9046. User information ................................2 A 1.1 Meaning of the pictograms ............................2 A 1.2 Important information ..............................2 A 1.3 Precautions ................................2 A 1.4 Purpose and potential applications ........................3 A 1.5 Technical data ................................3 Description ..................................4 Operation ..................................5 Maintenance ..................................6 Malfunctions ..................................6 Installation ..................................8...
The following should also be observed: request these from your responsible KaVo agent. Duplication and distribution of the • the work safety regulations, Important information for operator instructions for use (IU) require KaVo's and engineer.
9046. A 1.4 Purpose and potential applica- tions The extension set 9046 is used for extending the ABSORmatic plus from 3 to 6 suction points. Max. 4 suction points can be put into operation simultaneously. A 1.5 Technical data Suction path diverter...
9046. A 2 Description Supplementary control ” @ Jacks of external control inputs for suction points 2.1 to 2.4. ” Adjusting wheel for inlet air pressure reducer · # Manometer for regulated inlet air pressure £ £ Mains switch fi Mains jack Ì...
9046. A 3 Operation When products are operated for the first time and for every operation ” thereafter, it is essential to note the points mentioned in the section «A 1.3 Safety measures » ! Switch on the supplementary control ·...
9046. A 4 Maintenance Repair and maintenance work on the electrical part of the unit may be performed only by qualified specialists or trained per- sons who have been made aware of the safety regulations. For maintenance work, disconnect the mains plug or isolate the unit completely from the mains supply, thus cutting off power to the unit.
Page 9
9046. F 04 • CAN fault C • An error occurred during the software initialization of the CAN module. CAN is not ready. R • Switch the unit off and on at the mains switch. If this error occurs several times in succession, call the Customer Service.
Foot push button 1.000.3147 B 2 Requirements for installation and operation of the extension set • KaVo ABSORmatic plus from software version 3.0. (For checking of the software version, see ABSORmatic plus instructions for use.) With software versions 2.0 to 2.9, the replacement circuit board 1.000.3134...
9046. The installation is performed in four steps: B 3 Installation of the CAN printed circuit board in the suction path switching module of the ABSORmatic plus. B 4 Mounting of the suction path diverter and connection with the suction hoses.
9046. B 4 Mounting of the suction path diverter Mounting location Mount the suction path diverter under the bench on the technician’s workstation. The mobility of the Bunsen burner tubes may not be restricted through the mounting of the suction path diverter and the suction hoses.
Make CAN connection # between supple- mentary control and ABSORmatic plus. Connect external sealed keypads (option) · via KaVo cable fittings £ to the sub-D jacks. Check for correct assignment. Connect dust-generating consumers to the sockets fi. Connect control lines @ of the external...
9046. B 6 Commissioning Check for correct connection of all compo- nents. Using the adjusting wheel @, set the inlet pressure on the manometer ” to 0.8 bar. Switch on the supplementary control at the mains switch # and then switch on the ABSORmatic plus at the mains switch.
6 months from the date of purchase certified by the sales person. After expiry of the guarantee period, KaVo undertakes to provide a guarantee for a further 6 months for damage due to defects in the material or in the manufacture.