Page 2
In no event shall ThermoQuest or Forma Scientific Division be held liable for any damages, direct or incidental, arising out of or related to the use of this manual. When contacting the factory regarding this unit, have the model, serial number and date of purchase available.
Page 3
Monday through Friday from 8:00 a.m. to 7:00 p.m. Eastern Time. Please contact us by telephone or fax. If you wish to write, our mailing address is: ThermoQuest Forma Scientific Division Millcreek Road, PO Box 649 Marietta, OH 45750 International customers please contact your local ThermoQuest, Forma Scientific Division distributor.
Page 4
Forma® Model 4535/4536 _______________________________________________Safety General Safety Notes used in this Manual Important operating and/or maintenance instructions. Read the accompanying text carefully. Ce symbole attire l’attention de l’utilisateur sur des instructions importantes de fonctionnement et/ou d’entretien. Il peut être utilisé seul ou avec d’autres symboles de sécurité.
Page 5
Forma® Model 4535/4536 _______________________________________________Safety Hot surface(s) present which may cause burns to unprotected skin or to materials which may be damaged by elevated temperatures. Présence de surface(s) chaude(s) pouvant causer des brûlures sur la peau non protégée, ou sur des matières pouvant être endommagées par des températures élevées.
Forma® Model 4535/4536 _______________________________________Contents Table of Contents Page Section 1 – Receiving 1.1 Packing List ..................1-1 1.2 Pallet Hold-Down Shipping Brackets..........1-1 Section 2 - Installation 2.1 Location................... 2-1 2.2 Installing the Cabinet Stand-off bolts..........2-1 2.3 Chamber Drain ................2-2 a.
Page 7
Forma® Model 4535/4536 _______________________________________Contents Section 3 – Operation 3.1 Introduction ..................3-1 3.2 Factory Settings ................3-2 a. At Initial Start-Up ............... 3-2 b. Pre-Set Programs ................ 3-2 c. Other Factory Default Settings ............ 3-3 3.3 Operating the Control Panel............. 3-3 3.4 Quick Start Up .................
Page 8
Forma® Model 4535/4536 _______________________________________Contents Charts: Quick Reference Guide 3-1 Changing the Operating Time, RPM or Temperature Settings 3-2 Setting the Hold Time to Zero 3-3 Loading a Pre-Set Program 3-4 Editing the Pre-Set Programs 3-5 Setting or Changing the Access Code, Remote Alarms, or Unit Configuration 3-6 Reading the Accumulated Run Time 3-7 Setting the RPM Tracking Limit...
Forma® Model 4535/4536 ______________________________________Receiving Section 1 - Receiving 1.1 Packing List The Model 4535 Orbital Incubator Console Shaker is supplied with the following materials: 2 - Keys for the lid lock (packaged and attached to the outside of the unit) 1 - T-handle 5/32"...
Forma® Model 4535/4536________________________________________Installation Section 2 - Installation Figure 2-1 Forma Model 4535 Console Forma Scientific Incubator/Shaker Orbital Shaker 2.1 Location Locate the shaker on a firm, level surface in an area free of dust and dirt. To allow for lid opening, the top of the shaker must be at least 4-1/2 inches from the back wall or other obstruction.
Forma® Model 4535/4536________________________________________Installation 2.3 Chamber Drain A drain is provided in the bottom of the chamber for convenience when cleaning or removing spills. A clear vinyl hose and plastic valve is connected to the drain and is accessed by removing the front grille assembly and the lower front panel. The grille is removed by gently pulling it off.
Forma® Model 4535/4536________________________________________Installation The 2 liter, 2.8 liter, 4 liter, 5 liter and 6 liter flask clips use two metal springs and rubber spring tubes. On these larger clips, the springs are installed by hooking their ends together as illustrated in Figure 2-5. The upper spring and spring tubes should be installed prior to mounting the clip to the platform.
Forma® Model 4535/4536________________________________________Installation Figure 2-4 Installing the flask clip assembly with a single screw. 250/300 ml flask shown with cushion pad installed. 2.7 Installing the Test Tube Holders Forma Accessory Test Tube Racks and Test Tube Rack Holders are available in four sizes and are listed in Table 2-2.
Page 14
Forma® Model 4535/4536________________________________________Installation Figure 2-6 Forma Test Tube Rack showing the swing/tilt mechanism. Figure 2-7 Forma test tube rack holder with the test tube rack removed Table 2-2 Accessory Test Tube Racks and Adjustable Test Tube Rack Holder Part Number Description 950040 Test Tube Rack, 10-13 mm size...
Forma® Model 4535/4536________________________________________Installation 2.8 The RS-232 Interface Connector The Model 4535 Orbital Shaker is equipped with an RS-232 Serial Communication Interface for the remote transmission of data. An RJ-11 telephone style connector is located on the rear of the incubator. A cable with RJ-11 plugs, and an RJ-11 to DB-25 adapter are required.
Forma® Model 4535/4536________________________________________Installation The data transfer sequence is transmitted in the following format. X refers to the numerical time, RPM and temperature. (NUL)XX.XX(H)(SP)(SP)XXXRPM(SP)(SP)XX.XC(SP)(LF)(CR)(EOT) NUL - Null character (0) - Space - Line feed - Carriage return - End of text (4) - Hold Mode The Model 4535 transmits time, RPM and temperature information one minute after power is first applied to the unit and then every 60 minutes.
Forma® Model 4535/4536________________________________________Installation Figure 2-7 identifies the pin contacts. Figure 2-10 shows the location of the Remote Alarm Connector. A modular to modular cable, Forma Stock No. 190388 and a RJ-11 telephone style terminal converter, Forma Stock No. 190392 or equivalent may be used to convert to a screw terminal connection.
Forma® Model 4535/4536________________________________________Installation Figure 2-11 Rear and Side Panels of the Model 4535 Shaker/Incubator 2.11 Connecting the Unit to Electrical Power See the serial tag on the side of the unit for electrical specifications or refer to the electrical schematics at the end of this manual. Connect the orbital shaker to a grounded, dedicated circuit.
Forma® Model 4535/4536 ______________________________________Operation Section 3 - Operation Figure 3-1 Incubator Console Shaker Forma Scientific Orbital Shaker 3.1 Introduction The Forma Model 4535/36 is a microprocessor controlled console orbital shaker designed to accommodate a wide variety of flasks, test tubes and other glassware.
Forma® Model 4535/4536 ______________________________________Operation 3.2 Factory Settings a. At Initial Start-Up The Model 4535 has been shipped from the factory with the following settings which will appear on the display screen when the shaker is powered up for the first time.
Forma® Model 4535/4536 ______________________________________Operation Instructions on operating the shaker using these pre-set programs are found in Section 3.6. Instructions to edit these programs are in Section 3.7. c. Other factory default settings Function Default Reference Manual Section Audible Alarm Access Code 0 0 0 0 3.10 RPM Tracking Limit...
Forma® Model 4535/4536 ______________________________________Operation entered or the operating programs modified, the functions of the buttons will change. Their new function will be identified on the portion of the LCD above the button. Figures 3-3 and 3-4 illustrate how the functions of these three buttons are identified.
Forma® Model 4535/4536 ______________________________________Operation Figure 3-5 This display is a typical Start-Up or Operating screen. Any programming or setting changes are made starting from this screen. 3.5 Changing the Shaker Time, Speed and Temperature Settings When first turned on or when the shaker is operating, the Time, RPM and Temperature values are displayed on the LCD.
To create a new set of operating parameters for Pre-Set Program P1, P2 or P3, refer to Chart 3-4. 3.8 Changing/Setting the Access Code Refer to Chart 3-5 Record and file the new Access Code in a safe place. If the new code is lost or forgotten, contact the Forma Scientific Service Department, 1-888-213-1790.
Forma® Model 4535/4536 ______________________________________Operation 3.9 Setting the Remote Alarms Refer to Chart 3-5 The Forma 4535 and 4536 Orbital Shaker control system monitors and provides alarms for six operating parameters : Cycle Complete Drive Belt Integrity Power Failure Sensor Fault RPM Tracking Temperature High/Low Each of these alarms may be independently turned on or off to suit operator or...
Forma® Model 4535/4536 ______________________________________Operation Figure 3-6 Alarm warnings and messages The alarm condition and alarm messages shown in Figure 3-6 are interpreted as: • Accumulated operating time - 12 hours, 36 minutes • Operating speed - 250 RPM • Operating temperature - 35.0° C •...
Page 27
Forma® Model 4535/4536 ______________________________________Operation c. Over Temperature Alarm and Over Temperature Shutdown When the shaker's chamber temperature exceeds the alarm set point, both audible and visual warnings are initiated and the shaker heaters are shut down. The platform will continue to operate, however, and the system will reset once the temperature has lowered past the alarm point, although the warning message will continue to be displayed.
Forma® Model 4535/4536 ______________________________________Operation 3.13 Temperature Calibration The Model 4535/36 temperature monitoring system may be calibrated +4°C from the factory setting. From the Operating Screen: 1. Press Right Arrow twice. 6. When set, press Temp to return to the Over-Temp, Calibrate and Alarms screen.
Forma® Model 4535/4536 ______________________________________Operation 3.16 Reading the Over Temperature Probe To read the temperature being sensed by the Over Temperature Probe, follow the key-press sequence below. NOTE: This is a monitoring function only. Changes to the temperature alarm setpoint cannot be made from this display. From the Operating Screen: 4.
Forma® Model 4535/4536 ____________________________________Maintenance Section 4 – Maintenance 4.1 General Cleaning and Lubrication The Model 4535/4536 Shaker uses a brushless DC motor and oversized, permanently-lubricated bearings that require no maintenance. a. Shaker Platform and Cabinet The anodized brushed aluminum platform and powder-coated steel cabinet surfaces can be cleaned with common laboratory materials.
Forma® Model 4535/4536 ________________________________________Service Section 5 - Service The procedures outlined in this section must be performed by personnel qualified and knowledgeable in laboratory equipment service and repair. Whenever any panel is removed from the cabinet of this unit, electrical power must be shut off and the power cord unplugged.
Forma® Model 4535/4536 ________________________________________Service 5.2 Changing the HEPA filter P/N 760164 The HEPA filter is located on the left side of the chamber and is accessed by pulling up on the four, black press-in fasteners and sliding the cover off. Refer to the illustration in Figure 5-1.
Page 33
Forma® Model 4535/4536 ________________________________________Service To remove the panel beneath the grille, remove six Phillips screws, three on the bottom of the panel and three on the top. Loosen the two Phillips screws on the left side of the foot pedal to allow the panel to slide outward. Figure 5-2 Underside of the control panel frame...
Forma® Model 4535/4536 ________________________________________Service Do not substitute! Replace these fuses only with fuses of identical electrical ratings. 5.5 Circuit Boards (Qualified repair personnel only) Five circuit boards control the Orbital Console Shaker. Four boards are located in the relay tray compartment; the fifth is behind the liquid crystal display. Figure 5-4 identifies the circuit boards and other major components in the relay tray.
Forma® Model 4535/4536 ________________________________________Service 5.6 Temperature Sensors Two temperature sensors are located behind a perforated cover plate on the right side of the chamber. (Refer to Figure 5-5) To access these sensors, pull outward on the two black press-in fasteners on the top edge of the cover and lift the cover upward.
Forma® Model 4535/4536 ________________________________________Service Figure 5-7 Air plenum with heating and air moving components (Relay tray removed for illustration) 5.8 Heater Element Circuit Breaker (Qualified repair personnel only) A manual reset circuit breaker is located between the heating element electrical connectors on the side of the air plenum. (Figures 5-7 and 5-8) Reset the breaker by removing the front grille and inner cover and reaching over the relay tray.
Forma® Model 4535/4536 ________________________________________Service 5.9 Tuning the Cabinet Once the console cabinet is in place and level and the platform installed, turn the unit on and set the RPM to 300. Kneeling in front of the console, lightly touch the lower left and right corners of the cabinet. If one side seems to vibrate more then the other, raise or lower the corner support leg using a 3/4”...
Page 38
Forma® Model 4535/4536 ________________________________________Service Figure 5-10 Motor/belt drive system, cover tray removed. To remove the pulleys from their shafts, use a 1/8” Allen wrench to remove two set screws from the belt groove of the larger pulley; use a 5/16” Allen wrench to remove the single set screw from the belt groove of the smaller pulley.
Forma® Model 4535/4536 ___________________________________Specifications Section 6 - Specifications Models 4535 and 4536 a. Shaking Range 25-525 RPM ±1 RPM Accuracy Motion One inch/orbital Indicator LCD in 1 RPM increments b. Temperature Range 5°C above ambient to 60°C (140°F) ±0.1°C Control Uniformity ±0.2°C (in flask) Indicator...
Page 40
Forma® Model 4535/4536 ___________________________________Specifications g. Drive Triple counterbalanced. Handles unbalanced platform loads regardless of flask placement h. Drive Motor 1/3 HP Brussels DC, permanently-lubricated ball bearing i. Lid Counterbalanced, hand or foot operated, with tempered thermal pane window and key lock. j.
Page 41
Forma® Model 4535/4536 ___________________________________Specifications n. Certification Standard No. UL1262 (4535) CSA Standard C22.2 No. 151 (4535) CSA Standard C22.2 No. 1010.1 (4536) EN61010 (4536) o. Capacity Flasks from (91) 25ml up to (4) 6L p. Weights 518 lbs. (235.2kg) q. Optional Platforms Size 29.5"...
Page 42
Forma® Model 4535/4536 ___________________________________Specifications Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the category used for instruments in installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
Forma® Model 4535/4536 ______________________________________Parts List Section 7 - Parts List Model 4535 (Refer to Table 5-2 for electrical fuse information) Part No Description 129024 Pneumatic Spring, 80 lbs. 138009 Heater, Wirewound 450W 115V/230V 156089 Motor, Brushless 24V 190430 Motor Drive Board 190700 Temp Control Board 190562...
Page 44
Forma® Model 4535/4536 ______________________________________Parts List Model 4536 Part No Description 129024 Pneumatic Spring, 80 lbs. 138009 Heater, Wirewound 450W 115V/230V 156089 Motor, Brushless 24V 190430 Motor Drive Board 190525 Relay Board 190700 Temp Control Board 190562 LCD Display Panel 190769 Shaker Micro Board 290137 Probe, 2252 Ohm/25°C, 1/8 x 2 (2)