LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER TABLE OF CONTENTS SPECIFICATIONS DESCRIPTION OF FAB-2S FRIABILITY TESTER DESCRIPTION OF CONTROLS INSTALLATION AND SET UP OPERATING PROCECURES MAINTENANCE TROUBLE SHOTTING SPARE PARTS AND SERVICE INFORMATION Page 2 of 12...
LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER SPECIFICATIONS Drum Position: Drum Speed: Variable Speed 25-70 RPM Speed Accuracy Per Minute: ±1 RPM Maximum Testing Time: 9 HR 59 MIN. 59 SEC. Maximum Count of Revolutions: 99999 Revolutions Size: 12” x 16” x 14”...
The standard Friability drum causes the samples to roll and fall during the test while the abrasion drum continuously stresses the sample by abrasion of lamellas. FAB-2S Friability Tester can be used 1 or 2 drum. A noiseless gear motor rotates the drums with a variable speed of 25-70 RPM that programmed by users.
Page 5
LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER FAB-2S AUTOMATED DISSOLUTION TESTER FRONT VIEW Page 5 of 12...
Page 6
LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER CONTROL LOCATIONS ON FRONT PANEL OF FAB-2S Power Switch Location: Left side control panel Purpose: Turns the system on/off Display Purpose: To display timer or counter Timer Indicator Light Purpose: To indicate Friability Tester is in timer mode.
LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER INSTALLATION AND SET UP Remove the unit from packing box. Remove both drums from packing box. Remove drum nut from Friability Tester Install drums inf. Shafts. Both drums are marked. One marked “L” goes to left side.
Page 8
LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER Set Minutes: A. Press “Set” button, the third to last digit will start flashing. B. Press “Enter” to increase the flashing digit. (Skip step B if no change is required.) C. Press “Set” button, the fourth to last digit will start flashing.
Page 9
Once you have finished setting your time or count, press “Start/Stop” button. FAB-2S will start rotation for timer or counter mode. To Stop Test FAB-2S will automatically stop when it reaches the preset time or count. To stop test before it reaches its preset time or count, press “Start/Stop” button. Warning: Below information for the lab manager only, do not show technician how to change the rotation speed to avoid future problems.
Page 10
E. Press “Enter” to increase the flashing digit. (from 2 to 7.) F. Press “Reset” to enter the new speed setting into the FAB-2S. G. Turn off the FAB-2S and power on again to check the new speed setting on the display to start the test at the new speed.
LOGAN FRIABILITY TESTING SYSTEM FAB-2S FRIABILITY TESTER Maintenance Keep FAB-2S Friability Tester in a dry and flat surface Drum After test please clean by using a dry cloth. Do not use any solvent. Lay on clean flat surface. Do not store at vertical position.
Page 12
When power is on constantly long beep Check Buttons if join with case No rotation Check Motor and Belt If further assistance is needed please contact your local distributor or Logan Instruments: Address: 19 C School House Rd. Somerset, NJ 08873...