CONTENTS WARNINGS & CAUTIONS GENERAL INFORMATION 6-10 DEFINITION OF THE GROUPS MENTIONED NON-COMPLIANT USE SAFETY INSTRUCTIONS GENERAL SAFETY INSTRUCTIONS SAFETY INFORMATION FOR THE OPERATOR SAFETY INFORMATION FOR THE USER SYMBOLS USED CLEANING & DISINFECTION SERVICING & MAINTENANCE SERVICE LIFE & DISPOSAL TECHNICAL SPECIFICATION OVERVIEW 12 -13...
Do not place pump on or near a heat source or cover with bedding. • PremiumLift advise against smoking whilst the system is in use to prevent the accidental ignition of associated items which may be flammable, such as bed linen.
Page 5
Removal of the case by unqualified personnel will invalidate the warranty. Before cleaning the unit ensure that the electrical supply to the pump has been disconnected by removing the plug from the power supply. www.premiumlift.com.au...
Read through this instruction manual conscientiously from start to finish. Note the safety instructions which must be observed. PremiumLift products are ISO Certified and TGA Registered and meet all safety and functionality requirements. These safety requirements can only be met if the user is satisfied with the proper condition of the product (including accessories) before use.
Instruction Manual must be read conscientiously and in detail by the user/care personnel. Programming of the system should be carried out by competent trained personnel. Use only spares and accessories supplied or approved by PremiumLift. Only suitably trained personnel are allowed to operate the system. www.premiumlift.com.au...
If there is a suspected fault or damage, unplug the mains plug from the socket and follow the power down procedure (page 16). Clearly mark ‘Out of Order’, take out of service immediately and inform the person in charge without delay. www.premiumlift.com.au...
This product must be disposed of in a designated refuse bin for waste electronic devices (WEEE) in the European Union. Do not dispose of as normal domestic waste. No smoking. No naked flames. Read instructions / consult manufacturers guide. This symbol indicates electrical hazards. There is danger to life and health. www.premiumlift.com.au...
Any defects, damage or signs of wear must be rectified without delay. Only original spare parts from PremiumLift may be used, otherwise all guarantees or warranties will be invalidated. See pages 16-17 for ‘Routine Maintenance’ and ‘Routine Servicing’.
Power Rating Voltage AC 240V / 50Hz Fuse T 1AH 250V Medical Classification Medical Device Class 1 Safety Standards EN 60601-1. EN 60601-1-2. ARTG No.398711 For any further assistance, please contact a member of the PremiumLift team at sales@premiumlift.com.au www.premiumlift.com.au...
With the switch in the ON position press the power button on the control panel to activate the pump. To turn the pump off press the power switch on the control panel. www.premiumlift.com.au...
Page 13
The pump unit must be placed so that it can be clearly seen and monitored at all times. www.premiumlift.com.au...
Ensure that the comfort setting is altered correctly when the patient moves from a lying to a sitting position and vice versa. NOTE: The static mode is only to be used for patient care. Switch off after use. www.premiumlift.com.au...
The mattress should be loosely rolled lengthwise with the cover innermost, taking care not to strain the feed pipes. It can then be stored / transported in the carry bag with the pump, mains cable and this booklet. Do not stack bagged mattresses more than two high. www.premiumlift.com.au...
• The individual cells can be wiped clean with a mild antiseptic solution. • All cells are replaceable and can be sourced from PremiumLift. POWER UNIT Disconnect the power unit from the electricity supply before carrying out maintenance, repairs or cleaning.
Page 17
SERVICING SERVICING YOUR SYSTEM The AirCare Max-Pro pump should be serviced every 12 months by PremiumLift approved personnel using genuine PremiumLift spare parts. Failure to follow the service schedule may invalidate future warranty claims (‘Guarantees & Warranties’ can be found on page 23).
CHECK THAT STAGE 2 CHECK IF PROBLEM PERSISTS Pump shows no 1. Mains plug is plugged in and 1. Connect the pump to the Contact PremiumLift technical indication that it is power switched on. nearest (working) mains outlet. support. powered up.
Page 19
STAGE 2 CHECK IF PROBLEM PERSISTS The pump appears PLEASE NOTE 1. Disconnect and then re- Contact PremiumLift technical Inflation can take up to 30 to be running but the connect the hoses to the outlet support. mattress is not inflating minutes.
Page 20
TROUBLESHOOTING FAULT CHECK THAT STAGE 2 CHECK IF PROBLEM PERSISTS Some of the cells This is normal for alternating pressure therapy. appear to be deflated. The mattress is made up of individual air cells. The alternating section is split up into 2 sections consisting of odd cells e.g.
Page 21
FAULT CHECK THAT STAGE 2 CHECK IF PROBLEM PERSISTS The pump is fitted to the bed Reposition the pump unit. The pump is vibrating Contact PremiumLift technical or making excessive correctly. support. noise. Before calling: The mattress is Check the comfort setting on Set the pump to the correct uncomfortable.
Our standard terms and conditions of sale can be found on our website or by request to PremiumLift. If this device malfunctions and a serious incident occurs please contact PremiumLift. A member of our service team will be happy to help you and offer advice. Email: sales@premiumlift.com.au...
Need help?
Do you have a question about the AirCare Max-Pro ACMaxPP and is the answer not in the manual?
Questions and answers