Please read these operating instructions before unpacking and putting the unit into operation. Follow the instructions precisely as described herein. The instruction manuals on our website www.kobold.com are always for currently manufactured version of our products. Due to technical changes, the instruction manuals available online may not always correspond to the product version you have purchased.
The user assumes all risk for such usage. 5. General These operating instructions are valid for electrolytic conductivity cells as per data sheets ACS-Z. Electrolytic conductivity cells are used in conjunction with suitable transmitters for industrial chemical analysis, to determine the electrolytic conductivity of liquids, or the resistance in high-purity water.
5 µS/cm bzw. 20 MΩcm 0.01 bis 10 µS/cm bis 3000 µS/cm bis 15 mS/cm bis 30 m mS/cm 10.0 bis 200 mS/cm The actual range limits vary according to the electrode material, style and subsequent transmitter! ACS-Z K03/1022 page 5...
The connection cable must be run directly to the transmitter, and not via series terminals. Only use screened cables, preferably those recommended or supplied by the manufacturer. Please observe the instructions for the electrical connections in the operating instructions for the transmitter used. page 6 ACS-Z K03/1022...
(i.e. the conductive electrodes of the cell must always be completely enveloped by the liquid), and constructional measures must be taken to eliminate interruptions in the flow or gas bubbles. ACS-Z K03/1022 page 7...
Deposits can be removed using diluted hydrochloric acid, or by ultrasonic cleaning, for example. The use of electrolytic conductivity cells in strongly glutinous, oily or sticky media is not permissible. In such cases, inductive conductivity measurement should be used. page 8 ACS-Z K03/1022...
-gas bubbles necessary. The cell can also be checked for short-circuits or internal contact problems. This can be done using a continuity tester (e.g. the diode tester of a multimeter). ACS-Z K03/1022 page 9...
(Cd, Hg, Li or Pb) of the heavy metal that is decisive for the classification as containing pollutants: 1. „Cd" stands for cadmium 2. „Hg" stands for mercury 3. „Pb" stands for lead 4. „Li" stands for lithium Electrical and electronic equipment page 10 ACS-Z K03/1022...
ACS-Z 14. EU Declaration of Conformance We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Conductivity-/high-purity water cells with 2-electrode system model: ACS-Z to which this declaration relates is in conformity with the following EC guidelines:...