Do you have a question about the 909A and is the answer not in the manual?
Questions and answers
Summary of Contents for Keysight 909A
Page 1
Keysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 4
(Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation). K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 5
WEEE Directive Annex I, this product is classed as a “Monitoring and Control Instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local K eysight office, or see www.k eysight .com for more information. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 6
41 (21) 8113811 (Opt 2) Switzerland (German) 0800 80 53 53 (Opt 1) United Kingdom 44 (0) 118 9276201 Other European Countries: www.k eysight .com/find/contacts Or, go to www.k eysight .com/find/assist for more information. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
K eysight 909A/D Coxial Terminations Operating and Service Manual Introduction Product Overview 8 “909A Coaxial Termination" on page 8 “909D Coaxial Termination" on page 9 This chapter provides you the overview of K eysight 909A and 909D Coaxial Termination.
The K eysight 909A is furnished with a Precision 7mm connector. This is a sexless connector with low RF leakage and clearly defined reference plane. As an option, the K eysight 909A can be furnished with either male or female Type- N connector interfaces per IEEE standard 287 GPC.
909D Coaxial Termination The K eysight 909D has Precision 3.5mm connector interfaces per IEEE standard 287 GPC. Figure 1-2 K eysight 909D Coaxial Terminations K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 12
K eysight 909A/D Coaxial Terminations Operating and Service Manual...
K eysight 909A/D Coxial Terminations Operating and Service Manual Specifications Product Specifications 12 Physical Specifications 13 Pin Depth Specifications 13 This chapter provides you the specifications of K eysight 909A and 909D coaxial terminations.
• 6 to 26.5GHz : 0.06 (1.12 SWR) Power Rating 2W average 2W average 300W peak 100W peak (10ps max. pulse width) at 20°C * Derated to 1W average at 75°C K eysight 909A/D Coaxial Terminations Operating and Service Manual...
K eysight 909D Pin Depth in Micrometers (10 inches) Option 011 Option 040 Option 301 0 to -78.8 0 to -78.8 0 to -78.8 (0 to -20) (0 to -20) (0 to -20) K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 16
Relative humidity Always non-condensing • Operation 0% to 80% (26°C maximum dry bulb) • Storage 0% to 95% Vibration 0.015”, 5-55-5Hz, 15min, 3 axes Shock 100g, 1-2ms, 3 times each plane K eysight 909A/D Coaxial Terminations Operating and Service Manual...
K eysight 909A/D Coxial Terminations Operating and Service Manual Installation and Service Installation 16 “Initial Inspection" on page 16 Service Instructions 17 “Adjustment" on page 17 “Repair" on page 17 Maintenance 17 “Cleaning Connectors" on page 17 “Connections" on page 19 “Handling and Storage"...
Mark the container FRAGILE to insure careful handling. In any correspondence, refer to the instrument by model number and serial number. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
1 Use Compressed Air or Nitrogen Always use protective eyewear when using compressed air or WA RNI NG nitrogen. Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 20
Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low- pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
3 If necessary, clean the connectors. 4 Use a connector gage to verify that all center conductors are within the observed pin depth values in “Pin Depth Specifications" on page 13. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 22
135N- cm (12in- lb) 8710- 1766 3.5mm 90N- cm (8in- lb) 8710- 1765 * Type-N connectors can be tightened hand tight, but for better repeatability they should be torqued to 135N-cm (12in-lb) K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 23
(Reduces lift on connection) (Too much lift on connection) Figure 3-3 Wrench Positions 4 Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3- K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 24
Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more than the recommended torque. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
• Do not set connectors contact- end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Page 26
K eysight 909A/D Coaxial Terminations Operating and Service Manual...
Need help?
Do you have a question about the 909A and is the answer not in the manual?
Questions and answers