novel XSZ-107T User And Operational Manual

Biological microscope with an epi-fluorescence yx-2 attachment

Advertisement

Biological Microscope
Model XSZ-107T
With an Epi-Fluorescence (YX-2) attachment
USER AND OPERATIONAL
MANUAL
This manual is for the biological microscope XSZ-107T. To ensure the safety and optimum performance of this
unit, please ensure a thorough familiarization with this manual so as to become comfortable with the use of this
unit.
1 |
P a g e

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the XSZ-107T and is the answer not in the manual?

Questions and answers

Summary of Contents for novel XSZ-107T

  • Page 1 USER AND OPERATIONAL MANUAL This manual is for the biological microscope XSZ-107T. To ensure the safety and optimum performance of this unit, please ensure a thorough familiarization with this manual so as to become comfortable with the use of this unit.
  • Page 2: Table Of Contents

    Using the Epi-Florescence attachment for specimen observation ..............14 11.1 Preparation ..............................15 Technical Specifications ............................16 13.1 Main specifications of the microscope XSZ-107T ..................16 13.2 Main specifications of the epi-florescence attachment YX-2 ..............16 12.1 Fluorescent Filter Combinations ......................... 17...
  • Page 3: Application

    (to name a few). The microscope unit is used in conjunction with the epi-fluorescent attachment which has been designed for use with and including our biological microscopes such as the Novel N-400M, Novel N320M and the Novel XSZ-107CCD.
  • Page 4: Maintenance And Care

    Indicates that the main switch is OFF. ● This manual is written for the XSZ-107T with an epi-fluorescent attachment YX-2 and before equipping it with any other ordinary microscope, be sure to learn how to use the microscope. 3. Maintenance and care 3.1 All the lenses have been adjusted properly, please do not dismount any lenses by oneself as this...
  • Page 5: Components Of The Microscope

    4. Components of the microscope P a g e...
  • Page 6: Assembly

    5. Assembly Assembly Diagram The Following figure shows the completed installation of the components: The corresponding figures show the assembly steps. • Before installation, ensure the unit is clean as well as every component, do not store unpacked units in a room which has high temperatures, high humidity or a dusty environment. P a g e...
  • Page 7: Assembly Procedure

    6. Assembly Procedure 6.1 Installing the epi-florescence unit Place the epi-florescence unit on top of the microscope body and tighten the holding screw firmly to hold the unit in place Installing the binocular head Insert the binocular viewing head into the epi-florescence unit and then mount it onto the microscope and ensure it is in the correct position and bolt it down to ensure no movement P a g e...
  • Page 8: Installing The Eyepieces

    Installing the Eyepieces Insert the eyepiece into the eyepiece tube until they are level with the eyepiece tubes as shown Installing the objectives • Adjust the coarse focus knob until the support device of the mechanical stage reaches its lowest limit position. •...
  • Page 9: Connecting The Power Cords

    Connecting the power cords • The cables and power cords are vulnerable when bent or twisted; never subject the power cords to excessive force. • Set the main switches to the off “0” position before connecting any of the power cords. •...
  • Page 10: Adjustment Set Diagram

    Adjustment Set Diagram 10 | P a g e...
  • Page 11 11 | P a g e...
  • Page 12: Adjustment And Operation For Brightfield Microscopy

    9. Adjustment and Operation for brightfield microscopy Brightness Adjustment 9.1.1 Connect the power cord and set the main power switch on to the (on) position “I” . 9.1.2 Turning the brightness adjustment knob, increase the brightness of the illumination set to the users’...
  • Page 13: Focusing On The Specimen

    10. Focusing on the specimen 10.1 Focus on the specimen Focus on the specimen using the 10x objective. To avoid touching the specimen during focusing, you should raise the stage to its maximum height adjustment and observe whether it is close to the objective at first, then using the fine focus slowly lower the stage while observing through the eyepieces until a clear crisp image is seen.
  • Page 14: Adjusting The Interpupillary Distance

    10.4 Adjusting the interpupillary distance The interpupillary distance range of the microscope head is 55mm ~ 75mm. Push the eyepieces together and then while looking through the eyepieces, move the distance apart until the left and right fields of view coincide correctly.
  • Page 15: Preparation

    11.1 Preparation 11.1.1 Verify that the plug from the AC mains outlet is connected firmly into the connector of the power supply units. 11.1.2 Make sure the power cord is connected firmly. 11.1.3 When transmitted light observation is required, pull out the filter system and make the hole in the light path.
  • Page 16: Technical Specifications

    13. Technical Specifications Main specifications of the microscope XSZ-107T Optical System 160mm path length Viewing head Siedentopf Binocular Head, inclined at 30° Interpupillary distance of 55 – 75mm Eyepiece Wide Field Eyepiece WF 10x / 18 Magnification 10x Focal length 25mm...
  • Page 17: Fluorescent Filter Combinations

    ●Indoor Use Operating ●Altitude: Max. 2000m Environment ●Ambient Temperature: 5℃ to 40℃ (41℉ to 104℉) ●Maximum RelaPve Humidity 80% for Temperature Up to 31℃ (88℉) Decreasing linearly through 70% at 34℃ (93℉), 60 % at 37℃ (99℉) to 50% relative humidity at 40℃ (104℉) ●...