■ IQC Checking
IQC timeout at 8 AM daily
Power on the h 232
The measurement begins
automatically
■ Daily Cleaning (70% alcohol)
A) Clean the meter surface
with moistened tissue
B) Remove and wipe the
green cover
Quick Reference Guide
Press Control Test
The result is displayed
as Pass or Fail
Clean the inside area
with cotton swab
Insert the IQC test strips
(Low and High)
Return the test strip quickly
back into the container
* Be gentle & caution
when cleaning the
detection window on
the side