Page 2
Customer service 1. Install LP Dashboard from USB supplied with level transmitter or Tel: +1-800-633-7609 download from www.temposonics.com. Steps 1 through 11 can Fax: +1-800-498-4442 be done from the control room or any safe area and do not need E-mail: info.us@temposonics.com to be done on the tank.
Page 3
Level Plus Board Replacement ® Operation Manual 3. Open LP Dashboard and connect to level transmitter. Select the 5. Save a backup file of the settings for the level transmitter by correct Protocol from the drop-down menu. Select the correct clicking Save Settings to a File.
Page 4
Level Plus Board Replacement ® Operation Manual 10. If the customer supplied backup file is not used then you will 13. Pull out display. The display is connected with a cable. The need to set the address. To set the address click the three white display can hang from the housing by the cable or it can be bars and select Configuration.
Page 5
Level Plus Board Replacement ® Operation Manual 15. Remove the top board from the board stack. The top board will 17. Disconnect sensing element and temperature cables from the remain with the housing as the white ground wire is connected to board.
Page 6
Level Plus Board Replacement ® Operation Manual 20. Replace the ground lug, white ground wire, and top board. 21. Connect display to middle board (analog) or bottom board (digital). Fig. 15: Ground Lug Removed 22. Connect sensing element and temperature cable. The gold pins on the temperature cable should face down towards the board.