Only operate the C-Quant using original accessory parts supplied by us, and when
the device is in technically correct working order.
Never operate a C-Quant that is damaged. Should such a situation arise, contact the
supplier.
Do not operate the devices included in the delivery in areas where explosions may
occur, where there are inflammable anaesthetics, volatile substances such as alcohol
or petrol nearby.
Do not operate or store the C-Quant in damp rooms.
Keep the C-Quant away from water that may drip, surge or splash and make sure
that no liquids can enter the C-Quant. Do not place any containers with liquid either
close to or on the C-Quant.
When cleaning, use a damp cloth and make sure that no liquid enters the C-Quant.
Only operate the C-Quant in rooms used for medical purposes if VDE 0100 installa-
tion procedures have been observed.
Do not force any plug connections.
If you are unable to make a plug connection, check whether the plug fits the socket.
If you detect damage to the connection, you should let our service personnel repair
the damage.
When disconnecting electrical connections, pull on the respective plug instead of
the cable itself.
Attention
Any additional equipment that is to be connected to either analogue or digital interfaces
must verifiably fulfil appropriate EN or IC specifications. Furthermore, all configurations
must comply with the IEC 601 - 1 system norm.
Attention
If you connect the C-Quant with non-medical equipment (e.g. data processing equip-
ment) to a medical electrical system, doing so must not cause patient-relevant safety
levels to fall below IEC 601 - 1. If making this connection leads to the leakage current
threshold being exceeded, protective measures including a circuit breaker must be in
place.
Instruction Manual C-Quant ( G/80000/1007/en)