Introduction Introduction Priorclave autoclaves are a range of general-purpose laboratory autoclaves intended primarily for media preparation, the making safe of ordinary laboratory and pathological waste and other apparatus sterilisation purposes. The autoclaves are manufactured to a high standard and feature a number of patented innovative design features.
Contents Contents Introduction ..................................1 Notices and Important Information..........................5 Symbols used on the product and in this manual and their meanings ............... 5 General Safety Instructions ............................6 Important Notices and Warnings ..........................7 EC Declaration of conformity ............................. 10 Layout Diagrams ................................
Page 7
Contents Vacuum Options ................................ 34 Multi Program Memory Options ..........................35 Closing the pressure door ............................36 Starting a cycle ................................36 Vent button................................37 During the process time............................37 Power loss during a running cycle..........................38 Cooling ..................................
Notices and Important Information Notices and Important Information Symbols used on the product and in this manual and their meanings WARNING: Mechanical Hazard In this manual, warnings draw attention to the potential for Danger to personnel up to and including risk of severe injury or death.
Notices and Important Information Refer to Manual General Safety Instructions All cleaning and servicing requires the autoclave to be isolated from the power source and disconnected. This equipment weighs in excess of 18kg (40Ibs) and requires at least 2 persons to lift. ...
Always wear gloves, a facemask and adequate protective clothing when unloading an autoclave and ensure that the workload does not exceed safe limits. Priorclave are pleased to arrange training for operators in the use of their autoclaves at a small extra charge.
Page 12
Priorclave Ltd. cannot be held responsible for hazards or damage resulting from work carried out on the pressure system by untrained or unauthorised personnel. If in doubt please contact Priorclave Service or your nearest authorised dealership.
Page 13
Notices and Important Information Priorclave Ltd. 129 /131 Nathan Way West Thamesmead Business Park London SE28 0AB Environmental Conditions This equipment has been designed for safe operation within the following environmental conditions: Indoor Use. Altitude up to 2,000 M (~ 6500 feet).
Equipment Description Priorclave Laboratory autoclaves – RSC series, with Tactrol microprocessor control system. Model Numbers PS/RSC/EH450 and PS/RSC/EH700. For direct steam heated variants substitute S for E. All models may be suffixed DBL to denote double entry versions. Conformity Data...
Before proceeding please check the specification sheet at the front of this manual to establish which options and accessories, if any, are fitted to your Priorclave. This will determine whether you will need to read the instructions for these options later in this manual.
Page 18
Operating Summary Wait during the ‘Hold’ display until there is a bleep and the ‘Door’ indicator illuminates. Keep the key held and press the ‘Door’ button once to unlock the door. The key-switch can now be released and the door opened as above. If the key is released at any stage the procedure must be repeated to open the door and reset the display.
This equipment weighs in excess of 18kg (40Ibs) and requires at least 2 persons to lift. Reference should be made to your in-house procedures for heavy lifting before attempting to move unpack or lift Any shortages or damage must be reported to Priorclave within 7 days of delivery. Positioning When positioning the autoclave consideration should be given to proper access for servicing and maintenance purposes.
Positioning A space to the rear of the autoclave of up to 300mm can be required to accommodate pipe-work connections. This will reduce for more simple installations such as autoclaves without vacuum systems. For autoclaves with exhaust filters a space of at least 500mm should be allowed at the rear of the autoclave.
Page 21
Positioning mounted units if there is not sufficient space available in the autoclave location and in such cases consideration should be given to where the condenser is to be located as it may not be practical to site this to the rear of the autoclave. Drain condensers may operate at high temperature depending on final settings, and effectiveness of the cooling water supply.
Installation Drainage and Exhaust Gas Ventilation General Autoclaves used for processing laboratory waste must be provided with a drainage connection as described below. This is a requirement of British Standard 2646. A connection will also be required if the autoclave is fitted with any freesteaming or vacuum options. The hazard groups below are as defined by the Advisory Committee on Dangerous pathogens as published in Categorisation of Pathogens According to Hazard and Categories of Containment.
Page 24
The drain and vent pipe should be in place prior to commencement of installation by Priorclave. It will then be possible to make connections from the autoclave directly into the drainage services provided.
Page 25
It is Priorclave’s preferred policy to direct the safety valve outlet to discharge to the floor at the rear of the autoclave. Some establishments prefer safety valve outlets to be piped to a high level outside the building.
Page 26
Installation Autoclaves with water tanks and drip trays A separate drain is required for the connection of water tank and drip tray outlets. This may be a common connection to the same drainage system as above, but in such cases a sufficient level of isolation (such as a deep trap) is required to prevent the cross flow of steam between the drains.
The diagram overleaf shows a typical, but by no means definitive example of a reducing valve and trap set arrangement, which could be used to serve a single autoclave. Priorclave can provide assistance in planning a steam arrangement for a whole suite of autoclaves if required.
Page 28
Installation In the case of removal of an existing autoclave consideration should be given to which parts of the steam supply are integral to the old autoclave. It is possible that some parts may need to be retained, and re-mounted to suit the new autoclave. H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Page 29
Installation General All water supplies should be terminated on the wall to the rear of the autoclave, at a point within one metre of the point of connection to the autoclave. In hard water areas it will be necessary to use softened water for the supplies to the autoclave to prevent scaling of the autoclave vessel and heating elements.
Full Commissioning and Performance Qualification If you are having the unit commissioned by a Priorclave technician this will be a simple matter of checking for correct installation, checking that all functions are operating correctly, and familiarising you with the autoclave. There are, however some benefits that can be gained from having your Priorclave commissioned to suit your particular loads and requirements.
Page 31
Setting thermal lock release temperature to suit your particular load, to eliminate unnecessary cooling time whilst ensuring safety. If you feel that any, or all of the above would be of use to you then please contact Priorclave Service. H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Operation Operation Before using your Priorclave for the first time check that the circuit breakers and the isolator (at the back of the autoclave) are switched on and that the water supply (if required) is available and turned For Steam Heated models, check that the steam supply (or Internal steam generator if fitted) is ON and that there is sufficient pressure (3 to 3.5 Bar) to run a cycle.
Operation Release the locking castors before pulling the trolley away from the front of the autoclave. Great care should be taken when loading and unloading bulk loads with an autoclave trolley. Manual handling regulations should be observed at all times. Loading Trolleys and Cassette (Optional) Standard Loading Trolley The Trolley has been manufactured for loading baskets or containers onto the lower shelf of the...
Tests have shown that the depth of un-perforated discard containers should be no greater than 180mm (7”) for effective air displacement from the load. Suitable containers are available from Priorclave. Settings. Once the autoclave has been satisfactorily loaded the controls should be set for the process cycle that you require.
Page 35
Automatic Free-steaming function will perform this function automatically if the option is selected. If you have an interest in any of the options mentioned above, which can quite easily be retro-fitted, please contact Priorclave. In conclusion, when setting up the autoclaving cycle a large safety margin should be allowed within the settings.
Operation models) pressing the appropriate key will result in a visual and audible fault being signalled and the function will not be selected. Automatic Timed Free-Steaming What is free-steaming? Incorporating a period of free-steaming into a cycle can improve air removal in difficult loads and/or reduce temperature lag between the load and the autoclave, reducing process time at higher temperatures.
Load Sense Temperature Probes The load sensed process timing option utilises a temperature probe connected directly to the main processor board. Replacement temperature probes are available from Priorclave. See Maintenance - for details on temperature probe replacement. Rapid Cooling On most models a fan is fitted into the bodywork of the autoclave to direct cool air over the autoclave vessel, however on some models cooling is effected by cooling coils, a full or partial cold water jacket or a combination of both systems.
It is strongly recommended that to achieve optimum performance from Priorclaves fitted with vacuum options that commissioning and/or load validation tests are carried out by a trained Priorclave engineer. If no particular programs have been specified your autoclave will be factory set with the...
The indicators on the left are for programs 1 to 5 and those on the right for programs 6 to 10. If the Priorclave has been specified with a five program memory only the first five programs will be active.
Operation When the program memory option is fitted a three-position setting lock keyswitch is fitted. These setting positions allow different levels of access to settings as follows. Position 1. Only the currently selected program can be run. Program settings cannot be changed. Position 2.
Operation Once a cycle has been started the function selection settings cannot be changed; attempting to do so will cause a fault to be signalled. If changes are required the cycle should be aborted by pressing the start button again. Pressing the Free-steam Button during the freesteam part of the cycle will give an indication of free steam time remaining as a flashing display in the timer window.
Operation Power loss during a running cycle. Loss of power to the autoclave whilst the cycle is running can have one of two outcomes, depending on the stage of the cycle at which the power loss occurs: If the power loss occurs at any point before completion of the sterilizing stage of the cycle (process time) then the error code F006 will be shown in the temperature indicator and the cycle will be stopped.
Operation close procedure using the thermal lock override key. Operation will return to normal as soon as the next cycle is completed satisfactorily Cycle complete When cooling to the ‘thermal lock deactivation temperature’ is complete, the complete indicator will illuminate, and the autoclave will emit a bleep for a short time (about 10 seconds). If the Cooling System is selected it will automatically switch off at this point.
Operation with Options & Accessories The following descriptions detail how to operate and gain maximum benefit from the options and accessories that may be fitted to your Priorclave. Setting Lock Keyswitch Option Fitted on Priorclaves without program memory to give an optional level of security this keyswitch has two settings only, which are equivalent to positions 1 &...
Page 45
The paper roll carrier is mounted on the rear of the printer front panel and will swing out from the main body for ease of roll changing. Always use the correct paper. Suitable paper is available from Priorclave Service or your local distributor.
Unless specified otherwise before delivery of the autoclave or set up during the commissioning process the Media cooling function is set to operate by default with program 1. It can easily be re-set to operate on different programs by a Priorclave Engineer or your local approved Priorclave representative.
Operation with Options & Accessories Delayed start cooling (both lights lit on the cooling button) may be required if excessive media volume losses during cooling are observed. Chart Recorder / Data Recorder / Data Logger If fitted, the recorder power input is connected directly to the autoclave in such a way that the recorder will only operate during the autoclave cycle, i.e.
Page 48
Automatic Leak Test program If ordered the leak test program will be set up by a Priorclave trained technician and set up as one of the programs in the program memory. When this program is selected and run a leak test will be performed instead of a normal cycle.
Operation with Options & Accessories Bowie Dick test This is set up during commissioning by a Priorclave trained technician. It will normally be set in connection with a drying cycle. All programs set to be Bowie Dick test programs will be identical to normal cycles with the exception that at cycle complete “Bowie Dick test Complete”...
Changing Date & Time Changing Date & Time A number of additional control system settings can be accessed via a “Hidden Menu”. To access these settings turn & hold the thermal lock key in the override position. Press the time up or down keys.
F000 appears it may be cleared by the method described below. If the fault code will not clear, or continues to re-appear then the user cannot correct the fault. In such a case please contact Priorclave service or your local Priorclave approved service agent.
Page 52
Warning Indicators & Fault Codes F010 Air detector input activated. If fitted the air detector system has detected an over pressure condition symptomatic of excess air remaining in the load. F011 Printer Timeout / Malfunction. The control system has not received confirmation from the printer within its pre-set timeout.
Page 53
Warning Indicators & Fault Codes This page is deliberately blank H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Fault Finding & Rectification Guide Fault Finding & Rectification Guide Symptom Possible Cause Possible Solution No Power Power switched off at isolator Check Circuit Breaker Tripped Call Engineer Electrical Failure Call Engineer Emergency Stop Button in (if fitted) Release Emergency Stop Cycle does not commence when start Door is not closed properly Open &...
Page 55
If you have any doubts or If you do not feel competent to carry out any of the above procedures then do not hesitate to call Priorclave Service on 020 8316 6620 or your nearest Priorclave approved service agent.
Steam Table Steam Table Steam Generators Autoclaves (where fitted) Temperature Pressure (Bar) Temperature Pressure (Bar) 0.00 0.20 0.43 0.69 147.5 0.99 1.06 1.13 153.5 1.25 1.35 1.55 1.70 1.86 2.04 2.21 2.40 2.60 Correct Correlation between Temperature and Pressure shows correct operation of the autoclave and that air purging is satisfactory.
Ensure that any electrically locked doors are open before disconnecting power. In the event of any difficulty or doubt about any maintenance or service procedure contact Priorclave Limited or your nearest Priorclave approved agent or supplier immediately.
Routine Operator Maintenance Door Closure Bolts To ensure a free action of the closure bolts they should be kept lubricated and free from dirt. The threads of the bolts and the threaded holes in the header ring should be thoroughly cleaned and then lubricated with high melting point grease.
Maintenance Log. The pressure gauge and control system fitted to Priorclave are extremely reliable instruments and as such are unlikely to produce false readings. Therefore it is more likely that any deviation from the values given in the steam table is caused by incorrect air purging etc.
Equipment Specification Tables Equipment Specification Tables Priorclave Large Capacity Rectangular Section Laboratory Autoclaves Where the vacuum model is not listed separately the weight of the vacuum models is shown. For steam models substitute E with S Intended use. Principle uses: Preparation of Laboratory growth media and sterilization of laboratory waste prior to disposal.
Other Items Fitted Circuit Diagrams and Pipe-Work Schematics Output Board Designations Output Board 1 FUNCTION Output Board 2 FUNCTION Output Board 3 FUNCTION H:\Technical\MANUALS\Manual System\28138723-EN.docx...
The following Circuit Diagrams show the wiring configuration for the standard models and in addition for the more popular options and accessories. For Options and Accessories not covered by these standard drawings supplementary drawing sheets will be issued and these should be included within the manual and/or are available from Priorclave. (info@priorclave.co.uk) H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Page 67
Other Items Fitted H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Page 68
Other Items Fitted H:\Technical\MANUALS\Manual System\28138723-EN.docx...
The following Circuit Diagrams show the wiring configuration for the standard models and in addition for the more popular options and accessories. For Options and Accessories not covered by these standard drawings supplementary drawing sheets will be issued and these should be included within the manual and/or are available from Priorclave. (info@priorclave.co.uk) H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Page 70
Other Items Fitted H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Other Items Fitted Other Items Fitted H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Page 72
Other Items Fitted H:\Technical\MANUALS\Manual System\28138723-EN.docx...
Page 73
Specification Sheet Attention: Priorclave Service 129-131 Nathan Way Please read the manual before attempting to operate your Priorclave. West Thamesmead Business Park Connection to the wrong power supply could lead to an expensive and London SE28 0AB unnecessary call out.
Page 75
Please complete the details below and return by fax or email within 14 Priorclave Service days of delivery of your Priorclave Laboratory Autoclave. You will then be registered as a Priorclave user and your 12month* Parts 129-131 Nathan Way West Thamesmead Business Park and labour warranty will commence from the date entered below.