Page 1
Zaxis PD - Quick Start Guide This document will aid in the physical connections and navigation of the setup parameters for the Zaxis-PD test platform. Physical Connections Ethernet Interface RS232 Interface Clamp Valve Leak Standard Valve NO – Normally Open Staubli QC NC –...
Page 3
On initial power-up of the unit, the following screen will appear. This screen (About) provides the serial number as well as the firmware version of the display and main PCB. After five seconds the About screen will change into the run mode (below). Pressure vs.
Configuration Mode The configuration of the unit is separated into four main menus; PROG,UNITS, CALIBRATE, and OPTIONS. All four menus have user configurable settings. See the User Guide for more detailed instructions. PROG (Program Settings) Two menus control all the parameters associated with the test process. These screens are PRESSURE, and SETTINGS.
Page 5
Settings These paramters control test air presented to the test part. The Pre-Fill is used with larger part volumes (greater than 300cc) This timer controls the valve that will introduce a second source of air to the part, typically at a higher pressure. This function can be either timed, or Auto stopped at a value entered on the Pre Fill pressure value box.
Page 6
Some amount of trial and error will be involved in finding acceptable values. However Zaxis has a few guidelines to follow in aid of your task.