2.
SPECIFICATION
Range
Resolution
Accuracy
Stability
Stabilisation Time
Temperature Readout
Well Diameter
Immersion Depth
Heating Time
(Ambient to max.)
Cooling Time (Max. to 100°C)
Switch Test facility
Interchangeable
Thermowells
Housing / Mounting
Power Supply
Instrument Dimensions
(L x D x H)
Instrument Weight
Measurement Capability
NAG/IM/MTC350/01/22
50°C to 350°C
0.1°C
±0.4°C
±0.05°C
20 Minutes
°C / °F Selectable
25 mm
110 mm
20 Minutes approx.
33 Minutes approx.
Provided
Single hole to suit 1/4" & 1/2"
Probes
Bench-top
230V AC, 50 Hz
160 x 365 x 350 mm
8 Kg. (approx.)
ISA Types
Thermocouples
J, K, R, S, N, B & T
RTD
Pt100 IEC Std.
mA
0 to 25 mA
mV
0 to 100 / 1000 mV
Volt
0 to 10 V
Resistance
0 to 500 / 3.5
KOhm
6