  • Page 1 OWNER’S MANUAL Flat Panel Digital X-ray Detector Please read this manual carefully before operating your set and retain it for future reference. 17HQ701G-B www.lg.com Copyright © 2022 LG Electronics Inc. All Rights Reserved.
  • Page 2: Table Of Contents

    CONTENTS OPEN SOURCE SOFTWARE NOTICE INFORMATION ..3 ON CLEANING ............3 GENERAL DESCRIPTION ..........4 NAMES AND FUNCTIONS OF COMPONENTS ....7 ASSEMBLING BATTERY ..........15 REMOVING BATTERY ..........16 HOW TO USE MAIN CABLE HOLDER ......18 SPECIFICATION AND DIMENSION OF EACH PART ..19 ENVIRONMENTAL REQUIREMENT ......23 INSTALLING CALIBRATION SOFTWARE .......24 CONNECTION TYPE ..........25...
  • Page 3: Open Source Software Notice Information

    . • Isopropanol 70 % LG Electronics will also provide open source code to you on CD-ROM for a charge covering the cost of performing such • Ethanol 70 % distribution (such as the cost of media, shipping, and handling) upon email request to  opensource@lge.com .
  • Page 4: General Description

    GENERAL DESCRIPTION Overview This model is an x-ray imaging device, a system that can acquire and process X-ray images as digital images. It utilizes amorphous silicon and a high-performance scintillator to ensure sharp high-definition image quality with the resolution of 3.6 lp/mm and the pixel pitches of 140 um. This device is a flat panel based X-ray image acquisition device. This device must be used in conjunction with an operating PC and an X-ray generator. This device can be used for digitizing and transferring X-ray images for radiological diagnosis.
  • Page 5 Product Components Basic Accessories Detector Battery 2 EA AC Power Cord for the AC Power Adapter Inspection Report Owner's Manual/Regulatory Manual/Calibration Software Main Cable AC Power Adapter for Charger Battery Charger Control Box AC Power Cord for the Control Box Main Cable Holder...
  • Page 6 Optional Accessories Trigger Cable LAN Cable • Some models may not include optional accessories. CAUTION • You must use the authorized components as per the specification below. Unauthorized components may cause damage and/or cause the product to malfunction. Component Standard LAN Cable More than CAT5E Standard Power Cord...
  • Page 7: Names And Functions Of Components

    NAMES AND FUNCTIONS OF COMPONENTS Detector Front Back Battery...
  • Page 8 Side Connection to Main Cable • Power LED Indicator • Power button • Wired/Wireless LED Indicator • Wired/Wireless Connection Button • Function Switch LED Indicator • Function Switch Button OLED Indicator...
  • Page 9 Button Information Button Description Press the Power button to turn the power on or off. (Power button) – On: Press and hold for 1 sec or longer – Off: Press and hold for 5 sec or longer Press this button for at least one second to switch between the following connection modes, in respective order: Ethernet/Station/AP mode. (Wired/Wireless Connection button) (Function Switch button) Press this button for at least one second to switch between the following menus, in respective order: Check connection mode, video acquisition, image auto save.
  • Page 10 LED Indicator Description Displays the power and battery status of the detector. (Power LED Indicator) Power off White Power on Orange The battery level is greater than 10 % and less than 30%. Orange(blinking) The battery level is less than 10%. Displays the connection mode status of the detector.
  • Page 11 OLED Indicator The OLED indicator displays the information below. Connection Mode Information Ethernet Station Wired/STA/AP – Check Connection Mode (Ethernet/Station/AP). Dynamic On/Off – Video Acquisition Auto save On/Off (Up to 200 images) – Image Auto save NOTE • The information displayed on the OLED indicator will vary depending on the connection mode (Ethernet/Station/AP). •...
  • Page 12 Battery and Battery Charger Blinking Blinking Blinking LED Indicators on the Battery Charger LED Indicators on the Battery Charger Remaining 0 ~ 30 % 30 ~ 70 % 70 ~ 99 % 100 % Battery Levels Completion of Battery Status On charging charging NOTE •...
  • Page 13 Control Box Front Back Number LED Indicator LED Color Description Power Green Power normal operation Power off (AC power cord no connection or Power error) Ethernet Green Ethernet normal operation Green(blinking) On data communication Ethernet disconnected Ready Green Ready signal from X-ray Generator is active Ready signal from X-ray Generator is inactive Orange(blinking) Power error...
  • Page 14 Number Port Description DXD 1 Connects the Control Box to the Detector A. This connector supplies power (24 V 2.1 A) to the Detector and transmits the X-ray generator's sync signal and Ethernet image data. DXD 2 Connects the Control Box to the Detector B. This connector supplies power (24 V 2.1 A) to the Detector and transmits the X-ray generator's sync signal and Ethernet image data.
  • Page 15: Assembling Battery

    ASSEMBLING BATTERY Check the direction of the holes located inside the detector. Align the battery, and slide it into the holes inside the detector. Press the other side to assemble the battery.
  • Page 16: Removing Battery

    REMOVING BATTERY Slide the battery lock button in the direction shown in the figure. Remove the battery by lifting it in the direction shown in the figure.
  • Page 17 Battery Hot-Swap The detector is equipped with a hot-swappable battery. Hot-Swap is activated for one minute after the battery is removed from the detector. NOTE • The detector will turn off if a charged battery is not inserted on time. •...
  • Page 18: How To Use Main Cable Holder

    HOW TO USE MAIN CABLE HOLDER Clean the wall before attaching the holder to it. Attach the main cable holder using adhesive tape on the back. Store the main cable of the detector in the holder. NOTE • The strength of the adhesive may be weakened depending on the environment. Avoid attaching the holder too high on the wall. •...
  • Page 19: Specification And Dimension Of Each Part

    Images can be saved by the X-ray generator while the power of the detector is turned on without connecting to a PC. To produce images, X-ray is irradiated at intervals of more than 10 seconds. Check and load the saved images Typ. 66 % @ 0.1 lp/mm from LG Acquisition Workstation Software.
  • Page 20 Detector has been tested with below table’ s X-ray condition. This table is only for reference. The legally certified • Sensor Type: Oxide TFT, X-ray conditions radiologist expert should control X-ray dose. Adult Tube Current x Time • Sensor Type: a-Si TFT, X-ray conditions SID (inch / cm) Tube Voltage (KV) Tube Current (mA)
  • Page 21 GRID CAUTION • Always use the detector in places that meet the following environmental requirements. Item Recommended Specifications 100 cm / 130 cm / 150 cm / 180 cm Item Units (39.3 inch / 51.1 inch / 59 inch / 70.8 inch) Temperature (Storage) -20 (-4) 60 (140) °C (°F)
  • Page 22 Control Box Item Specifications Item Specifications Model LG Battery Charger Model LG Control Box Size (Width x Height x Depth) 268.4 x 54.9 x 186.9 mm (10.5 x 2.1 x 7.3 inch) Size (Width x Height x Depth) 125.0 x 109.8 x 255.0 mm (4.9 x 4.3 x 10.0 inch) Weight Typ.
  • Page 23: Environmental Requirement

    ENVIRONMENTAL REQUIREMENT Cables Item Length PC System Requirements Main Cable 7 m (22.9 ft) LAN cable (Optional) 10 m (32.8 ft) PC Specifications Power cord (110 V or 220 V) 1.5 m (4.9 ft) Intel i5 Trigger Cable (Optional) 15 m (49.2 ft) 8 GB Disk capacity 500 GB or more recommended...
  • Page 24: Installing Calibration Software

    Click and launch Control Panel in your Start menu. Select Programs and Features in Control Panel. Select [LG DXD Calibration] from the list of the installed programs. Select the [Delete] button when the screen for installation and uninstallation of the program appears.
  • Page 25: Connection Type

    CONNECTION TYPE Connection of X-ray Generator - Detector Select the trigger mode according to the image acquisition method. – Auto mode: The detector automatically detects and acquires X-ray images. – Manual mode: The detector recognizes the X-ray generator signals and acquires images. Connection of Detector - PC The status of the connection between the detector and a PC –...
  • Page 26 Trigger Cable The trigger cable is connected to the control box and X-ray Generator. It can only be used in manual mode, and not in automatic mode. PREP_REQ CANCEL_REQ_GND EXP_ACK_GND SEN_LEVEL_A SEN_LEVEL_B_COM POWER EXP_REQ PREP_ACK_GND EXP_ACK SEN_LEVEL_C EXT_TX CANCEL_REQ_VCC X-ray Generator PREP_ACK EXT_RX SEN_LEVEL_B...
  • Page 27 Number Pin Label Description PREP_REQ Prepare the signal from the X-ray generator to the control box. CANCEL_REQ_GND Cancel REQ ground. EXP_ACK_GND Signal ground SEN_LEVEL_A READY SEN_LEVEL_B_COM READY POWER Power: X-ray generator supply voltage (DC 12 V to 24 V) EXP_REQ Expose the signal from the X-ray generator to the control box. PREP_ACK_GND Prepare ground check.
  • Page 28 Block diagram of Trigger Cable connection X-ray Generator Control Box DC 12-24 V PREPARE PREP_REQ EXPOSE EXP_REQ <Connection of X-ray Generator - Control Box>...
  • Page 29 X-ray Generator Control Box EXP_ACK EXP_ACK_COM EXPOSE <Assembly Diagram> EXP_REQ EXP_ACK <Timing Chart>...
  • Page 30 Connecting Detector to PC (Wired Mode) A: Detector D: LAN Cable G: X-ray Switch J: X-ray Generator Interface B: Main Cable E: PC H: Power Cord (AC 100-240 V~) K: Trigger Cable C: Control Box F: Object I: X-ray Generator Auto Mode Manual Mode...
  • Page 31 Connecting the Cables Main Cable: Connects the control box to the detector. Up to two detectors can be connected, and both ports can be used by one detector. LAN Cable: Connects between Control Box and PC. Trigger Cable: Connects between control box and X-ray generator. In case of auto mode, this connection is not necessary.
  • Page 32 Launch Calibration Software, go to (Settings) > [DXD] > [Connection & File Save] and enter the detector's Connect - Wired Connection IP (10.10.10.100). Click the [Ping] button to check the connection. Use the LAN cable to connect a PC to the Control Box and connect the detector to the Control Box with the main cable. After connecting the cables, proceed to set up the PC as shown below.
  • Page 33 Connecting Detector to PC (Wireless Mode) A: Detector D: PC G: X-ray Generator J: Control Box B: AP E: Object H: X-ray Generator Interface K: Power Cord (AC 100-240 V~) C: LAN Cable F: X-ray Switch I: Main Cable L: Trigger Cable Auto Mode 1.
  • Page 34 Manual Mode...
  • Page 35 • AP mode (connection via detector AP) - SSID: LGEDXD_SOFTAP It is possible to change the Wireless Settings using LG DXD Calibration Software. - Please see "Wireless AP Settings" for details. Press (Wired/Wireless Connection button) for at least one second to switch between the following connection modes, in respective order: Ethernet/Station/AP mode.
  • Page 36 The model and serial number of the product are located on the back and on one side of the product. Record them below in case you ever need service. Responsible Party LG Electronics USA, Inc. Address 111 Sylvan Avenue, North Building, Englewood...

Table of Contents