ABB AZ200 770 Manual
ABB AZ200 770 Manual

ABB AZ200 770 Manual

Endura az20 oxygen monitor pumped reference air unit

Advertisement

Quick Links

Information INF09/027 Rev. D
Endura AZ20 oxygen monitor
Pumped reference air unit
1 Introduction
This
publication
details
maintenance of an ABB Pumped Reference Air Unit. This pump
supplies reference air to any of the following ABB oxygen probes:
AZ20, AZ25, AZ20/ZFG2 replacement, ZFG2 and ZGP2
Note. The pump cannot be used with multiple probes
simultaneously.
4 pump models are available – see Table 1.1:
Description
1
/
BSP (Metric) 230 V AC 50 / 60 Hz
4
1
/
BSP (Metric) 115 V AC 50 / 60 Hz
4
1
/
NPT (Imperial) 230 V AC 50 / 60 Hz
4
1
/
NPT (Imperial) 115 V AC 50 / 60 Hz
4
Table 1.1 ABB Pumped Reference Air Unit Part Numbers
The internal air filter is a replaceable item (ABB part
number 0217463).
the
installation,
connection
ABB Part Number
AZ200 770
AZ200 771
AZ200 772
AZ200 773
2 Safety
and
Information in this document is intended only to assist our
customers in the efficient operation of our equipment. Use of this
manual for any other purpose is specifically prohibited and its
contents are not to be reproduced in full or part without prior
approval of the Technical Publications Department.
2.1 Safety precautions
Please read the entire document before unpacking, setting up, or
operating this pump.
Pay particular attention to all warning and caution statements.
Failure to do so could result in serious injury to the operator or
damage to the equipment.
To ensure the protection provided by this equipment is not
impaired, do not use or install this equipment in any manner other
than that which is specified in this manual.
2.2 Health & Safety
To ensure that our products are safe and without risk to health,
the following points must be noted:
The relevant sections of these instructions must be read
carefully before proceeding.
Warning labels on containers and packages must be
observed.
Installation, operation, maintenance and servicing must
only be carried out by suitably trained personnel and in
accordance with the information given.
Normal safety precautions must be taken to avoid the
possibility of an accident occurring when operating in
conditions of high pressure and / or temperature.
Safety advice concerning the use of the equipment described
in this manual or any relevant Material Safety Data Sheets
(where applicable) may be obtained from the Company address
on the back cover, together with servicing and spares
information.

Advertisement

Table of Contents
loading

Summary of Contents for ABB AZ200 770

  • Page 1 ABB Pumped Reference Air Unit. This pump customers in the efficient operation of our equipment. Use of this supplies reference air to any of the following ABB oxygen probes: manual for any other purpose is specifically prohibited and its...
  • Page 2: Restriction Of Hazardous Substances (Rohs)

    Protective earth (ground) terminal. Note. The following only applies to European customers. Functional earth (ground) terminal. ABB is committed to ensuring that the risk of any Direct current supply only. environmental damage or pollution caused by any of its products is minimized as far as possible. The Alternating current supply only.
  • Page 3: Safety Recommendations

    Only personnel herein be scrupulously respected. If danger warnings are not from ABB or its approved representative(s) is (are) authorized to heeded to, serious material or bodily injury could occur. attempt repairs to the system and only components formally approved by the manufacturer should be used.
  • Page 4: Installation

    3 Installation 3 Installation 3.3 Electrical safety 3.1 Siting Warning. Site the pump adjacent to the probe – see Fig. 3.1 for environmental  Isolate the incoming power supply before accessing the requirements. pump or making connections.  The pump is not fitted with a switch therefore a 55 °C disconnecting device such as a switch or circuit breaker (131 °F)
  • Page 5: Electrical Connections

    3 Installation 3.4 Accessing the terminal block 3.5 Electrical connections Referring to Fig. 3.3: 3.5.1 Preparation Referring to Fig. 3.4: 1. Loosen the 4 pump cover retaining screws 1. Prepare the incoming power supply cable as follows: 2. Remove the pump cover a.
  • Page 6 3 Installation 3.5.2 Making connections 3.6 Reference air connections Referring to Fig. 3.4, make mains connections as detailed in steps Referring to Fig. 3.5: 1 to 5. The mains power supply must be 115 or 230 V AC, 50 / 60 1.
  • Page 7: Maintenance

    4 Maintenance 4 Maintenance 4.1 Replacing the air filter Note. Isolate the pump from power and air supplies before replacing the air filter. 1. Remove the pump cover – see section 3.4, page 5. Referring to Fig. 4.1: 2. Disconnect the top tube from air filter 3.
  • Page 8 Oldends Lane notice. With regard to purchase orders, the agreed Stonehouse particulars shall prevail. ABB does not accept any Gloucestershire GL10 3TA responsibility whatsoever for potential errors or possible lack of information in this document.