Baby Radiant Warmer
User Manual
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KR-1000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for novos KR-1000

  • Page 1 KR-1000 Baby Radiant Warmer User Manual...
  • Page 2 This document contains information on which NOVOS assumes the property rights. The information may not be reproduced in whole or in part unless authorized in writing by NOVOS. This information is the property of NOVOS, for its intended use only.
  • Page 3: Table Of Contents

    Novos KR-1000 Baby Radiant Warmer User Manual Table of Contents About This Manual ..............................7 Safety Information ............................12 1.1. User Obligations for Patient Safety ......................12 1.1.1. Patient Monitoring ..........................12 1.2. Limitation of Liabilities ..........................12 1.3. Usage Restrictions ............................12 1.3.1.
  • Page 4 Novos KR-1000 Baby Radiant Warmer User Manual 2.7.3. Accessory Mounting Rail (Optional) ....................34 2.7.4. Power Cable Coil ..........................35 2.7.5. Acrylic Protective Panels ........................35 2.7.6. Rotating Bed (Optional)........................35 2.7.7. X-Ray Tray............................36 2.7.8. E Cylinder Slots (Optional) ........................36 2.7.9.
  • Page 5 Novos KR-1000 Baby Radiant Warmer User Manual 6.4. Preparation ..............................64 6.5. Operating ..............................70 6.6. Cleaning and Maintenance......................... 70 Bililed Mini+ Phototherapy Unit (Optional) ....................71 7.1. Usage Purpose ............................71 7.2. User’s Responsibilities for Patient Safety ....................71 7.3.
  • Page 6 Novos KR-1000 Baby Radiant Warmer User Manual 9.6. Manufacturer ............................. 98 NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 7: About This Manual

    The intended users of this guide are end users of the device, healthcare providers in delivery rooms and neonatal intensive care units, and hospitals’ biomedical and clinical engineering services. The product is intended for professional use. Aim: This manual is a guide to the following items of the KR-1000 Baby Radiant Warmer System: • Installation •...
  • Page 8 Novos KR-1000 Baby Radiant Warmer User Manual Symbols This symbol means "Attention, see relevant documentation" and may appear on any hardware component of the device. This symbol means the device is in the BF class. This symbol is used to warn the users of the device against possible risk or injury.
  • Page 9 Novos KR-1000 Baby Radiant Warmer User Manual Trendelenburg position button Reverse Trendelenburg position button Product ID. Manufacturer. Input voltage. Date of Manufacture. Electrical and electronic equipment waste should not be disposed of in general municipal waste, but should be collected separately.
  • Page 10 Novos KR-1000 Baby Radiant Warmer User Manual Labels on the Device Product Identity Technical Service Warning Outer Box Label NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 11 Novos KR-1000 Baby Radiant Warmer User Manual Reusable Skin Probes Disposable Skin Probe NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 12: Safety Information

    NOVOS, product violation, etc. limit the liability of NOVOS regardless of whether it was the fault of NOVOS and regardless of the manner in which the fault occurred. NOVOS shall not assume any liability for compensation for damages arising from or related to the products and for the buyer's liabilities to third parties;...
  • Page 13: Operating

    1.3.3.Servicing • Servicing should only be carried out by persons who have been authorized by Novos to repair this device, in accordance with the procedures specified in the service manual. •...
  • Page 14: Warnings For The Use Of The Device

    • All devices to be used with KR-1000 must comply with the IEC 60601 standard. • Do not use any unapproved accessories with the KR-1000. Do not connect such a device to the accessory mounting arm. • KR-1000 is not recommended to be used with an extra thermoregulation device.
  • Page 15: Bearing Capacity

    There is no integrated oxygen monitor on the device. When using the Easypuff T-Piece resuscitator, it is recommended to monitor the set oxygen concentration with an independent oxygen monitor. Line all electronic devices, the KR-1000 Baby Radiant Warmer system should be handled with care to prevent damage. Consider the following measures. 1.3.7.Bearing Capacity Bed Weight Capacity The maximum capacity of the baby bed is 10 kg.
  • Page 16: Emc Restrictions

    The use of electrosurgical devices or other devices that emit high electric fields may adversely affect the operation of the KR-1000. In this case, the skin probe cable should be kept as far away from these devices as possible. The excess part of the power cord should not be placed on the baby bed. The use of such devices may cause inaccurate measurement (display of temperatures above normal) on the KR-1000 skin probe due to the electrical energy they emit.
  • Page 17: Transportation Restrictions

    Novos KR-1000 Baby Radiant Warmer User Manual Any device to be connected to the KR-1000 must comply with the following standards. • IEC 60601-1 (EN 60601-1) Medical Electrical Equipment Part 1: General Safety Requirements • IEC 60601-1 (EN 60601-1) Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Basic Performance Guarantee Standard: Electromagnetic Compatibility;...
  • Page 18: Warranty

    2 weeks before the end of the warranty period. The above provisions belong to the sole warranty which is applicable and provided by NOVOS. It cannot be aimed to describe or explain the warranty status in any other way.
  • Page 19: Product Description

    2. Product Description 2.1. Usage Purpose KR-1000 is a combined open bed system with a radiant warmer designed for different needs of term and preterm newborns, especially thermoregulation, resuscitation (optional), weight monitoring (optional) in delivery rooms, emergency and neonatal intensive care units of hospitals.
  • Page 20 • In infant mode (servo-controlled mode), for optimal temperature regulation, the KR-1000 measures the newborn's skin temperature using a double thermistor probe. In order to maximize the safety and accuracy of warming during operation, the newborn skin temperature is read with the help of the dual thermistors in the skin probe and these values are verified at every measurement.
  • Page 21: Controls And Gauges

    Novos KR-1000 Baby Radiant Warmer User Manual 2.5. Controls and Gauges 1 Warmer Head Grommet 11 Height Adjustment Pedals 2 Control Panel Side Protector 12 Wheel 3 Balance Control Unit Trendelenburg Control Unit 13 IV Pole Accessory Mounting Arm +...
  • Page 22: Overview To Control Panel

    Novos KR-1000 Baby Radiant Warmer User Manual 2.5.1.Overview to Control Panel The control panel can be divided into 5 different zones. 1. Timers 2. Usage modes 3. Alarms 4. Temperature settings 5. Lower buttons group 2.5.1.1. Timers 1 Timer 1 set/start button: The first timer value is set by holding down and when the button is released the timer starts counting.
  • Page 23 Novos KR-1000 Baby Radiant Warmer User Manual 2.5.1.2. Usage modes Green LED indicators in the upper right corner of the usage modes buttons indicate the active usage mode. 1 Pre-warming mode button: Used to activate pre-warming mode. 2 Manual stop button: Used to activate manual mode.
  • Page 24 Novos KR-1000 Baby Radiant Warmer User Manual 2 Target temperature setting: The target temperature in baby mode or the warmer power percentage determined in manual mode is displayed on this indicator. 3 Set temperature>37 button: If a skin temperature above 37 °C is targeted in Baby mode, the selection is confirmed by pressing this button.
  • Page 25: Balance Control Unit

    2.5.2.Balance Control Unit KR-1000 has an optional balance unit that allows it to measure the baby's weight and store up to 6 records. This unit can also show the difference between consecutive recordings. 1 Tare button: This is used to measure tare weight manually.
  • Page 26 Novos KR-1000 Baby Radiant Warmer User Manual Beware of situations where an object weighing 200g or more is removed from the cradle, otherwise the balance unit will start the tare process even if the baby is still in the cradle. This may cause inaccurate measurement.
  • Page 27 • After measuring, you can store it as described in the following section. 2.5.2.3. Weight Measurement Records The KR-1000 balance unit can store up to 6 measurements in memory and can show the difference between these records. • After a fixed value appears on...
  • Page 28 Novos KR-1000 Baby Radiant Warmer User Manual The weighing unit also shows the difference between previously taken records. For example, assume the 6 measurements are as follows; 1. 1.200 kg 2. 1.260 kg 3. 1.300 kg 4. 1.320 kg 5. 1.380 kg 6.
  • Page 29: Easypuff Resuscitation Device (Optional)

    2 Light Source Joint Body Kit: Allows the light source to be fixed to the spiral arm. 3 Spiral Arm: This is used to flexibly position the Bililed Mini+ light source. 4 Spiral Arm Fixing Screw: Allows the spiral arm to be fixed to the KR-1000 body. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 30: Trendelenburg Control Unit

    Novos KR-1000 Baby Radiant Warmer User Manual It is important to position the Bililed Mini+ light source so that it is not affected by the heat source. Otherwise, the warmer light source may cause permanent damage to the device. 2.5.5.Trendelenburg Control Unit 1 Up arrow button: Raises the foot end of the bed to give it a trendelenburg position.
  • Page 31: Power Input

    Novos KR-1000 Baby Radiant Warmer User Manual 2.5.6.Power Input Primary power switch: This is the primary on/off switch that energizes the electronic components of the device such as the control panel, height adjustment motor, and scale control unit. 2 AC power input: This is the connector where the AC power is supplied to the device.
  • Page 32: Height Adjustment Pedals (Optional)

    Novos KR-1000 Baby Radiant Warmer User Manual 2.5.8.Height Adjustment Pedals (Optional) 1 Lifting Pedal: Press the pedal lightly to increase the device height. 2 Lowering Pedal: Press the pedal lightly to decrease the device height. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 33: Other Mechanical Components

    Novos KR-1000 Baby Radiant Warmer User Manual 2.6. Other Mechanical Components 2.7. Warmer Head Ceramic resistances: Ceramic warmers with 400W power provide homogeneous warming on the baby bed surface. LED examination lamps: 5W power embedded examination lamps can be positioned by applying light force to their front or rear ends.
  • Page 34: Monitor Tray (Optional)

    Novos KR-1000 Baby Radiant Warmer User Manual 2.7.2.Monitor Tray (Optional) Monitor Tray: The 33cm x 37cm rotating monitor tray can be used to place devices such as patient monitor SpO2 monitor. Monitor Tray Fixing Screw: Loosen the fixing screw to adjust the position of the monitor tray.
  • Page 35: Power Cable Coil

    Novos KR-1000 Baby Radiant Warmer User Manual 2.7.4.Power Cable Coil 1 Cable Coil: The remaining part of the device's power cable can be wrapped around this coil. 2.7.5.Acrylic Protective Panels There are acrylic protective panels surrounding all sides of the baby bed. The front, right and left panels can be folded down by gently pushing in the direction of the arrow.
  • Page 36: X-Ray Tray

    Novos KR-1000 Baby Radiant Warmer User Manual 2.7.7.X-Ray Tray Maximum 14x14 '' X-Ray cassette can be placed on the X- Ray tray of the KR-1000. • Open the front protective panel by pushing it in the direction of the arrow.
  • Page 37: Rotating Drawer (Optional)

    Novos KR-1000 Baby Radiant Warmer User Manual Accessory Hanger: There is an accessory hanger that can rotate around itself at the back of the baby bed. 2.7.10.Rotating Drawer (Optional) 1 Upper rotating drawer: Rotating drawers can be opened by rotating freely 360 °...
  • Page 38: Wheel Locks

    Novos KR-1000 Baby Radiant Warmer User Manual Auxiliary tray has a maximum load bearing capacity of 10 kg. 2.7.12.Wheel Locks Wheel Locks: The wheels can be locked by lightly pressing the end of the lock pedal. Use opposite force to unlock.
  • Page 39: Air Probe

    Novos KR-1000 Baby Radiant Warmer User Manual 2.8.2.Air Probe Allows the device to measure the ambient air temperature. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 40: Operating Procedure

    Novos KR-1000 Baby Radiant Warmer User Manual 3. Operating Procedure Follow the steps below to ensure the most effective use of the KR-1000 Baby Radiant Warmer System. • Read all the information in this manual carefully. • Note the WARNING and CAUTION statements in this manual.
  • Page 41 Novos KR-1000 Baby Radiant Warmer User Manual 7. Connect the black balance and gray Trendelenburg unit cables under the cradle to the column. 8. Lift the baby bed with the bed tray and remove the cylindrical shaped load protectors on the balance module (optional) by unscrewing the 4 hex screws shown below.
  • Page 42 Novos KR-1000 Baby Radiant Warmer User Manual 10. Fix the phototherapy unit to the spiral mounting arm (1) and the spiral mounting arm to the KR-1000 (2). (Optional) 11. Fix the monitor tray with 4 screws. (Optional) 12. Place and secure the IV pole. (Optional)
  • Page 43 Novos KR-1000 Baby Radiant Warmer User Manual 13. Remove the 2 screws that hold the warmer head fixing plate on the back of the appliance and the plate. 14. Place the accessory mounting arm in its slot in the device column. (Optional) 15.
  • Page 44: Pre-Start Check

    Novos KR-1000 Baby Radiant Warmer User Manual 3.2. Pre-Start Check After the installation of the device, any maintenance / repair or modification, follow the steps below to confirm the normal operation of the device. 3.2.1.Alarm System Check Before you begin, make sure that: •...
  • Page 45 Novos KR-1000 Baby Radiant Warmer User Manual 7. Check if the power failure indicator light flashes and audible alarm sounds. 8. Plug in the power cord. 9. Check the restart of the warmer. 10. Connect the skin probe and make sure there is no alarm on the alarm indicator.
  • Page 46: Control Panel Check

    • Primary power switch is in the ON position. The following procedure can be used to verify that the electronic systems of the KR-1000 are working. 1. Press the power button on the control panel. 2. Make sure the audible start alert is activated.
  • Page 47: Preparation

    The use of electrosurgical devices or other devices that emit high electric fields may adversely affect the operation of the KR-1000. In this case, the skin probe cable should be kept as far away from these devices as possible. The excess part of the power cord should not be placed on the baby bed.
  • Page 48 Novos KR-1000 Baby Radiant Warmer User Manual Phototherapy devices can cause the patient's skin temperature to increase. Monitor skin temperature continuously with the help of a skin probe. It is recommended to use baby mode during phototherapy application. Radiant energy generated in the warmer may increase the insensible fluid loss in the patient.
  • Page 49: Essential Use

    Novos KR-1000 Baby Radiant Warmer User Manual 3.4. Essential Use In this section, essential usage procedure of KR-1000 Baby Radiant Warmer System will be explained. 3.4.1.Operating Device 1. Connect the power cord to a properly grounded socket and the power input socket (1).
  • Page 50: Placing/Removing Skin Probe

    Novos KR-1000 Baby Radiant Warmer User Manual 3.4.2.Placing/Removing Skin Probe Placing Skin Probe • Gently clean and dry the skin in the area where you intend to stick the skin sensor. • Remove the paper behind the sensor retaining washer.
  • Page 51: Operation Modes

    Novos KR-1000 Baby Radiant Warmer User Manual 3.4.3.Operation Modes 3.4.3.1. Baby Mode In baby mode, the KR-1000 keeps the baby's skin temperature constant by automatically adjusting the warmer power to compensate for heat loss caused by changing physiological and environmental conditions.
  • Page 52 3.4.3.2. Pre-Warming Mode Turn on the KR-1000 at least 15 minutes before placing the baby on the device and switch to pre- warming mode so that the baby does not lie on a cold surface. •...
  • Page 53 Novos KR-1000 Baby Radiant Warmer User Manual • The active mode indicator LED in the upper right corner of the pre-warming button will turn green. • If the air temperature is below 25 °C, the microcontroller adjusts the warmer power to 40%.
  • Page 54 Novos KR-1000 Baby Radiant Warmer User Manual • In manual mode, the warmer power is adjusted by the user with the up and down arrows in the temperature settings section of the control panel. • Press the up arrow to increase the warmer power percentage.
  • Page 55 Novos KR-1000 Baby Radiant Warmer User Manual • After the warmer's power automatically drops to 50%, the “control baby alarm” is triggered. • Press the Manual button again to reset the previous warmer percentage level. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 56: Alarms

    Novos KR-1000 Baby Radiant Warmer User Manual 3.5. Alarms Alarm silencer Alarm Mode Cause Action function More than 80% energy has Warmer Bypasses the Baby been given to the warmer power drops audible alarm for 15 minutes. to zero. only...
  • Page 57 Novos KR-1000 Baby Radiant Warmer User Manual Warmer Bypasses the Skin probe is faulty or Baby power drops audible alarm disconnected to zero. only Sensor Failure Manual Pre- warmer Warmer Bypasses the Axis Failure All modes Warmer head not aligned...
  • Page 58: Troubleshooting

    Novos KR-1000 Baby Radiant Warmer User Manual 4. Troubleshooting 4.1. Fast Troubleshooting Guide The troubleshooting tables provide the user with general conditions, possible causes, and recommendations. If, in special cases, these tables cannot help solve the problem, the warmer should be sent to the authorized service.
  • Page 59: Routine Cleaning And Maintenance

    5. Routine Cleaning and Maintenance 5.1. General Cleaning The KR-1000 Baby Radiant Warmer System should be cleaned at intervals specified in the hospital protocol. Consider the following items for cleaning. Before cleaning, always disconnect the appliance from the power supply and wait for one hour to make sure that the warmer’s components have cooled.
  • Page 60: Cleaning Skin Probe

    Novos KR-1000 Baby Radiant Warmer User Manual • Use only NOVOS approved cleaning solutions. Liquid solutions can be used for cleaning purposes. For example, hospital disinfectants and microbactericides. • Remove dust from all plastic surfaces with a clean damp soft cloth.
  • Page 61: After Cleaning

    1 year. Records must be kept during these protective services. A technical services contract with NOVOS is recommended for a continuous service in line with the rules and standards. It is recommended to contact the NOVOS Technical Service Department for maintenance and repair of the KR-1000 Baby Radiant Warmer System.
  • Page 62 Novos KR-1000 Baby Radiant Warmer User Manual • The functionality of the membrane buttons of the control panel, trendelenburg control unit and optional balance control units is checked. • 0 kg and 5 kg calibrations of the balance unit are performed.
  • Page 63: Easypuff T-Piece Resuscitation (Optional)

    Novos KR-1000 Baby Radiant Warmer User Manual 6. Easypuff T-Piece Resuscitation (Optional) 6.1. Usage Purpose The Easypuff T-Piece Resuscitator is an easy-to-use, manually operated, gas-powered resuscitator that provides controlled and accurate resuscitation of newborn babies in delivery rooms and neonatal intensive care units.
  • Page 64: Preparation

    Novos KR-1000 Baby Radiant Warmer User Manual NOVOS cannot guarantee or approve the safety performance of third party accessories. Use only original parts and accessories of NOVOS. The use of unauthorized parts or accessories may cause serious damage to the patient and the device.
  • Page 65 Novos KR-1000 Baby Radiant Warmer User Manual Flowmeter System Air/Oxygen Mixer It is recommended to set the total current outlet to 8 lpm. If a flowmeter system is to be used, make sure that the total flow output of the air / oxygen mixture does not exceed 15 lpm. For example, setting both oxygen and air flow to 4 lpm will provide a total flow of 8 lpm.
  • Page 66 Novos KR-1000 Baby Radiant Warmer User Manual Oxygen concentration 100% Air - 1 Air - 0.87 Air - 0.75 Air - 0.62 Air - 0.5 Air - 0.37 Air - 0.25 Air - 0.12 Air - 0 - 0.12 - 0.25 - 0.37...
  • Page 67 Novos KR-1000 Baby Radiant Warmer User Manual 3. Connect the T-Piece patient circuit. Please be careful not to damage the gas inlet when connecting the gas supply to Easypuff. A gentle force will be sufficient for the connector probe. •...
  • Page 68 Novos KR-1000 Baby Radiant Warmer User Manual 5. Check the maximum pressure. PEEP Head • Close the *PEEP head using your thumb and turn the PIP control clockwise all the *PEEP: Positive End Expiratory Pressure **PIP: Peak Inspiratory Pressure •...
  • Page 69 Novos KR-1000 Baby Radiant Warmer User Manual 6. Set PIP Value • While closing the PEEP cap with your thumb, turn the PIP control counterclockwise until the desired peak inspiratory pressure is set. 7. Set PEEP Value • Take your thumb off the PEEP cap and adjust the PEEP value by turning the PEEP cap.
  • Page 70: Operating

    Novos KR-1000 Baby Radiant Warmer User Manual 6.5. Operating 1. Adjust the gas flow to the desired flow rate (5-15 lpm) and oxygen concentration. 2. Select an appropriately sized neonatal resuscitation mask and connect the T-piece to the circuit. Place the mask over the baby's mouth and nose or connect the T-piece to the endotracheal tube.
  • Page 71: Bililed Mini+ Phototherapy Unit (Optional)

    This may cause an increase in temperature as well as prevent the radiation of the therapy light. NOVOS cannot guarantee or support the safe performance of third-party accessories for use with Bililed Mini+ phototherapy device. Use only original NOVOS accessories.
  • Page 72: Use Restrictions

    Bililed Mini+ Light Source: For phototherapy lamp operation, see the Operating Device section. Spiral Arm: This is used to flexibly position the Bililed Mini+ light source. Spiral Arm Fixing Screw: Allows the spiral arm to be fixed to the KR-1000 body. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 73: Power Supply Connection

    Novos KR-1000 Baby Radiant Warmer User Manual 1. Loosen Spiral Arm Fixing Screw and connect Spiral Arm. 2. Tighten Spiral Arm Fixing Screw and fix Spiral Arm. 3. Place the Bililed Mini+ as followings; Not to be exposed to direct heat b.
  • Page 74: Operating

    Novos KR-1000 Baby Radiant Warmer User Manual 7.5. Operating 7.5.1.Phototherapy Operation • Connect the power adapter (12V 0.7A) to Bililed Mini+ and plug it in. • Turn on the device by pressing the "Power" button. • Flashing green LEDs indicate that the device has started.
  • Page 75 Novos KR-1000 Baby Radiant Warmer User Manual Bililed Mini+ must be properly positioned and operated before therapy is initiated. • There are 5 preset therapy light intensity levels in total. • Adjust the light intensity of the blue therapy LEDs using the "Up and Down"...
  • Page 76 Novos KR-1000 Baby Radiant Warmer User Manual • Set the target therapy hour digit using the "Up" and "Down" arrow buttons. • Press the “Confirm” button. • Set the target therapy minute digit using the "Up" and "Down" arrow buttons.
  • Page 77 Novos KR-1000 Baby Radiant Warmer User Manual • Press the "confirm" button to exit the menu and return to the main screen. • Wear an appropriately sized eye mask so that the baby's eyes are completely covered. • Press the "Therapy Start/Pause/Stop”...
  • Page 78 Novos KR-1000 Baby Radiant Warmer User Manual • Stop therapy by holding down the "Therapy Start/Pause/Stop” for 3 seconds. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 79 Novos KR-1000 Baby Radiant Warmer User Manual • A long beep will be heard. • Blue phototherapy and status LEDs will turn green. • The therapy started icon in the upper right corner of the screen will disappear. • The therapy time counter reverts to the target therapy time.
  • Page 80 Novos KR-1000 Baby Radiant Warmer User Manual To return to the therapy screen • Press the confirmation button 3 times, waiting for transition times between menus. • Finally, the therapy screen will be displayed. When the targeted therapy time is reached, the message "Therapy Ended" will be displayed on the screen and this message will remain on the screen until any button is pressed.
  • Page 81: Use Of Examination Lamps

    Novos KR-1000 Baby Radiant Warmer User Manual 7.5.2.Use of Examination Lamps • Bililed Mini+ has 4 white examination LEDs. • Press “Examination lamp” button. • Enter the menu by pressing the "Confirm" button. • Go to "Lamp Intensity" option in the menu by pressing the "Down"...
  • Page 82 Novos KR-1000 Baby Radiant Warmer User Manual • There are 5 different preset levels for the examination lamp’s light intensity. • Set the appropriate light intensity level for the white inspection LEDs using the "Up and Down" arrow buttons. •...
  • Page 83: Cleaning And Maintenance

    Such disinfectants can damage the phototherapy unit. WARNING! NOVOS recommends the use of Ammonia Quaternary components that do not have any harmful effects on the materials used in the device, that clean the surfaces and fight bacteria. NVS.S9.KK04 / Rev.06 / 27.07.2021...
  • Page 84 Novos KR-1000 Baby Radiant Warmer User Manual Cleaning of Air Channels Air channels should be protected from dust for effective use of the device and long-lasting operation of the LED module. Clogged cooling channels due to dust cause the internal temperature of the device to increase, and this can lead to damage to electronic components.
  • Page 85: Maintenance

    A technical services contract with NOVOS is recommended for a continuous service in line with the rules and standards. It is recommended to contact the NOVOS Technical Service Department for repair of the Bililed Mini+ Phototherapy device. WARNING! Technical service and maintenance services should only be provided by technically qualified persons.
  • Page 86: Venturi Vacuum Unit (Optional)

    Novos KR-1000 Baby Radiant Warmer User Manual 8. Venturi Vacuum Unit (Optional) 8.1. Usage Purpose The venturi vacuum unit is used to create, adjust and measure a suction from a compressed gas source at the wall or a cylinder (oxygen or medical air). It allows suction of liquid or mucus in the absence of a vacuum pipeline network.
  • Page 87: Operating

    Novos KR-1000 Baby Radiant Warmer User Manual The venturi vacuum unit is not suitable for use in the MRI environment. 8.3. Operating 1. Make sure the device is not damaged and the connector is compatible with the supply connection. 2. Make sure the adjustment knob is in the closed position (turn fully clockwise).
  • Page 88: Technical Specifications

    Novos KR-1000 Baby Radiant Warmer User Manual 1. Remove the collection container and unscrew the lid from the container. 2. Immerse the parts of the collection container (following the manufacturer's instructions for soaking, rinsing and drying times) in a disinfectant solution.
  • Page 89: Annexes

    Novos KR-1000 Baby Radiant Warmer User Manual 9. Annexes 9.1. Technical Specifications 9.1.1.Environmental Properties Operating Temperature 20°C-30°C Operating Humidity 5-95% RH, non-condensing Storing Temperature -20°C - 60°C Storing Humidity 5-95% RH, non-condensing 9.1.2.Mechanical Properties Height 177 cm Height (Adjustable, optional)
  • Page 90: Other Properties

    Novos KR-1000 Baby Radiant Warmer User Manual conditions) Skin Temperature Display Range 10 – 50 °C Temperature Measurement Accuracy 0,1 °C Fuse 2x4A Skin Temperature Sensor Resolution 0,1 °C Touchless Alarm Silence 10 min. Preheat Mode Heater Power Percentage 40% at temperatures below 25°C, otherwise 30%...
  • Page 91: Accessories And Consumables

    Novos KR-1000 Baby Radiant Warmer User Manual 2. Level: >20 μW/cm^2/nm – 1 mW/cm^2 – 1000 μW/cm^2 1. Level: >15 μW/cm^2/nm – 0.75 mW/cm^2 – 750 μW/cm^2 Wavelength 440 – 460 nm Lamp Life 50,000 hours Maximum Noise Level <10 dBA Software Version: 3.x...
  • Page 92: Compatibility

    Novos KR-1000 Baby Radiant Warmer User Manual 9.2. Compatibility NOVOS baby warmers and accessories are designed to meet the requirements of: IEC 60601-2-21 Safety requirements for baby radiant warmers IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety...
  • Page 93: Guidance And Manufacturer's Declaration Of Electromagnetic Compatibility

    This device is designed for use in below conditions and environments. 9.3.1.Electromagnetic Environment KR-1000 is intended for use in the electromagnetic environment specified below. The customer or user of the KR-1000 must guarantee that the device will be used in such an environment.
  • Page 94: Electromagnetic Immunity

    Novos KR-1000 Baby Radiant Warmer User Manual 9.3.2.Electromagnetic Immunity KR-1000 is intended for use in the electromagnetic environment specified below. The customer or user of the KR-1000 must guarantee that the device will be used in such an environment. IEC 60601...
  • Page 95 Novos KR-1000 Baby Radiant Warmer User Manual Electromagnetic Immunity Test IEC 60601 test level Compatibility level Environment Main power quality should be in typical commercial quality or suitable for hospital environment. 95% drop for 0.5 95% drop for 0.5 If the user of the KR-1000...
  • Page 96 RF transmitters, electromagnetic site survey should be considered. If the measured field strength in the location in which the KR-1000 is used exceeds the applicable RF compliance level above, the KR-1000 should be observed to verify normal operation. If an...
  • Page 97: Recommended Separation Distance

    Recommended separation distance between portable and mobile RF communications equipment and the KR-1000 The KR-1000 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the KR-1000 can prevent...
  • Page 98: Equipment Classifications

    IP Class IP20 Not suitable for use in the presence of flammable anesthetics 9.5. Trademark Registrations KR-1000™, Easypuff™, Bililed Mini+™ models are Novos branded commercial products of Novos Tıbbi Cihazlar Sanayi ve Ticaret İthalat ve İhracat Limited Şirketi. 9.6. Manufacturer Novos Tıbbi Cihazlar Sanayi ve Ticaret İthalat ve İhracat Limited Şirketi...