BaoLai Medical BCL-A Instruction Manual

BaoLai Medical BCL-A Instruction Manual

Built-in led curing light

Advertisement

Quick Links

成 型 尺 寸 :1 0 0×1 5 5 m m 材 质: 8 0 g双 胶 纸 黑 白 印 刷 骑 钉 成 本
B C L -A英文说明书
物料名称
物料编号
PA-BCL.A-M-E-1
版本
BCL-A
Built-in LED curing light
Instruction Manual
A1
设计
审核
批准
20190924
日期
日期
日期

Advertisement

loading

Summary of Contents for BaoLai Medical BCL-A

  • Page 1 BCL-A Built-in LED curing light Instruction Manual 成 型 尺 寸 :1 0 0×1 5 5 m m 材 质: 8 0 g双 胶 纸 黑 白 印 刷 骑 钉 成 本 B C L -A英文说明书 20190924 物料名称 日期...
  • Page 3 The light emittel may be harmful to the eyes Do not stare at the light source...
  • Page 5 Technical specifications :AC24V 50Hz. Power input :10VA. Input power :26×221mm. Dimensions :0.18kg. Net weight Components: details in packing list. LED light performance 4.6.1 LED power: 5W. :440nm~480nm. 4.6.2 Wave length : I. 4.6.3 Category 4.6.4 Light intensity :1000mW/cm²~1300mW/cm². Low power :1600mW/cm ~1800mW/cm .
  • Page 6 7.Cleaning and sterilization Cleaning 7.1.1Pull out of the lamp cap from BCL-A curing light, wipe it with medical alcohol or with special dental disinfection towel. Sterilization 7.2.1 Pull out of the lamp cap and pack it in sterilization pouch or sterilization box to sterilize it(135℃...
  • Page 7 8.Operation 8.1 Open the packing and check if all the parts and accessories are complete according to the packing list. 8.2 Ensure protection measures before use.For example, wear anti-glare glasses. 8.3 Pull out of the lamp cap and pack it in sterilization pouch or sterilization box to sterilize it in autoclave.
  • Page 8 If light intensity is not in the scope of setting, please send it back to us for repair. 9.8 Please use BCL-A curing light in the scope of cable length. Don't drag it please. Put it back to the hanger of dental unit after use.
  • Page 9: Maintenance

    10.3 Wipe surface of the lamp cap by purified water or disinfectant. 11.Trouble shooting Fault Possible cause Solution Check if BCL-A curing light BCL-A curing light connects dental unit without electric power. correctly or not. indication Curing light stops working Curing light works for a Non-act.
  • Page 10 13.After-sale service From the date of this equipment sold,if the equipment can't work normally because of the quality problems,our company will be responsible for the maintenance base on the warranty card. Please refer the warranty scope and period from warranty card. 14.Environmental protection The equipment has no any harmful ingredient.
  • Page 11 Guidance and manufacturer’s declaration-electromagnetic emissions The model BCL-A is intended for use in the electromagnetic environment specified below.The customer or the user of the model BCL-A should assure that it is used in such an environment. The model BCL-A is suitable for use in all...
  • Page 12 Guidance & Declaration — electromagnetic immunity The model BCL-A is intended for use in the electromagnetic environment specified beliw. The customer or the user of the model BCL-A shouid assure that lt is used in such an environment. Not use mains power.
  • Page 13 3 Vrms 3 Vrms 150 kHz to 150 kHz to 80MHz 80MHz 6 Vrms in ISM 6 Vrms in ISM and amateur and amateur radio bands radio bands d=[3,5/V1]×P 3 V/m 3 V/m 80 MHZ to 2.5 80 MHZ to 2.5 10 V/m 10 V/m 80 MHZ to 2.7...
  • Page 14 BCL-A BCL-A BCL-A...
  • Page 15 BCL-A BCL-A BCL-A BCL-A...
  • Page 16 PA-BCL.A-M-E-1 Rev.A1...