Summary of Contents for Avante Volumax VOS
Page 1
Knob Run/Standby button This guide provides you with basic instructions on how to set up and operate the Volumax VOS. Before operating the Avante Volumax VOS please review the Operators Instruction Manual. 1578-2022-03-31 2601 Stanley Gault Pkwy, Suite 101 Louisville, KY 40223 USA •...
Page 2
Open door and install Disposable Water Path into docking station • Hang sterile water bag • Wipe water spike with alcohol pad and insert into sterile water bag • • Volumax VOS is ready for start up START UP AND ADJUSTING PARAMETERS • Install oxygen sensor • • •...
Page 3
Avante Volumax VOS Quick Reference Guide ALARM ICON INDICATES CAUSE ACTION Malfunction of sensor or Internal component Cannot be corrected by user: control system failure send for service. GENERAL FAULT (FLASHING) Depleted or defective GENERAL FAULT % O₂ (FLASHING) DISPLAYS DASHES ( - - )
Page 4
Avante Volumax VOS Quick Reference Guide ALARM ICON INDICATES CAUSE ACTION Cartridge and/or RUN mode: faulty Disconnect patient. Remove disposable DPC not sensor or cartridge patient circuit. Check cartridge detected. Unit will not detected. installation. Check sensor windows not run.