[Route:]; Close - Agilent Technologies 75000 Series User Manual

50 ohm rf multiplexer module 75 ohm rf multiplexer module
Hide thumbs Also See for 75000 Series:
Table of Contents

Advertisement

[ROUTe:]

Subsystem Syntax

CLOSe

Parameters
Comments
Example
48
Agilent E1366A/E1367A RF Multiplexers Command Reference
The
ROUTe
subsystem controls switching and scanning operations for RF
multiplexer modules in a switchbox.
[ROUTe:]
CLOSe <
channel_list
CLOSe? <
channel_list
OPEN <
channel_list
OPEN? <
channel_list
SCAN <
channel_list
:MODE NONE | VOLT | RES | FRES
:MODE?
channel_list
[ROUTe:]CLOSe <
specified by channel_list. The channel_list has the form (@ccnn) where
cc = card number (01-99) and nn = channel number (00-03 and 10-13).
Parameter Name
<channel_list>
Closing Channels: To close:
a single channel, use
multiple channels, use
sequential channels, use
groups of sequential channels, use
[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn)
or any combination.
Closure order for multiple channels with a single command is not
guaranteed.
Related Commands:
*RST Condition: All channels open.
Closing Multiplexer Channels
This example closes channels 100 and 213 of a two-module switchbox (card
numbers 01 and 02).
CLOS (@100,213)
>
>
>
>
>
closes the RF multiplexer channels
>
Parameter Type
numeric
[ROUTe:]CLOSe (@ccnn)
[ROUTe:]CLOSe (@ccnn,ccnn)
[ROUTe:]CLOSe (@ccnn:ccnn)
,
[ROUTe:]OPEN
[ROUTe:]CLOSe?
! Close channels 100 and 213.
100 closes channel 00 of card #1;
213 closes channel 13 of card #2
Range of Values
cc00-cc03, cc10-cc13
Chapter 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1366aE1367a

Table of Contents