Short Programming Guide - Flow Mode (mL/hr)
Caution:
Users must familiarise themselves with the full user
instructions. These notes provide only a reminder of main functions.

Pump activation and mode selection

The default delivery setting for the S-PID 100 is timed mode. This
allows programming in time of infusion. It is possible to change the
mode of operation by removing and replacing the battery.
The pump will then perform a short self test procedure during which
a full screen display of all elements should appear.
When the symbol [-t-] is displayed press and hold the [P] button until
the display flashes. Press the [-] button until the display changes to
[-F-]. Press [P] to confirm and the pump will switch to [OFF] mode.
.

Locking/unlocking the keypad

By default keyboard lock is not activated on this device.
With the pump displaying [OFF] press the (-) key for approximately
7 seconds. A number representing the number of completed
infusions will appear on the screen.
While still holding the [-] key depress the [+] key. The screen will
display flashing [L.0] and keyboard is unlocked.
Use the (+) or (-) button to select required value. (L1 is on, L0 is
off)
This symbol will display when the keyboard is locked.
Crono
S-PID 100

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Crono S-PID 100 and is the answer not in the manual?

Questions and answers

Summary of Contents for AMT Crono S-PID 100

  • Page 1 Crono S-PID 100 Short Programming Guide - Flow Mode (mL/hr) Caution: Users must familiarise themselves with the full user instructions. These notes provide only a reminder of main functions. Pump activation and mode selection The default delivery setting for the S-PID 100 is timed mode. This allows programming in time of infusion.
  • Page 2 Programming the pump With the keyboard unlocked, and the pump displaying [OFF] pressing the [P] button will allow you to step through a menu to set the following parameters in turn. Set Flow rate 0.2 mL/hr to 300 mL/hr End of infusion alarm On/Off Infusion Break Points IP.1 to IP.5 (Normally set to IP.1)

Table of Contents