User Manual
OPERATING TABLE FENIX
OT-01
Serial numer ...................................
Edition 3.0
June 2022

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FENIX OT-01 and is the answer not in the manual?

Questions and answers

Summary of Contents for INFIMED FENIX OT-01

  • Page 1 User Manual OPERATING TABLE FENIX OT-01 Serial numer …………………………….. Edition 3.0 June 2022...
  • Page 2 Manufacturer: INFIMED Ltd. Kabaty 1 Street 34-300 Żywiec Poland WWW.INFIMED.PL Tel/fax +48 33 861 40 96 SRN: PL-MF-000015633 NIP 553 251 29 67 REGON 243274947 In accordance with the classification of medical devices according to Regulation (EU) 2017/745 of the European Parliament and of the Council, the lamp has been classified as class I.
  • Page 3: Table Of Contents

    Table of contents Safety ............................... 5 1.1. General safety remarks ........................ 5 1.2. General notes on the safe use of the product ................5 1.3. Technical parameters ........................6 1.4. General requirements ........................7 1.5. Product description ........................8 1.6. Description of the components of an example table ..............8 2.
  • Page 4 12. Liquidation of the product ......................17 13. Additional accessories ......................18 14. Table labels ..........................21...
  • Page 5: Safety

    1. Safety The product has been designed and manufactured to ensure its safe use and operation. A prerequisite for the safe use of the table is to read, understand and follow the rules included in this manual. 1.1. General safety remarks - Do not use, operate or service the table in a manner inconsistent with this manual.
  • Page 6: Technical Parameters

    - Do not use agents containing alcohol to wash and disinfect mattresses. - If the product will not be used for an extended period of time, it should be stored under the following environmental conditions: - temperature: 25⁰C (77⁰F) ± 10⁰C (18⁰F) - humidity: 50% ±...
  • Page 7: General Requirements

    Table mass 250 kg Maximum working load 250 kg Period of use 10 years At the special request of the customer, it is possible to manufacture a product with changed technical parameters, without reducing its safety. 1.4. General requirements Using, operating and servicing the table inconsistently with the guidelines of the manual is absolutely forbidden and may lead to danger and irreparable damages due to the user's fault, for which the manufacturer is not responsible.
  • Page 8: Product Description

    1.5. Product description The operating table FENIX OT-01 is made of stainless, acid-resistant steel and is designed to support the patient during all surgical and specialised procedures and operations. The operating table FENIX OT-01 is equipped with a multi-segment table top mounted on a mobile base with central wheel locking.
  • Page 9: Transport And Start-Up

    Side rail Footrest segment Footrest angle adjustment lever Side rail Optional kidney bridge Adjustment for side recline Kidney bridge adjustment Footrest attachment to seat Lever attachment socket 2. Transport and start-up 2.1. Transport For the time of transport of the table by means of transport the table must be placed in the transport position.
  • Page 10: Unpacking And Initial Start-Up

    8-10°C per hour. Products must not be stacked in layers if not clearly marked on the transport packaging. 2.2. Unpacking and initial start-up The table is supplied by the manufacturer in a box with a total weight of approximately 250 kg. The table must not be unpacked outside the building.
  • Page 11: Operation And Performance

    3. Operation and performance 3.1. Raising and lowering of the tabletop The height of the tabletop is variable within the range specified in the product parameters. To set the worktop to the desired height, select the appropriate function of the drive selector "5i" and then press the appropriate foot pump lever "5c"...
  • Page 12: Tabletop Function Adjustment

    appropriate lever of the foot pump "5c" (one tilts the tabletop into the Trendelenburg or right position and the other into the Antitrendelenburg or left position) until the desired position is reached. When the end positions are reached, the tabletop will stop. When carrying out the Trendelenburg and reverse-Trendelenburg positions as well as lateral tilting of the tabletop, one should always remember to secure the patient to prevent him/her from uncontrolled sliding off the operating table.
  • Page 13: Footrest Rotation Adjustment

    Angle adjustment of the footrests is carried out within the range specified in the technical parameters of the table. The change of the angle should be carried out separately for each footrest. To do this, stand in front of the footrest, grasp the footrest with one hand and the lever underneath with the other.
  • Page 14: Assembly And Disassembly Of Selected Components

    It is absolutely necessary to ensure that the locking mechanism works and that the worktop cannot be moved despite the release of the handles. 3.4. Assembly and disassembly of selected components 3.4.1. Assembly and disassembly of the headrest The headrest is mounted to the backrest segment. Special headrest mounting clips are used for this purpose.
  • Page 15: Assembly And Disassembly Of Mattresses

    Ensure that the footrests are properly secured before use. 3.4.3. Assembly and disassembly of mattresses The mattresses are disassembled without the use of tools. Both in the case of Velcro and pin fastening, their removal consists of applying the appropriate muscle force and detaching the mattress from the base of the respective segment.
  • Page 16: Defect And Faults

    b) Check the operation of the mechanisms controlled by the pneumatic springs and the corresponding retarders. Once the position has been properly adjusted (see sections Operation and functioning), it should not be possible to change the position of the individual table segments without pressing the appropriate retarders and applying sufficient force.
  • Page 17: Condition Checks And Inspections

    Any repairs to the product shall be carried out by the relevant service centre or the manufacturer's direct representative. The user is not entitled to make any modifications or repairs to the product himself without special training and authorisation. After the customer has obtained the manufacturer's written consent for a possible repair to be carried out by the customer's technical personnel, the manufacturer will make available all the necessary information needed to carry out the repair.
  • Page 18 13. Additional accessories - One position clamp EB-01 EB-02 - Multi position clamp EB-03 - Knee support EB-04 - Angle arm support - Arm support EB-05 - thigh holder EB-06 EB-07 - Side-shoulder support EB-08 - Side support EB-09 - Anesthetic frame EB-10 - Anesthetic frame with adjustable width - Infusion holder...
  • Page 19 - X-ray tray ES-11 - Chest support ES-12 - Attachment for knee operations ES-13 - Pubic support ES-14 - Leg support, not-divided ES-15 - Proctological attachment ES-16 - Side rail extension ES-17 - Tabletop extension ES-18 - Strap for anesthetic frame ES-19 - X-Ray translucent plate /mounted instead of leg rests/ ES-20...
  • Page 20 - Trolley for installation of orthopedic attachment EO-12 - Mobile stand for additional equipment of orthopedic attachment EO-13 - Leather shoes for orthopedic attachment EO-14 - Small leather shoe EO-15 - Attachment for hand operating EO-20 - Support roller for hand surgery EO-21 - Attachment for shoulder operations EO-30...
  • Page 21 14. Table labels The following labels can be placed on the housing and tabletop structure: Read the user's manual Transport position Headrest angle adjustment Backrest angle adjustment Foot rest angle adjustment Longitudinal slide...
  • Page 22 Rotation of footrests Disinfection Brake lever Rating label Potential equalization socket...
  • Page 23 Code UDI-DI-PI...
  • Page 24 Annex no. 1 Agents recommended for cleaning and disinfection of surfaces of products and elements made of stainless steel and antistatic polyurethane. ANTISTATIC AGENT TAINLESS STEEL ISTRIBUTOR RODUCER POLYURETHANE Aesculap-Chifa Sp.z o.o. ELISEPTOL st. Tysiąclecia 14 64-300 Nowy Tomyśl phone: 061 4420100 fax: 061 4437505 Bochemie PL Sp.
  • Page 25 Annex no. 2 Card of repairs and inspections of product Operating table type ......Serial no …...... Purchase date ......…… Inspection Inspection type Sign of person Inspection Person performing Comments concerning inspection or or repair (annual, six- performing inspection or repair repair date month)