ULTRASOUND EQUIPMENT USER MANUAL Foreword Thank you for purchasing The -type ultrasonic diagnostic apparatus In order to make sure that you can be a safe and effective use of the instrument, please read this manual carefully. After reading, please keep this manual, so that can check at any time when you needed. This manual contains the product user manual and technical specifications.
ULTRASOUND EQUIPMENT USER MANUAL Form 1 Safety symbol table Graphics, Symbols, Words Name Meaning Used to indicate if you ignore it will result in serious Warning personal injury, death or actual damage to property. Warning Used to indicate if you ignore it will produce slight Careful personal injury or property damage.
ULTRASOUND EQUIPMENT USER MANUAL Identification of Graphics, Symbols, Abbreviations Name Meaning Position Reduce Reduce/decrease On the panel Time gain Time gain control On the panel At the back of the VIDEO OUT Output interface PAL video output panel At the back of the VGA OUT Output interface SVGA video output...
ULTRASOUND EQUIPMENT USER MANUAL Safety precautions In order to prevent the damage caused by improper use, please observe the following safety precautions. Safety precautions Warning 1. The user should according to the provisions in this manual operation, the manufacturer does not assume in violation of the instructions may cause a variety of security risks.
Page 9
ULTRASOUND EQUIPMENT USER MANUAL Safety precautions Careful 1. Should prevent the probe far away from the sharp, hard object, it is forbidden to collision probe acoustic window, otherwise it will cause damage of the probe. 2. Probe for a long time can cause fever, during the suspension diagnosis, should press the "frozen" button to make the probe to stop working.
ULTRASOUND EQUIPMENT USER MANUAL Chapter 1 Outline 1.1 Functional Characteristics This Ultrasonic diagnostic instrument (here in after referred as instrument) using 3.5 MHz electronic convex array probe, adopts full digital beam former(DBF); real-time dynamic aperture imaging (RDA); full digital dynamic receive focusing (DRF); frequency conversion; 8 segments TGC; dynamic digital filtering; dynamic digital filtering;...
3) The probe driven by the high voltage will change the electric signal into ultrasonic vibration, and the ultrasonic wave is transmitted into human body. 4) The ultrasonic wave produces the reflection signal in the body, then the probe will change it into electric signal and send to ultrasonic receiver.
ULTRASOUND EQUIPMENT USER MANUAL Monitor Digital scanning Ultrasonic conversion emission Control Ultrasonic Probe panel reception Figure 1-2 Functional Block Diagram 1.4 Field of Application and Contraindication 1.4.1 Field of Application The instrument mainly applies in the dynamic image diagnosis of liver, gallbladder, spleen, kidney, urinary bladder, pancreatic gland, as well as gynecology, obstetrics examinations, pregnancy and contraception examination.
Annotation function: hospital name, patient’s name, age and gender; Human body marks (with probe’s position); Full-screen character annotation; Real-time clock display; Puncture guide: 3.5MHz convex array probe can display puncture guide line in B mode.
ULTRASOUND EQUIPMENT USER MANUAL Permanent storage: ≧128 images(Pseudo color display: External monitor 16 kinds of color display) Image playback: series playback or check one by one. Image processing: γ correction, frame correlation, point correlation, line correlation, digital filtering, digital edge enhancement and pseudo color processing, etc. Output interface: SVGA video output offers connection to SVGA color display;...
ULTRASOUND EQUIPMENT USER MANUAL Chapter 3 Operation Panel 3.1 Operation Panel Figure The operation panel is as follows: Gain Trace ball Figure 3-1 Control Panel 3.2 Key text/graphics : enter numbers or marks ~ : enter characters or marks ~ : Space : Make notes including hospital name, name and age of patient with full-screen character : Menu selection...
Page 16
ULTRASOUND EQUIPMENT USER MANUAL : Select focus point : Freeze image : Enter, confirm : Distance measurement : Circumference/ area measurement : Heart rate measurement : Gestational weeks (gestation) measurement mode : Escape from the measurement and note : Time set :...
Page 17
ULTRASOUND EQUIPMENT USER MANUAL : Under BM mode : Move scanning beam to the left under M mode : Move scanning beam to the right under M mode : Magnifies image : Partial image magnification : Puncture guidance function : Image memory : Up : Down : Clear the measurement and the note...
Page 18
ULTRASOUND EQUIPMENT USER MANUAL : Upper character conversion : Select probe : Up and down flip : Reset the unit GAIN : Adjust the overall gain : Adjust the gain applied to the near field of the image : Adjust the gain applied to the far field of the image Brightness: Brightness control Contrast: Contrast control...
ULTRASOUND EQUIPMENT USER MANUAL Chapter 4 Instrument Installation 4.1 Open packing box Should take the instrument from the top of the packing box, after open the plastic bags, take the top handle of the instrument at upward direction. 4.2 check Should check whether there is any shipping damage on instrument or not.
Page 20
ULTRASOUND EQUIPMENT USER MANUAL Before use new probe, should tear down the protective tape gently. When take off the probe from the host, should rotate the locked handle gently to the loosen state, to pull out the probe plug. 4.4.3 External Video Equipment The video output socket may connect with the video output equipment recommended by the company.
ULTRASOUND EQUIPMENT USER MANUAL 4.5 Use of electromagnetic environment Guidance of Using the electromagnetic environment Attention . The instrument cannot be used in conjunction with the high frequency surgical equipment When using, should stay away from such as generator, X-ray equipment, radio stations and transmission lines, in order to avoid electromagnetic interference in the process of using, otherwise it will make the instrument image interference signals in diagnosis.
ULTRASOUND EQUIPMENT USER MANUAL Careful 1. Sundry, water cup, etc. are not permitted on the operation panel. It is forbidden to pull the plug before shut off the power switch. The correct use of the probe Careful Before starting up the instrument, make sure the installation of the probe is correct and reliable. Power on, but without the probe, don't touch metal parts of the probe socket on the host to prevent electric shock.
ULTRASOUND EQUIPMENT USER MANUAL Turn on the monitor power switch, and then the main engine power switch (in "|" condition), and the instrument works. The panel indicator lights on. If the screen is not bright, please press the “POWER” at the bottom of the monitor, in order to make display light, then the instrument work. After a few seconds, the screen display boot logo.
ULTRASOUND EQUIPMENT USER MANUAL alternatively (this mode works with the magnification of 0.8, 1.0, and 1.2. Press and the instrument enter the working condition with the magnification of 0.8, 1.0, 1.2. when it is with the magnification of 1.5 and 1.8 or 2.0. c) Press : The instrument works in 4B mode;...
ULTRASOUND EQUIPMENT USER MANUAL measurement; the measurement data (+D XX.X (cm)) will real-time display with the cursor; e) Press to exchange the start and end point of the measurement; f) Repeat Step d) and Step e) to carry on another measurement; g) Repeat Step a) –Step f) to fix the measurement of the first group of cursor “+”...
Page 26
ULTRASOUND EQUIPMENT USER MANUAL f) Press to connect the tracks to the start point of the measurement and the measurement data will be displayed as C: XX.X (cm) and A: XX.X (cm ) and the measurement ends; g) Repeat Step a) and Step e) to carry on another measurement; h) Press to exit from the C/A measurement;...
ULTRASOUND EQUIPMENT USER MANUAL Heart Rates Measurement 5.9 Gestational Weeks Measurement Press to select the measurement pattern of gestational weeks as BPD, GS, CRL, FL and AC, measurement pattern is highlighted on the right of the screen and it is only effective under B mode, B/B mode and 4B mode.: 5.9.1 (BPD) Bipartitely Diameter (Press...
Page 28
ULTRASOUND EQUIPMENT USER MANUAL e) Press to end the measurement; exchange the start and end point of the measurement; f) Repeat Step d) and Step e) to repeat the measurement; g) Press to exit from the gestational weeks measurement; h) Press to clear the measurement mark after exited from the measurement.
Page 29
ULTRASOUND EQUIPMENT USER MANUAL c) Press to confirm the start point of the measurement; d) Press or touch the trackball to select the end point of the measurement and the gestational weeks, expected date of confinement and the data of fetal weight will be displayed;...
Page 30
ULTRASOUND EQUIPMENT USER MANUAL XX/XX (month/date) G: XX XX±XX(Gram) e) Press to end the measurement; exchange the start and end point of the measurement; f) Repeat Step d) and Step e) to carry on another measurement; g) Press to exit from the gestational weeks measurement; h) Press to clear the measurement mark after exited from the measurement.
Page 31
ULTRASOUND EQUIPMENT USER MANUAL b) Press arrow keys or touch the trackball to select the start point of the measurement; c) Press to confirm the starting point of the measurement; d) Press arrow keys to select the end point of the measurement; and displays the gestational weeks, expected date of confinement and the data of fetal weight: HL: XX.X (cm) G.A: XX (weeks)
ULTRASOUND EQUIPMENT USER MANUAL h) After exit gestational week measurement, press to clear the measurement mark of TTD. 5.9.8 (AC) Abdominal Circumference a) Press to highlight AC on the right side; cursor “+” appears in the screen; b) Press or touch the trackball to move the cursor to the center of the fetus’...
Page 33
ULTRASOUND EQUIPMENT USER MANUAL Press continuously to enter the noting condition of ID, AGE and the full- screen character in turn when no measurement or setting up is carried on. Characters from 0-9, from A-Z, from a-z and some special characters may be input randomly at the point of the cursor. Press the direction keys or touch trace ball to move the cursor when noting characters in full-screen.
Press and the body mark will be seen at the left-bottom of the screen. The instrument has set up with Human body marks; continuously press and 64 body marks will be displayed circularly.
ULTRASOUND EQUIPMENT USER MANUAL During the setting, press , it means exit setting, the date is invalid. 5.14 Setting of Focus Point The instrument has four sections of electron focusing, includes section 1, section2, section 3, and section4 when focusing: a) Press : FO: 1 2 3 4 shows on right flank of the screen demonstrates.
ULTRASOUND EQUIPMENT USER MANUAL Press to freeze the image or return to real time image at any time. When image is frozen, the “Freeze” mark will display at the right-bottom of the screen. User is suggested to make the instrument in freeze mode before checking when starting up. 5.17 Menu Function Selection Press to enter the menu function, and selections show at the bottom of the screen.
ULTRASOUND EQUIPMENT USER MANUAL the trackball to select and review the image one by one. Cine loop is not available in BM or M mode. 5.21 Image Memory and Display When image is frozen, press , a numeral appears at the top right corner of the screen, press direction key to change the numeral.
ULTRASOUND EQUIPMENT USER MANUAL When need to scan on real time, press dynamic range increase; press dynamic range reduced. ←, →, Dynamic range size according to the information is showed on the right side of the screen area. 5.25 Gain Adjustment Gain control including 8 period of the TGC, Gain (Gain) Adjusting the TGC sliding handle, can adjust paragraphs respectively image gain strength;...
ULTRASOUND EQUIPMENT USER MANUAL Chapter 6 Simple Maintenance 6.1 Attention In the maintenance of the instrument, please observe the following precautions: Preventive maintenance and matters needing attention Warning 1. Before cleaning the host and probe appearance, in order to avoid the accident, should unplug the power cord is completely cut off power supply net.
Page 40
ULTRASOUND EQUIPMENT USER MANUAL Chapter 7 Fault Checking and Ruled Out 7.1 Fault Checking Need Attention When Fault check in, please observe the following precautions: Fault Checking Need Attention Warning According to this chapter, if the problem can not be solved successfully, please send back the instrument to repair.
Page 41
ULTRASOUND EQUIPMENT USER MANUAL Chapter 8 Transportation and Storage 8.1 Transportation and storage attention During transportation and storage, please observe the following precautions: Transportation and storage attention Warning 1. Before equipment during transport or storage, the probe should be removed to a safe place. During the transportation and storage, it is forbidden to inversion or horizontal position.