EDIZIONE 01/12 OPERATING MANUAL SPARE PART LIST GP + GP VACUUM GROUPS GENERAL EUROPE VACUUM Srl VIA DEL COMMERCIO 9 20090 BUCCINASCO (MI) ITALY TEL. ++39 02 48841120 FAX ++39 02 4453957 e-mail: info@gevac.com http://www.gevac.com...
Page 2
___________________________________________________________________________ 2 of 43 GP+GP – Maintenance book Pag.
Page 3
It is absolutely mandatory that these operating instructions be read and understood prior to the vacuum group installation and start up. The GEV vacuum groups have been manufactured according to the newest technical standards and safety regulations as per “Machines directives 2006/42/ CE” and additions.
Page 4
ASSISTANCE If you send a pump to General Europe Vacuum indicates whether the pump is free of substances damaging the health or whether it is contaminated. If contaminated, indicate the nature of hazard. Warning The pump must be packed in a way that will not be damaged during shipping and that no harmful substances can escape from the package.
Page 5
GENERAL INDEX DESCRIPTION PARAGRAPH GENERAL INSTRUCTIONS: RISKS LIST - Mechanical risks - Electrical risks - Temperature risks - Acoustic risks OPERATING INSTRUCTIONS DESCRIPTION - Operating principle - Main parts into the vacuum system - Lubricants - Transport, lifting, manipulation and stock ...
Page 6
The presence inside the vacuum group of moving parts leades to the evaluation of the possible mechanical risks due to the same parts. For this reason, GEV has equipped all their machine with protective casing firmily screwed on pumps and fit for protecting dealing with the pumps. It is recomended to provide for skilled personnel.
Description Operating principle : GEV vacuum groups GP+GP or GPx2 are assembled with two or more oil-sealed rotary vanes vacuum pumps, single stage, type GP. Each vacuum group is calculated and manufactured in order to work with one vacuum pump with technical datas enough for the vacuum application. A second pump, installed in stand by, will start-up whenever the vacuum level won’t be keep with just one...
Page 8
General Europe Vacuum recommends its special oil “TYPE G 880” to allow vacuum pump works properly If you use a non-approved oil, GEV cannot guarantee that the pumps will meet their operating specifications ( ultimate pressure, pumping speed, operating temperature etc. ) However the warranty is voided only if a non-approved oil adversely affects the operation and reliability of the pump.
The standard vacuum pump does not fit for being installed in explosion and hazard areas. Take contact with GEV if you are planning to install it in such areas. Before installing pump, or in case of maintenance, disconnect it from the electric mains to prevent the pump from running up suddently.
B.2.2 Exhaust side The vacuum pumps type GP are equipped in exhaust by one or more oil remover cartridges (according to the pump size) having a filtration in the range of 0,01 micron. When the oil remover cartridges are clogged, smokes in exhaust is visible, the pump temperature and the absorbed power gets increased and it may happen that the motor thermal protection switches on.
Page 11
Warning If any motor or electric switch cuts the pump out. Re-start up of the pump can be carried out only manualy. After having carefully checked the stopping reason. Vacuum group can start up also when suction is already under vacuum. After connecting motor and every time you alter the wiring, check motor rotation direction, check the arrow shown on motor lable.During check, suction must be open.
Page 12
The differential of pressure could be changed, from 1 to 10, as you can see opening the instrument. GEV suggests to put it around 3-4, with value of 120-150mmHg.
Pump stop GP vacuum pumps are equipped in suction by an anti-suckback valve closing the suction when vacuum pump is off keeping vacuum steady in all the connected appliances thus avoiding the flush back of oil into the appliance same. Therefore, under normal circumstances, pump stops after disconnection to mains.
Maintenance schedule The maintenance schedule stated in the maintenance sheet underneath are approximate values for a normal function of the group. Difficult environmental conditions and/or aggressive gases can increase significally the maintenance interventions. CHECK AND MAINTENANCE FREQUENCY Oil level check Daily Oil conditions check Depending upon process...
Oil change Oil must be changed according to the following chart : VACUUM PUMPS MOD. GP – GPM 11 or 21 200 hours running Or every 1 / 2 months GP – GPM 45-65-100 500 hours running Or every 3 / 4 months GP –...
Page 16
Electric motor disassemblig WARNING Check that motor cannot start and the electric main is off For disassembling the electric motor unscrew the three screws which are coupling pump to turret. Pull out the joint with fun from the motor’ shaft Unscrew the 4 fixing screws from motor turret After check out and replacement of the electric motor, reassemble all operating inversaly Beware...
VACUUM SWITCH D.1 Regulation Open and close the valve (pos.5) up the Vacuum switch code no. 62007. If rotate it, the switching vacuum level change (check the value thru the glass). Remove the mask and change from 1 to 10 the wheel (pos.19).
TROUBLESHOOTING GUIDE FAULTS CAUSES REMEDY Pump does not Motor is not properly connected Connect motor correctly start The thermal device has not been Readjust the thermal device thermal device properly adjuested starts after few The Volts do not fit with the motor’s one Replace motor Pump is stucked Replace pump...
Page 19
The pump is too Cooling air is not enough Place the pump in a more aired location Piping or fan is dirt Clean with air or solvent Too high temperature of the gas aspired Verify the system and the reiability of the pump Insufficent oil quantity in the pump Restore oil level...
Page 20
___________________________________________________________________________ 20 of 43 GP+GP – Maintenance book Pag.
Page 21
Vacuum pump model GP/ GPM 11-21 Components list (refer to the enclosed split draw) QUANTITA DESCRIZIONE CODICE CODICE QUANTITA DESCRIZIONE CODICE GP/M GP/M GP/M GP/M GP/M GP/M GP/M GP/M ELECTRIC MOTOR 650221075 650220001 SCREW 624206025 624206025 STRUCTURE/MOTOR FLANGE 190007 190007 BODY OUTLET GASKET* 190020 190020...
Page 22
___________________________________________________________________________ 22 of 43 GP+GP – Maintenance book Pag.
Page 36
___________________________________________________________________________ 36 of 43 GP+GP – Maintenance book Pag.
Page 37
Vacuum pump model GP/ GPM 308 Components list (refer to the enclosed split draw 2) QUANTITA DESCRIZIONE CODICE QUANTITA DESCRIZIONE CODICE GP/M GP/M WASHER 624171006 MOTOR SIDE PUMP COVER 230006 SCREW 624064816 O-RING * 628104587 INLET FLANGE 220011 BEARING 645900030...
Page 38
PUMP TYPE GP /M - 11/21 1 VACUUM INLET 2 OUTLET TO ATMOSPERE 3 OIL DISCHARGE CAP 4 OIL SIGHT GLASS 5 OIL FILL UP CAP PUMP TYPE GP 11 GPM 11 GP 21 GPM 21 Speed: 50 Hz 60 Hz 50 Hz 60 Hz 50 Hz...
Page 39
PUMP TYPE GP/M - 45/65 1 VACUUM INLET 2 OUTLET TO ATMOSPERE 3 OIL DISCHARGE CAP 4 OIL SIGHT GLASS 5 OIL FILL UP CAP 6 MOTOR WIRING BOX GPM 45 PUMP TYPE GP 45 GP 65 GPM 65 Speed: 50 Hz 60 Hz 50 Hz...
Page 40
PUMP TYPE GP/M - 100 1 VACUUM INLET 2 OUTLET TO ATMOSPERE 3 OIL DISCHARGE CAP 4 OIL SIGHT GLASS 5 OIL FILL UP CAP 6 MOTOR WIRING BOX GP 100 GPM 100 TIPO Speed: 50 Hz 60 Hz 50 Hz 60Hz Displacement m3/h...
Page 41
PUMP TYPE GP/M - 210 1 VACUUM INLET 2 OUTLET TO ATMOSPHERE 3 OIL DISCHARGE CAP 4 OIL SIGHT GLASS 5 OIL FILL UP CAP 6 MOTOR WIRING BOX GPM 210 PUMP TYPE GP 210 Speed 50 Hz 60 Hz 50 Hz 60 Hz Displacement...
Page 42
1 VACUUM INLET 2 OUTLET TO ATMOSPHERE 3 OIL DISCHARGE CAP 4 OIL SIGHT GLASS 5 OIL FILL UP CAP 6 MOTOR WIRING BOX PUMP TYPE GP 308 GPM 308 Speed 50 Hz 60 Hz 50 Hz 60 Hz Displacement m3/h...
Page 43
SAFETY INSTRUCTIONS FOR INSTALLATION IN HAZARDOUS AREAS Before to install rotary vane vacuum pumps type GP / GPM in potentially explosive areas, classified accordingly to ATEX Directive 94 / 9 / EC please read carefully these instructions. Read carefully Operation and Maintenance Manual ...