Fresenius Vial OPTIMA PT Operator's Manual

Fresenius Vial OPTIMA PT Operator's Manual

Volumetric infusion pump

Advertisement

Quick Links

OPERATOR'S GUIDE
Volumetric
Infusion Pump
Introduction
Great care has been taken in the design and manufacture of the
Optima PT, a volumetric infusion device.
The configuration flexibility of the Optima PT provides overall
improvement in the working conditions of medical teams, thus increasing
patient safety.
The use of this material requires great care. The user must be able to handle the instrument
properly and must know how to fully operate.
Please read the operator's guide carefully before putting the device into use.
237,0$

Table of contents

1.
Operations for use.......................................................................2
2.
Advanced functions ....................................................................4
3.
Optima PT - Internal safety features...........................................7
4.
Optima PT - Performances........................................................10
5.
Optima PT - Technical characteristics .....................................11
6.
Configuration menu ..................................................................13
7.
Precautions to be taken ............................................................15
8.
Maintenance recommendations ...............................................16
9.
RS 232 connection ....................................................................17
10.
Nurse call connection ...............................................................17
11.
Drip sensor ................................................................................17
12.
Function with the internal battery ............................................17
13.
Accessories ...............................................................................18
14.
Disposable .................................................................................19
15.
Conditions of guarantee ...........................................................20
16.
Useful addresses.......................................................................24
37

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OPTIMA PT and is the answer not in the manual?

Questions and answers

Summary of Contents for Fresenius Vial OPTIMA PT

  • Page 1: Table Of Contents

    Optima PT - Internal safety features...........7 Optima PT, a volumetric infusion device. Optima PT - Performances............10 The configuration flexibility of the Optima PT provides overall Optima PT - Technical characteristics ........11 improvement in the working conditions of medical teams, thus increasing Configuration menu ..............13...
  • Page 2: Operations For Use

    1.1. Installation of the Optima PT infusion set Special attention must be paid to the stability of the Optima PT. Open the pump door by lifting the door handle. The pump automatically The Optima PT can be used on mains or battery switches on when connected to mains (if configurated).
  • Page 3 Select and prepare new set as indicated above. Program and start infusion as indicated above. 1.8. Turning off the Optima PT Press , then for more than 2 seconds. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 3 -...
  • Page 4: Advanced Functions

              P P O O WRWD D O O L L Q Q   K K     Confirm by pressing . Then, press to escape. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 4 -...
  • Page 5 Î Drop sensor Note: the set must be equipped with a one way valve to prevent the Í secondary solution from filling the MS 80 primary container. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 5 -...
  • Page 6 The sequential program is not modified. If sequence program is modified during sequential infusion, only forthcoming sequences will be modified. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 6 -...
  • Page 7: Optima Pt - Internal Safety Features

    Optima PT - Internal safety features Optima PT has a continuous inspection system which functions as soon as the pump is in use. Any internal failure or anomaly detected will be immediately displayed. Nevertheless, the qualified personnel in your establishment or our After-Sales Department should always be notified of any abnormal function where no specific cause can be found.
  • Page 8 The pump has been out of Set date and time and charge the battery. Set date and time.   mains power for a long time. 3URFHHG WR EDWWHU\ FKDUJH NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 8 -...
  • Page 9 3.2. Quick check This test is recommended or mandatory (in case of in-force law). It allows a complete alarm and safety features check of the Optima PT (no patient connected). Serial number (ID/N): ......Date: ../ ../ ..
  • Page 10: Optima Pt - Performances

    Activated when the volume to infuse is reached. 3 ml/h for 30 minutes for flow rate selection above 3ml/h, set flow rate for 30 minutes for flow rate below 3 ml/h. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 10 -...
  • Page 11: Optima Pt - Technical Characteristics

    13.5 x 17.5 x 14.5 cm Height / Width / Depth The last 764 dated events are stored and can be read with the Weight approx. 2.9 Kg maintenance software. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 11 -...
  • Page 12 Set flow rate Set flow rate 1000 1500 2000 2500 3000 3500 4000 1000 1500 2000 2500 3000 time (seconds) time (second) NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 12 -...
  • Page 13: Configuration Menu

    Configuration menu Fresenius Vial recommends the presence of its qualified personnel or a member of the Technical Department of your establishment to help you implement the configuration procedures you wish to choose. Note: press to cancel modification at any time - Press OFF to leave configuration mode at any time.
  • Page 14           % % D D XG XGV V           % % D D XG XGV V NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 14 -...
  • Page 15: Precautions To Be Taken

    EN 60 601-1 Standard. The pump may only be connected to the mains with the power cord Fresenius Vial will not be liable for any damages or claims, supplied by the manufacturer. Check that the mains voltage corresponds medical or otherwise, of any nature whatsoever, whether direct or with the value indicated on the label placed underneath the device.
  • Page 16: Maintenance Recommendations

    Use a cloth soaked in DETERGENT-DISINFECTANT, previously Fresenius Vial is not liable for loss or damage to the device during diluted with water if required, to destroy micro-organisms. Avoid transport to our After-Sales Department.
  • Page 17: Rs 232 Connection

    Connect the drop detector to the drip chamber. Important: control the right positioning of the drip chamber and check there are no drops on the drip chamber walls. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 17 -...
  • Page 18: Accessories

    Added to high stability rolling stand 180, for live loads higher than 16 kg. Transrail 120 - Cat # 073071 Two clamps which firmly fix a burnished pole on 2 parallel rails. NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 18 -...
  • Page 19: Disposable

    14.1. VS range 14.2. Other PVC giving sets In order to use the Optima PT pump in best conditions and cover Fresenius Vial recommends the use of the following sets : most applications, we have developed a complete range of sterile MCM400 infusion set (Fresenius MCM) disposables: the VS series.
  • Page 20: Conditions Of Guarantee

    15. Conditions of guarantee Fresenius Vial guarantee that this product is free from defects in materials and workmanship (excluding batteries and accessories) for a period of one year from the date of invoice. If you comply to benefit from the materials and workmanship guarantee from our...
  • Page 21 Notes NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 21 -...
  • Page 22 Notes NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 22 -...
  • Page 23 Notes NU OPTIMA PT IS - 1273 rev 0 – 03/12/02 - 23 -...
  • Page 24: Useful Addresses

    As standards and equipment change from time to time, the features shown and described in this document must be confirmed by our departments. This document may not be reproduced in whole or in part without the written consent of Fresenius Vial. Fresenius Vial - Le Grand Chemin - F-38590 BREZINS (FRANCE)

Table of Contents