3.2 PRESET OTHER SPECIFICATIONS
Press MODE to enter the specification configuration interface, which
including many specification adjusting options like FILE NUMBER,
MEASURING MODE, VIEW MODE, PROBE SETUP, MINIMUM
ALARM, MAXIMUM ALARM, NORMAL THICKNESS, THE MINIMUM
OF B SCAN, THE MAXIMUM OF B SCAN, RECTIFICATION,
RECTIFICATION WAVEFORM, RESOLUTION, UPDATE RATE,
LANGUAGE, UNITS, AUTO POWER DOWN, ERASE ALL FILES AND
DEFAULT SETUP. See the following figure:
3.3 Specification Adjusting Steps
9
FILE NUMBER – Select the current file. Total 400 files and each file
could save 252 thickness values or wave-forms.
PROBE SETUP: D15X (means single crystal delay probe)
MEASURING MODE:
The Thickness Gauge has 4 measurement modes:
I-E: interface wave- Echo, measurement range 2mm-27mm;
E-E: Echo-Echo, measurement range 0.2mm-15mm;
ME-E: Multiple-wave check method, measurement range is
0.2mm-10mm. According to 3(Min. times) to 9Max.times) times echos
every time when measuring, the more measuring times, the higher
precision will get. Furthermore the reading values are all through
checkout, to make sure the results are accurate.
AUTO: Automatic mode, measurement range is from 0.2mm-27mm. On
the basis of thickness, it selects the measuring model automatically.
When H≤10mm, it adopts ME-E measuring mode; when
0.2mm<H≤5mm, it adopts E-E measuring mode; and when H>2mm, it
adopts P-E mode. Under this mode, the user only needs to adjust the
Gain. It is easy to operate, we recommend the common users to adopt
this mode. H is the thickness of the material to be measured.
VIEW MODE: normal mode, difference mode and limit scanning mode.
MINIMUM ALARM: Set the minimum thickness alarm value, range of
0.15-635mm. The result will be displayed in red if the actual thickness is
less than the minimum value preset.
MAXIMUM ALARM: Set the maximum thickness alarm value, range of
0.15-635mm. The result will be displayed in red if the actual thickness is
more than the maximum value preset.
NORMAL THICKNESS: Set the normal thickness, range of
0.15-635mm. The real concrete application will be introduced in the
difference mode.