Summary of Contents for USA Scientific ErgoOne

  • Page 3 L o r e m I p s u m A D J U S T A B L E V O L U M E P I P E T T E...
  • Page 5: Table Of Contents

    4. Cleaning and Decontamination ______________________ 5. Technical Data ___________________________________ 6. Calibration ______________________________________ Maintenance _____________________________________ 7.1 Maintaining Single-Channel Pipettes __________________ 7.2 Maintaining Multi-Channel Pipettes __________________ Ordering Information _____________________________ 9. Pipette Tip Compatibility __________________________ 10. Warranty ________________________________________ ErgoOne · User Manual ·...
  • Page 6: Introduction

    • Volume lock single-channel, eight-channel and twelve-channel models. • Minimal operating forces • Extremely accurate and precise • Lightweight • Low-maintenance high-quality product • Fully autoclavable · ErgoOne · User Manual...
  • Page 7: Operation

    – Declaration of conformity with test record – Labels for personalization/ TipOne® pipet tips are recommended to identification obtain the best results with ErgoOne®. – Auxiliary tool (single-channels up to 1ml only) – Pipette filters (5ml only) ErgoOne · User Manual ·...
  • Page 8: Setting The Volume

    Yellow 300µl 30 – 300µl Green 1000µl 100 – 1000µl Blue 5000µl 500 – 5000µl Violet Table 1 Examples of typical volume displays: 10µl Pipette 200µl Pipette 1000µl Pipette 10.00µl 0.50µl 200.0µl 20.0µl 1000µl 100µl · ErgoOne · User Manual...
  • Page 9: Pipetting

    – Allow the volume control button – Remove the pipette tip from the to return slowly back to the starting liquid. position. – Press the tip ejector button eject the pipette tip. ErgoOne · User Manual ·...
  • Page 10 TipOne® RPT tips have optimized surfaces that ensure improved delivery of detergents, viscous liquids, small volumes, and other difficult samples. Ask your local USA Scientific representative for more information or go to www.usascientific.com. · ErgoOne · User Manual...
  • Page 11: Pipette Identification

    This serial number can liquids that strongly differ from the be found behind the tip ejector button physical properties of water, eg. when it is pressed down. with different density (see also section 6 Calibration). ErgoOne · User Manual ·...
  • Page 12: Troubleshooting

    The following table provides information about the possible causes of problems with suggestions as to how these can be fixed. If you are not able to fix the problem please call USA Scientific at 800-522-8477 or 352-237-6288. You may also contact your local representative or send an email to infoline@usascientific.com.
  • Page 13: Cleaning And Decontamination

    Internal cleaning ejector sleeve. – Regularly check the pistons, springs, Note: The ErgoOne® piston material is seals, and the tip cones. For internal an innovative, wear-resistant plastic cleaning, we recommend a mild...
  • Page 14: Technical Data

    30 – 300µl 150µl ±1.0% ±0.5% 300µl ±0.6% ±0.3% Measurement data determined in compliance with the test conditions for piston-stroke pipettes according to EN ISO 8655 Measurement carried out with original TipOne pipette tips ® · ErgoOne · User Manual...
  • Page 15: Calibration

    If for example 1000µl is realigned by 10µl (1%) for a 100 –1000µl pipette, then for 100µl, the pipette must also be corrected by 10µl = 10%. ErgoOne · User Manual ·...
  • Page 16: Maintenance

    Note: If you make any changes to the pipette or replace individual parts, a check with recalibration is required. If the pipette cannot be calibrated successfully or if a fault continues, contact USA Scientific. · ErgoOne · User Manual...
  • Page 17: Maintaining Single-Channel Pipettes

    (Fig. 4). – Carefully unscrew the tip cone by turning the auxiliary tool clock- wise (ErgoOne 5ml tip cone can be ® removed without the auxiliary tool). Fig. 4 ErgoOne · User Manual ·...
  • Page 18 – Degrease the piston using an alcoholic disin- fectant and re-lubricate the piston with original USA Scientific piston grease (see section 8 Ordering Information). A damaged piston must be replaced. – Reassemble the pipette in the reverse order. Fig. 5 Replacing seals for models up to 1000µl...
  • Page 19 – Degrease the O-ring of the piston unit using an alcoholic disinfectant and re-lubricate it with original USA Scientific piston grease (see section 8 Ordering Information). – The pipette filter is easy to remove by pulling it out of the tip cone .
  • Page 20: Maintaining Multi-Channel Pipettes

    . This will allow you to slightly lift the ejector fixing which will release the tip ejector – Pull both clips down and rotate the lower housing through 180° (Fig. 10). Fig. 9 Fig. 10 · ErgoOne · User Manual...
  • Page 21 Fig. 11 disassembled. In the event of damage, it should be fully replaced (see section 8 Ordering Information for a replacement piston unit). – Reassemble the pipette in the reverse order. Fig. 12 Fig. 13 ErgoOne · User Manual ·...
  • Page 22: Ordering Information

    7112-3300 Green ErgoOne Accessories ® Description Cat.No. Pack Size Universal Pipette Carousel Stand for up to 8 pipettes 7200-0000 Silicone Grease for Pipette Piston 7200-0100 Tool for Single-Channel Pipettes 7200-0101 Tool for Multi-Channel Pipettes 7200-0102 · ErgoOne · User Manual...
  • Page 23 10 – 100 µl 7200-0310 7200-0321 7200-0331 30 – 300 µl 7200-0310 7200-0322 7200-0332 ErgoOne 12-Channel ® 0.5 – 10 µl 7200-0320 7200-0335 10 – 100 µl 7200-0310 7200-0321 7200-0336 30 – 300 µl 7200-0310 7200-0322 7200-0337 ErgoOne · User Manual ·...
  • Page 24: Pipette Tip Compatibility

    (1000 µl XL Filter) 10 x 50 10 x 50 Tips Tips Tips 5000 µl Graduated, 500 – 5000 µl 1050-0000 1050-0700 1050-0710 Macro When using TipOne RPT tips please recalibrate your pipette for best accuracy. · ErgoOne · User Manual...
  • Page 25 1180-1840 1180-1740 1123-1840 1123-1740 1183-1840 1183-1740 1120-8810 1120-8710 1180-8810 1180-8710 1111-0210 1111-1210 1161-1800 1161-1700 1163-1800 1163-1700 1120-9810 1120-9710 1160-9800 1160-9700 1180-9810 1180-9710 1126-7810 1126-7710 1121-1830 1122-1730 1161-1820 1161-1720 1182-1830 1182-1730 10 x 50 Tips 1050-0810 ErgoOne · User Manual ·...
  • Page 26: Warranty

    Warranty is effective for 36 months from the date of delivery when the product is used according to the instructions. Damage caused by improper use is not covered. This Limited Warranty covers, at the sole discretion of USA Scientific, repairs to or replacements of the equipment. Our “General Terms and Conditions” apply.