Problem
Cause
Solution
Problem
Cause
Solution
Problem
Cause
Solution
Problem
Cause
Solution
Problem
Cause
Solution
Problem
Cause
Solution
Problem
Cause
Solution
Problem
Cause
Solution
The flow rate control is variating
Potentiometer issue
Contact Spectra Technician for replacement
Pump only flows in one direction
Flow direction Toggle switch issue
Contact Spectra Technician for replacement
The Pump head rollers are spinning but the
silicone tubing is dragged out of pump head
One of the rollers must be stuck due to corrosion
Contact Spectra Technician for Pump head
maintenance
The Motor is running but the Pump head rollers
are not spinning
Coupling Displacement
Contact Spectra Technician
The pump does not draw water
Water level too deep
Check for Water levels - Max at 25 Feet (7 m)
The pump does not draw water
Check for Silicone Tubing
Make sure it matches the pump head
specifications - LS 15, 24, 35, 36
It does not have much drawing power
Check for Battery Voltage
Charge the Battery
Unit does not turn ON
Popping of Circuit Breaker
Press the Circuit breaker inside the Vial
Compartment