3.14. Phase Sequence Indication (for PF > 0.1)
Fig.17
37

3.16. Reading Data in Memory

To retrieve data in memory, press the READ button as long
as rotary switch is not set at 3ffunction.
Once the read button is pressed, the record number will be
first displayed. then data is displayed after. Once in the READ
function, the REC. NO. symbols will be displayed in LCD.
To exit the READ function, turn the rotary switch to change
function.
NOTE:
If users observe both REC and NO. symbols are shown in LCD,
users should know that the readings shown in LCD are not cur-
rent data measured. They are data stored in the memory.

4. Specifications (23

C ± 5
C)
0
0
AC ± DC True Power ( PF 0.2 - 1.0, 3f3W, 3f4W, 1f2W and
1f3W) :
Range
Resolution Accuracy (of rdg)
0 -
0.01KW
±2.0%±0.05KW AC 600V, DC 800V,
99.99KW
100 -
0.1KW
±2.0%±0.5KW
999.9KW
1000 -
1KW
±2.0%±5KW
1200KW
KW Autoranging Map (PF 0.2 - 1.0, 3f3W, 3f4W, 1f2W
and 1f3W)
0V - 200V
0A - 200A
0.00 - 40.00KW
200A - 2000A
0.0 - 400.0KW
39
Range
ACA/DCA 2000A
AC 600V, DC 800V,
ACA/DCA 2000A
AC 600V DC 800V,
ACA/DCA 2000A
200V - 600VAC, 200V - 800VDC
0.00 - 99.99KW
100.0 - 160.0KW
0.0 - 999.9KW
1000 - 1600KW
NOTE :
The indication of phase sequence is not limited to 3f balanced
power. It can be used in general as long as the PF displayed is
greater than 0.1.
Set the power clamp In 3f balanced power mode, the power
clamp will automatically detect phase.
3.14.1. Set the rotary at 3f.
3.14.2. Connect the voltage input (red terminal) to R (or L1)
phase, connect the COM input (black terminal) to S
(or L2) phase, and clamp on the T (or L3) phase. When
clamp on the T (or L3) phase, the power clamp must
face the power source.
3.14.3. If the phase sequence is clockwise, R, S, T sequence
will be shown and flashed. (refer to figure 17 )
3.14.4. If the phase sequence is counterdockwise, T, S,
R sequence will be shown and flashed. (refer to
figure 17)
3.14.5. If the current signal (where jaws clamp on) is not
detected, only one of the R, S or T will be left in LCD.
NOTE :
The "+" sign printed on jaw must face the power source for correct
measurement.

3.15. Recording Data in Volatile Memory

The power clamp can store 4 data in its own memory. To
store any data displayed in LCD, press the REC button. If
the power clamp has stored 4 data already, LCD will display
FULL. The data is lost if power is turned off.
The REC symbol will be displayed in LCD if any data is
stored.
NOTE :
If users observe REC (only) is shown in LCD, that means some
data is stored in memory.
AC+DC True Power ( Power Factor 0.2 - 1.0, 3f Balanced
Power) :
Range
Resolution Accuracy (of rdg)
0 -
0.01KW
99.99KW
100 -
0.1KW
999.9KW
1000 -
1KW
2000KW
Power Factor (PF)
PF =
AC+DC Voltage ( True RMS, Crest Factor < 4, Autorange,
Overload Protection 800VAC for all range)
Accuracy (of reading)
Range
Res.
DC, 50/60Hz
0-200V
0.1V
±1.5%±5dgts
200-500V
0.1V
±1.5%±5dgts
500-600V
1V
±1.5%±5dgts
AC+DC Current (True RMS, Crest Factor < 4) :
Accuracy (of reading)
Range
Res.
DC, 50/60Hz
0 - 200A
0.1A
±1.5%±5dgts
200 - 500A 0.1A
±2.0%±5dgts
500-2000A
1A
±2.5%±5dgts
AC+DC KVA (Apparent Power)
(KVA) =
AC+DC KVAR (Reactive Power)
(KVAR) =
4500 : 28-09-2012
Range
±2.0%±0.5KW
AC 600V, DC 800V.
ACA/DCA 2000A
±2.0%±0.5KW
AC 600V, DC 800V,
ACA/DCA 2000A
±2.0%±5KW
AC 600V, DC 800V,
ACA/DCA 2000A
KW
KVA
Input
40-400Hz
Impedance
±2.0%±5dgts
10MV
±2.0%±5dgts
10MV
±2.0%±5dgts
10MV
Overload
40-400Hz
Protection
±2.0%±5dgts AC 3000A
±2.5%±5dgts AC 3000A
±3.0%±5dgts AC 3000A
V*A
1000
(KVA)
2
- (KW)
2
38