– Eliminate or reduce EMI with technical solutions (such as shielding).
– Restrict use of personal communicators (cell phones, computers) in areas with devices susceptible to EMI.
– Share relevant EMI information with others, particularly when evaluating new equipment purchases which may
generate EMI.
– Purchase medical devices that comply with IEC 60601-1-2 EMC Standards.
This Equipment is designed to comply with IEC 60601-1-2. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with instructions, may cause harmful interference to other devices in the
vicinity; however, there is no guarantee that interference will not occur in a particular installation. Harmful interference to
other devices can be determined by turning this equipment ON and OFF. Try to correct the interference using one or more
of the following:
– Reorient or relocate the receiving device.
– Increase the separation between the equipment.
– Connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected and
consult the BD service technician for help.
– Consult BD for help.
Symbol Glossary
Consult
Temperature
instructions
Limitation
for use
Serial
IP33 – protection from solid
Number
objects and falling water
Caution: Federal law (USA) restricts this
device to sale by or on the order of a licensed
healthcare practitioner.
Waste of electrical and electronic equipment must not be disposed as unsorted
municipal waste. It must be collected separately, and must be disposed as
per local regulations. Contact BD authorized representative for information
concerning the decommissioning of your equipment.
Electromagnetic Compatibility (EMC) Tables For RF Emissions
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The BD™ Intraosseous Powered Driver is intended for use in the electromagnetic environment specified below.
The user of the BD™ Intraosseous Powered Driver should assure that it is used in such an environment.
Emissions Tests
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations IEC 61000-3-3
Complies
5 (continued)
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Non-life supporting ME Equipment
The BD™ Intraosseous Powered Driver is intended for use in the electromagnetic environment specified below. The customer
or the user of the BD™ Intraosseous Powered Driver should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Compliance Level
Level
Conducted RF
AC Mains
AC Mains
IEC 61000-4-6
3V
3V
80% AM at 1 kHz or
80% AM at 1 kHz
risk frequency
or risk frequency
150 kHz – 80 MHz
150 kHz – 80 MHz
DC & I/O & Patient
DC & I/O &
Coupled (>3m)
Patient Coupled
(>3m)
3V with 6V ISM,
Home: 6V Amateur
3V with 6V
radio
ISM, Home: 6V
Amateur radio
80% AM at 1 kHz or
risk frequency
80% AM at 1 kHz
or risk frequency
150 kHz – 80 MHz
150 kHz – 80 MHz
Radiated RF
10V/m,
(E1) = 10V/m
IEC 61000-4-3
80% AM at 1 kHz 80
MHz – 2700 MHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the "Product Name" is used exceeds the applicable RF compliance
level above, the "Product Name" should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the BD™ Intraosseous Powered Driver.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
7 (continued)
Humidity
Atmospheric
Limitation
Pressure
Limitation
Direct Current
Type BF
5V - 1 A
Applied Part
Manufacturer
Date of
manufacture
Electromagnetic Environment - Guidance
The BD™ Intraosseous Powered Driver uses RF energy only for
its internal function; therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
The BD™ Intraosseous Powered Driver is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the BD™
Intraosseous Powered Driver, including cables, than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
3,5
— —
d = [
]
√ P
V 1
3,5
— —
d = [
]
√ P
80 MHz to 800 MHz
E 1
7
— —
d = [
]
√ P
800 MHz to 800 MHz
E 1
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
should be less than
a
the compliance level in each frequency range.
b
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The BD™ Intraosseous Powered Driver is intended for use in the electromagnetic environment specified below. The customer
or the user of the BD™ Intraosseous Powered Driver should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic
± 8 kV for Contact Discharge
± 8 kV for Contact Discharge
discharge
(ESD) IEC
±2 kV, ±4 kV , ±8 kV and
±2 kV, ±4 kV, ±8 kV and
61000-4-2
±15kV for Air Discharge
±15kV for Air Discharge
Electrical fast
±2kV for power supply lines
±2kV for power supply lines
transient/burst
±1kV for input/output lines
±1kV for input/output lines
IEC 61000-4-4
Surge IEC
AC Mains Line to Ground
AC Mains Line to Ground
61000-4-5
±0.5kV, ±1kV and ±2kV
±0.5kV, ±1kV and ±2kV
AC Mains Line to Line ±0.5kV
AC Mains Line to Line ±0.5kV
and ±1kV
and ±1kV
DC Input (>3m) Line to Line
DC Input (>3m) Line to Line
±0.5kV and ±1kV
±0.5kV and ±1kV
DC Input (>3m) Line to
DC Input (>3m) Line to
Ground ±0.5kV, ±1kV and
Ground ±0.5kV, ±1kV and
±2kV
±2kV
I/O, Line to Ground ±2kV
I/O, Line to Ground ±2kV
(Outdoor Line only)
(Outdoor Line only)
Voltage
<5% UT (100% dip in UT) for
<5% UT (100% dip in UT) for
dips, short
0.5 cycles
0.5 cycles
interruptions
0% UT (100% dip in UT)
0% UT (100% dip in UT)
and voltage
for 1 cycle
for 1 cycle
variations on
70% UT (30% dip in UT) for
70% UT (30% dip in UT) for
power supply
25/30 cycles
25/30 cycles
input lines IEC
61000-4-11
<5% UT (100% dip in UT) for
<5% UT (100% dip in UT) for
250/300 cycles
250/300 cycles
Power
30 A/m
30 A/m
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications equipment and the
BD™ Intraosseous Powered Driver Non-life supporting ME Equipment
The BD™ Intraosseous Powered Driver is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the BD™ Intraosseous Powered Driver can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the BD™ Intraosseous Powered Driver as recommended below, according to the maximum
output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated
maximum
150 kHz to 80 MHz
80 MHz to 800 MHz
output power
3,5
— —
of transmitter
d = [
d = [
]
√ P
V 1
(W)
0.01
0.12
0.12
0.1
0.38
0.38
1
1.2
1.2
10
3.8
3.8
100
12
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Warranty
BD warrants to the original purchaser that this product will be free from defects in material and workmanship for a period of two (2)
years from the date of purchase. IF this product proves to be so defective, purchaser may return same to BD for repair, replacement,
refund, or credit at BD's option. All returns must be authorized in advance in accordance with BD's Returned Goods Policy found in
its then current Price List. The warranty on the repaired or replaced unit continues from the purchase date of the original unit. The
liability of BD under this limited warrantly does not extend to any abuse, misuse, modification, improper storage, alteration, further
manufacture, packaging or processing of the product or repair by anyone other than a BD representative.
The following will also void this limited warranty:
Opening or servicing any component of the BD™ Intraosseous Power Driver by anyone other than BD authorized service personnel.
Removing system labels by anyone other than service personnel authorized by BD.
Connecting the BD™ Intraosseous Power Driver to any AC adapter other than the system adapter.
This limited product warranty is in lieu of all other warranties, whether express or implied, (including, but not
limited to, any implied warranty of merchantability or fitness for a particular purpose). The liability and remedy
stated in this limited product warranty will be the sole liability of BD and remedy available to purchaser for this
product, whether in contract, tort (including negligence) or otherwise, and BD will not be liable to purchasers for
any indirect, special, incidental, or consequential damages arising out of its handling or use.
Some states/countries do not allow an exclusion of implied warranties, incidental or consequential damages. You may be
entitled to additional remedies under the laws of your state/country.
Manufacturer: Bard Access Systems, Inc., 605 North 5600 West, Salt Lake City, UT 84116 U.S. 801.522.5000
BD and the BD Logo are trademarks of Becton, Dickinson and Company or its affiliates. ©2022 BD. All rights reserved.
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the BD™ Intraosseous Powered
Driver requires continued operation
during power mains interruptions,
it is recommended that the BD™
Intraosseous Powered Driver be
powered from an uninterrupted power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
6 (continued)
800 MHz to 2,7 GHz
3,5
7
— —
— —
]
d = [
]
√ P
√ P
E 1
E 1
0.23
0.73
2.3
7.3
23
0754809 2022-02