If a core reaches the end of the conveyor, an electronic eye will stop the belt. The software will continue to collect data and correct for the elapsed time. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
The software will continue to collect data and correct for the elapsed time. 8. When the last core sample has left the shield, click the “Stop” button. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 6
Pos (rel-ft): The distance to move the belt when Drive Mode is set to Position. Execute: Check this box to begin moving the belt. Uncheck it to stop the belt. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 7
Pause: Check this box to pause the test. The unit will stop collecting data and the belt will stop moving the samples. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
10. Enter the starting depth in the “Top of Interval” field. Enter the ending depth in the “Bottom of Interval” field. Make sure the “Positive Depths” field is checked. 11. Click OK to compile the log. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
High Voltage Setting: This is the amount of voltage being sent to the photo multiplier. Gain Fine: The amplifier gain. Get Versions: Populates the “DigiBASE Info” box OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 10
Stop: Stop logging data. Standard: Choose a standard for a manual calibration. Export Standard: Save the current data as a calibration for the selected standard. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
1. Place the three standards on the belt with Potassium first, then Uranium, then Thorium. Place the Potassium standard centered under the shield and leave 12 inches between the standards. Shield Height Sensor OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 12
3. Click the “Accept” button. The software will scan each standard for 10,000 seconds and the automatically advance the belt to the next standard. 4. When the calibration is complete, click the “Save” button. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 15
5. Click the “Start” button to start logging. 6. When the True Time has elapsed, logging will stop automatically. Click the “Export Standard” button to save the calibration. Export Standard OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
4. When the belt stops moving, measure the distance between the two marks. The belt will travel approximately 6 feet. 5. Enter the value in the software and click the “Accept” button. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 18
5. Click the Calculate button to save the calibration. 6. Place the 2.376" Calibration Gauge (#700-410-069) under the scanner. The height should be 2.00" - 2.75". If not, repeat this procedure. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 19
Y TickExt: The length of the tick marks on the Y-axis. Border: Select a border color. X Grid: Select a color for the X-axis gridlines. Y Grid: Select a color for the Y-axis gridlines. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 20
Least Max value, whichever is higher. Enter NaN to set it to the highest data point on the graph. Nom Num of Increments: The number of increments on the Y-axis. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
3. Use a torque wrench to re-tighten the four (4) Socket head cap screws to 18 ft-lbs. 4 Flat Coutersink Head Cap Screws Left Knob Belt Tensioner Right Knob Belt Tensioner 4 Socket Head Cap Screws OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
During normal operation a red light on the transmitter and green light on the receiver should both be on. If one or both is not on, the system must be serviced by an OFITE technician. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
10. If the calculated speed is 5% over or under the set speed, adjust the belt tension (see page 20) and repeat steps 6 – 10. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Turning freely without bearing noise, product build-up, or carry back. Belt tracking correctly c. Carry Side Idler turning freely. d. Head Pulley and or Drive Pulley running smoothly. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...