Trouble shooting
(1) 1) No lights on power unit
Check the power unit is connected to the mains
power supply and that the mains switch is turned ON.
Check power unit for any blown fuses.
Low Pressure indicator is flashing
(2) 2) Low Pressure indicator is flashing (RED)
Check whether power was suddenly shut down.
Check that the connection between air tube and
power unit is tightly secured.
Check that all tubing connections along the mattress
are secured.
If all of above steps have been checked.
Touch "Alarm Mute" for system to be verified again.
(3) 3) Power Failure Alarm Indicator is flashing (RED) and
sounding
If in use - Check the power unit is connected to the
mains power supply and that the mains switch is
turned ON.
If being taken out of use - Check pump has been
switched OFF on the panel before at power source.
Reconnect to power supply and turn pump ON, then
OFF.
Please wait a few minutes to see if the issue
resolves.
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
(4) 4) The power unit is operating but the mattress is
not alternating
Ensure that the mattress inflation process is
complete.
Check that the CLP indicator on the control panel
is not illuminated. If it is turn CLP off and alternating
mode will engage. Please wait a few minutes and
check that the cells are now alternating.
(5) 5) Power unit is noisy
• Ensure that the power unit is resting against a solid
surface.
If the problem persists, contact Ultimate Healthcare or
your local service provider.