MT100 BLOOD REFRIGERATOR / INCUBATOR INSTRUCTIONS FOR USE...
Page 3
Transient overvoltages up to the levels of overvoltage category II; • Temporary overvoltages occurring on the mains supply; • Pollution degree 2. • The MT100 will be routine tested for grounding and dielectric strength according to Annex F of the above mentioned IEC 61010-1.
• B Medical Systems does not guarantee the safety of the stored products, if the appliance is used for any purposes other than its intended use or if any procedures other than those mentioned in this operating manual are used.
2.1.1 Warnings • Do not use the device outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. • Before cleaning or carrying out maintenance work, always switch the device off and disconnect the mains plug.
• Not activating the breaks, may cause the unit to roll away and cause damage to the unit and physical injury. • Not securing the device with the appropriate belt system, may cause damage to the unit during transit and physical injury.
2.2 Labels and Symbols on the Packaging • Follow the instructions for transport and storage of the product marked on the packaging: 2.3 Unpacking and Inspection CAUTION • Check whether the device has been delivered undamaged. If you find that damage has occurred in transit, immediately contact the delivery service or relevant sales outlet, submitting the delivery note or proof of purchase.
2.4 Intended use 2.4.1 Intended Purpose Blood refrigerator/incubators are devices intended to be used in the blood transfusion medicine for the support on the diagnosis, prevention and treatment of diseases or injuries by keeping whole blood or blood components at a protective and stable temperature between 2°C to 6°C, or platelet components between 20°C to 24°C, until they are ready for use.
Blood components intended for intrauterine and exchanged transfusion (e.g. IUT, ET, PR/ET) must be stored between 2°C and 6°C for no longer than five days prior to use. After concentration and irradiation, the storage time must be no longer than 24 hours. If transported or stored without agitation, platelet components must be kept at a controlled temperature between 20°C and 24°C for a maximum period of 24 hours before use.
3 Installation Procedure WARNING • All installation work and adjustments to the device must be carried out by qualified personnel. Work performed by persons with insufficient technical knowledge may adversely affect the performance or cause physical injury or damage to the equipment. •...
3.2 Castors • The device is equipped with 4 wheels. Two of these wheels are equipped with brakes. Once it is in its final position, the brakes of the front wheels must be locked to immobilise it and stop it from rolling away . 3.3 Fixation •...
3.6 Electrical Connection WARNING • Only connect to a grounded socket. If the power supply outlet is not grounded, it will be necessary to have a ground installed by qualified personnel. • Before connecting, check whether the details on the nameplate in the interior correspond to local values. Use of any other voltage or frequency other than that on the nameplate may cause fire or electric shock.
4 Use and Operation WARNING • The device must only be used by adults. Do not allow children to play with it or touch the controls. • Do not store flammable or explosive substances inside these appliances (e.g. aerosols with flammable propellant).
• The unit is designed as storage for blood bags and platelets. It is not designed for cooling down the stored product. • The maximum storage capacity of the MT100 is 21,5L of blood (48 blood bags of 500ml filled with 450ml each).
5 Control Panel 5.1 Overview Electronic Control Unit Power 4ßC Power Indicator Main Switch 22ßC Mute Button Temperature for Powerfail-Alarm set-point key switch 5.2 Main Switch • Switch on the unit by turning the switch to the “I”- position. 5.3 Power Indicator •...
6 Alarms CAUTION • Upon the occurrence of an alarm situation, attempts must be made to discover the reason for the alarm and resolve it as quickly as possible. If that is not successful the necessary measures must immediately be taken to avoid damaging the stored goods.
• The rechargeable battery is a lead accumulator which, if defective, must be disposed of separately. 7.2 Automatic Defrost • The MT100 is equipped with an automatic defrost function. After a run time of 8 hours, the compressor is switched off until the evaporator temperature reaches +4°C. If this temperature is not achieved after a timeout of 20 minutes, the appliance is switched on again.
7.6 Vehicle Battery Protection • In DC-operating mode, the MT100 is equipped with a voltage management system. In order to ensure that the vehicle battery is not completely unloaded and it can be started without problems, the MT100 is switched off.
8 Maintenance and Cleaning WARNING • Before cleaning or carrying out maintenance work, always switch the unit off and disconnect the mains plug. Working on the connected appliance may cause electric shock or fire. • When disconnecting, always pull the plug and never the cable. Pulling the cord may result in electric shock or fire due to a short circuit.
9 Disposal CAUTION • Before scrapping an old unit, remove the door so that children cannot lock themselves inside while playing. • Before scrapping the unit, remove the lead accumulator and dispose of it seperately. • When disposing of the unit, make sure that it does not get too hot, as combustible gas would cause the insulating foam to froth up.
11 Technical Data Model MT100 Gross volume [L] Net volume [L] 21.6 External dimensions (HxWxD) [mm] 520x800x1000 Supply voltage [V ] AC 100-130 / 220-240 - DC 13.7 / 27.4 Frequency [Hz] 50/60 Fuse [A] Set point [°C] 4 / +22 Preset cold alarm [°C]...