Page 1
This instrument is manufactured in the European Union.The Company reserves the right to change the specification or design without prior notice. MEGGER is the registered Trade Mark of AVO INTERNATIONAL LIMITED. This data uses the comma as the decimal marker to align with general European usage.
Page 2
® MEGGER FlexiClamp 200 User Guide SAFETY WARNINGS • The section of wire under test must be clean, free of debris, and undamaged. • Measurements beyond maximum selected range must not be taken. • When taking measurements above 30 V RMS, extreme care must be taken, with the hand kept behind the hand guard.
Page 3
When measuring the current in a flat cable, the cable must be positioned as follows: The proximity of other current carrying conductors may affect the reading. Keeping a minimum distance of 125mm (5”) from these conductors will minimise this. Read the current directly from the display. The ‘Hold’ button holds the reading on all ranges.
Page 4
Specification Measuring Ranges: Single wire mode 0 to 199,9 Amp 2-or-3-wire mode 0 to 40,0 Amp Resolution: 0,1 Amp. Accuracy: At 22˚C (72˚F) and rated frequency Single wire 2% of reading ± 2 digits 2-or-3-wire mode Flat or Twin cable 5% of reading ±...