 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Recalling /Running (Operating) a pre-programmed test. For example: TRIG CREAT MENU ALBUMIN GAMMAGT FIXED TIME Mode screen (After selecting a RUNQC UREA GLUCOSE saved test.) To view the programmed test touch” VIEW”...
  • Page 69: Kinetic

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 During Online graph; Point plotting per sec when (lag + read) < 100 sec Point plotting after 3 sec when (lag + read) < 298 sec Point plotting after 5 sec when (lag + read) <...
  • Page 70 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 KINETIC Mode screen before programming would look like the above screen Refer chapter13 for entering the test parameters. On completion, touch “SAVE” to save the programmed test. Recalling /Running (Operating) a pre-programmed test...
  • Page 71: Endpoint

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 9.4 Endpoint The instrument reads absorbance of the sample and calculates concentration using fed Factor or calculates the factor from concentration of the standard. PROGRAMMING / ADDING a NEW TEST NAME...
  • Page 72 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Recalling /Running (Operating) a pre-programmed test For example: ENDPOINT Mode screen (After Selecting a TRIG CREAT MENU saved test.) To view the programmed test touch” VIEW” ALBUMIN GAMMAGT To run the programmed test, touch “RUN” To print, touch “PRINT”...
  • Page 73 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 For duplicate or triplicate run Select “DUP” OR “TRI” after referencing. NAME: GLUCOSE MODE : EP NAME: GLUCOSE MODE : EP Factor: 1.0000 Factor: 1.0000 Sr.No:1 WASH WASH Aspirate Sample Read Sample...
  • Page 74 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 NAME: GLUCOSE MODE : EP Factor: 1.0000 Sr.No:1 0.164 Res: 47.10 mg/dl WASH AVG. RES: 47.264 mg/dl Read Sample Page 74 of 104...
  • Page 75: Differential

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 9.5 DIFFERENTIAL Differential of Sample and Sample Blank is taken. Concentration is calculated either from the factor fed by using standard. PROGRAMMING / ADDING a NEW TEST NAME MODE BLANK BIL-D...
  • Page 76 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Recalling /Running (Operating) a pre-programmed test For example: DIFFERENTIAL Mode screen (After TRIG CREAT MENU selecting a saved test.) To view the programmed test touch” ALBUMIN GAMMAGT VIEW” RUNQC UREA GLUCOSE To run the programmed test, touch “RUN”...
  • Page 77: Ratio

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 9.6 RATIO PROGRAMMING / ADDING a NEW TEST NAME MODE BLANK UREA FIL-P READ SAVE FIL-S FACT 1.0000 UNIT TEMP LIMIT No_Unit 1.0000 RATIO Mode screen before programming would look like the above screen Refer chapter13 for entering the test parameters.
  • Page 78 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Recalling /Running (Operating) a pre-programmed test For example: TRIG CREAT MENU RATIO Mode screen (After Selecting a saved test.) ALBUMIN GAMMAGT To view the programmed test touch” RUNQC UREA GLUCOSE VIEW”...
  • Page 79: Qc Run In Biochemistry Mode

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 10 QC Run in Biochemistry mode: NAME MODE BLANK GLUCOSE FIL-P READ SAVE FIL-S FACT 1.0000 UNIT TEMP LIMIT No_Unit 1.0000 EP Mode screen before programming would look like the above screen Refer chapter13 for entering the test parameters.
  • Page 80 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Recalling /Running (operating) a pre-programmed test TRIG CREAT MENU End point Mode screen (After ALBUMIN GAMMAGT Selecting a saved test.) RUNQC UREA GLUCOSE To view the programmed test touch” VIEW” HBA1C...
  • Page 81 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 NAME: GLUCOSE MODE: EP Sr.No: Factor: 1.0000 WASH Sample Readings Lag Time NAME: GLUCOSE MODE: EP Sr.No: Factor: 1.0000 0.153 WASH Res: Q1 44.495 Read Sample Page 81 of 104...
  • Page 82: Coagulation

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 11 COAGULATION This mode is used to report the Prothrombin Time (PT) and Activated Partial Thrombin Time(APTT). For PT, results is based on the ISI (International Sensitive Index) of the thromboplastin reagents and INR (International (ISI) Normalized Ratio).
  • Page 83 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 NAME: PT MODE: COAG MENU APTT RUNQC WASH PRINT Close the Lid &Press Run NXT PRV NEW VIEW DEL BIO COA RFID NAME: PT MODE: COAG NAME: PT MODE: COAG ISI Value: 1.0000 ISI Value: 1.0000...
  • Page 84 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 NAME: APTT MODE: COAG MENU APTT RUNQC WASH PRINT Close the Lid &Press Run NXT PRV NEW VIEW DEL BIO COA RFID NAME: APTT MODE: COAG NAME: APTT MODE: COAG Sr.No.: 1 Sample APTT : 9.1130 sec...
  • Page 85: Rfid

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 RFID User can read the data of card by clicking new option present in RFID screen. and saved data/test can be view in RFID screen. 1. Click on NEW 2. Click on YES...
  • Page 86 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 NOTE:- If test count is over then it will display the message “Max test count limit over” If days is over or key is expired, then it will display the message “Max days count limit over”...
  • Page 87: General Functions

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13. GENERAL FUNCTIONS 13.1. Enter Test Name: Touch “Name” under the selected mode on the Test Screen. For example: In “FIXTIME” mode NAME MODE BLANK NAME MODE BLANK FIL-P READ SAVE...
  • Page 88: Temperature Selection

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.2. Temperature Selection: For example: In “Fix time” mode. NAME MODE BLANK NAME MODE BLANK FIXTM FIXTM FIL-P READ SAVE FIL-P READ SAVE FIL-S FACT FIL-S FACT 1.0000 1.0000 UNIT TEMP...
  • Page 89: Filter Selection

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.3. Filter Selection: To select the interferential filter required for the test being programmed. For example: In Fix time mode. NAME MODE BLANK FIXTM FIL-P READ SAVE FIL-S FACT 1.0000 UNIT...
  • Page 90: Selection Of Units

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.4. Selection of Units: For example: In “Fix time” mode. NAME MODE BLANK NAME MODE BLANK FIXTM FIXTM Select the measuring unit for the test Touch “UNIT”. being programmed. On proper selection the analyser...
  • Page 91: Blank Selection

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.5. Blank Selection For example: In “Fix time” mode. NAME MODE BLANK NAME MODE BLANK FIXTM FIXTM FIL-P READ SAVE FIL-P READ SAVE FIL-S FACT FIL-S FACT 1.0000 1.0000 UNIT TEMP...
  • Page 92: Selection Of Aspiration/Sipping Volume

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.6. Selection of Aspiration/Sipping volume For example: In “Fix time” mode. NAME MODE BLANK NAME MODE BLANK FIXTM FIXTM FIL-P READ SAVE FIL-P READ SAVE FIL-S FACT FIL-S FACT 1.0000 1.0000...
  • Page 93: Normal Value Selection

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.7. Normal Value Selection: Normal Values: The normal range for the concentration or the activity of the assay can be entered. If a test result is out of this range, the result will be flagged with L or H.
  • Page 94: Lag Time

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.8. Lag Time: Delay interval (in seconds) before readings are executed. For example: In Fix time mode. NAME MODE BLANK FIXTM Touch” LAG”. FIL-P READ SAVE On proper selection the analyser responds the lag time selection screen.
  • Page 95: Read Time

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.9. Read Time Time interval (in seconds) between subsequent readings.(For kinetic and fixed time mode) For example: In FIXED TIME mode. NAME MODE BLANK FIXTM Touch “READ”. FIL-P READ SAVE On proper selection the analyser...
  • Page 96: Limit Set (End Point , Differential & Ratio )

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.10. Limit Set Init O.D: High or low absorbance level. This limit is entered for those samples which may exceed the limits of the reagent system. Max Delta/min :Upper limit for the Delta/min Max Delta: Upper limit for the Delta Rgnt.
  • Page 97: Factor / K-Factor / Standard Selection

    F13, User Manual , V3 09.08.2020 13.11. Factor / K-Factor / Standard selection Selection of Standard or Factor MICRO TOUCH 1300 has the flexibility to calculate the results by- 1) Factor Method 2) Single Standard/ Multi-standard For example in Fix time mode,...
  • Page 98 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 Single Standard/ Multi-standard On selecting this particular touch zone it will display following screen where user can enter number of Standards by placing the cursor in the textbox. It will display a keypad where user can enter min. 1 STD &...
  • Page 99 MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 STD.NUM= PLOT PRINT PLOT STD.NUM= PLOT PRINT PLOT CON 1 ABS 1 GRAPH CON 1 ABS 1 GRAPH 10.000 0.0000 PT TO PT 10.000 0.0000 CON 2 ABS 2 CON 2...
  • Page 100: Qc Values

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 13.12. QC values: This parameter consists of three levels namely- QC Level 1 QC Level 2 QC Level 3 Following screen appears on selecting this test parameter. NAME MODE BLANK NAME...
  • Page 101: Trouble Shooting

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 14. TROUBLE SHOOTING: MESSAGES CAUSE /CORRECTIVE ACTION If printer connection is not proper, then it will display Check printer power this message. check the printer connection. Disable Printer YES / Printer paper absent Disable...
  • Page 102: Decontamination

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 15. DECONTAMINATION 15.1. Decontamination Procedure: If the instrument is to be shipped after being exposed to potentially hazardous material, it should be decontaminated. The following procedure outlines the method of decontaminating the instrument before packaging and shipment.
  • Page 103: Procedure

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 15.4. Procedure: A solution of .5% Sodium Hypo Chlorite (NaOCL) solution (Bleach) is used. Commercial bleach is 5% NaOCL; household bleach is 3% NaOCL. When using commercial bleach, use a 10:1 mixture; if using household bleach, a 6:1 mixture is required.
  • Page 104: Safety Clearance Certificate

    MICRO TOUCH 1300 F13, User Manual , V3 09.08.2020 16. SAFETY CLEARANCE CERTIFICATE Please complete all information requests on this form prior to returning the instrument to the manufacturer or your local distributor for servicing, repairs or return. Thank you for your co- operation.