CONTENTS 5.4.9 Auto Power-Off Setting ............. 15 1.Principle of Leeb Hardness Testing Method........1 5.4.10 Delete All Files..............15 ..........2 Instrument and Impact Device Diagram 5.5 Measuring..................15 2.1 Instrument Diagram ..............2 5.6 Average Value..................16 2.2 D Type Impact Device Diagram............3 5.7Delete the Gross Error Value Manually 2.3 Types of Impact Device..............
1.Principle of Leeb Hardness Testing Method 2. Instrument and Impact Device Diagram The impact body, which is equipped with tungsten carbide, impacts into the 2.1 Instrument Diagram work piece and rebounds back. The rebound and impact velocities are measured at the 1mm point from the work piece in the following way: the integrated permanent magnet will produce directly proportional voltage with the impact velocity.
Average Surface 2.5 Standard Configuration 1.6μm 1.6μm 0.4μm 6.3μm 1.6μm Roughness Ra: Min.Weight.of LM500 Main Unit (kg) Sample: D type Impact Device >5 >5 >1.5 >15 >5 Directly Measurable HLD Value Standard Test Block 0.5-1.5 5-15 Need Stable Support 0.05-2...
Impact Device Type Workpiece Surface Roughness Ra 4.2 Support of Test Piece D、 DC、DL 、D + 15 2 μm The heavy workpiece do not need support. 7 μm The Mid-heavy workpiece must be placed on a flat, sturdy surface, and 0.4 μm absolutely smooth placement, without any shaking.
5.3 Turn on the Instrument Press “on” to enter into the big font testing interface.It displays the default parameter for the very first start, and it displays the parameters automatically before power off for the next start. 5.4 Parameters Setting Press “MODE”...
Hardness Tensile Strength Steel and Cast Steel Figure 5.7 GC. IRON CrNi 5.4.5 Impact Times NC. IRON CrMo Under the parameter configuration interface can set the impact times, which C. ALUM is available from 1-8 times. 5.4.6 Set Upper Limit and Lower Limit Alarm BRASS CrMnSi Under the parameter configuration interface can choose whether to use the...
5.4.7 Impact Devices Under the parameter configuration interface can set the type of the probes among D,C,G and DL. 5.4.8 Languages Under the parameter configuration interface can interchange the languages among Chinese, English, Japanese, French and Italian. 5.4.9 Auto Power-Off Setting Under the parameter configuration interface can set the standby time among 5.6 Average Value 5 Min, 10 Min, 20 Min and automatic power off.
LM500 Leeb hardness tester is a high-tech product, maintenance work After starting up, press CAL to enter the calibration mode, then impact the should be operated by professionals, please do not operate by yourself.