Intelligent hart temperature transmitter (28 pages)
Summary of Contents for Microcyber NCS-TT106H-R1
Page 1
NCS-TT106H-R1 HART DIN Rail Temperature Transmitter User Manual...
Page 2
Warning 1. It is forbidden for users to disassemble the temperature module by themselves. 2. Please check if the supply voltage of temperature transmitter meets the power supply voltage requirements in the manual...
NCS-TT106H-R1 uses digital technology, so it can connect with many types of thermocouple and thermo resistive sensors. It has wide range and simple interface between field and control room, which reduces the expense of installation, operation and maintenance.
2. INSTALLATION 2.1 Dimension Figure 2.1 Dimension of Temperature Transmitter (Unit: mm) 2.2 Wiring Figure 2.3 Wiring HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 6
The shielded wires of bus ends shall be connected to the ground. HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
1) Access to the upper level control system through the AI module and HART communication device; 2) Mixed use of analog and digital communication methods; 3) The device address is 0. HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
The following is an example of Baldota's HartMPT configuration software to introduce the configuration method of the intelligent temperature transmitter. It mainly includes the following functions: 1) Basic information configuration: configure the basic HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 9
ID, manufacturer, device type, device ID. , long address and version information, as shown in Figure 3.3. HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 10
6) The maximum length of the assembly number is 6 specified characters. ⚫ Configuration information configuration The smart transmitter configuration information can be read or modified via the configuration information tab. HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 11
When modifying the unit, you cannot modify the upper and lower limits of the main variable range at the same time, and should be modified separately. HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 12
The sensor information tab allows you to view the currently configured sensor information (upper, lower, and minimum spans) and to configure each sensor type, line, and other information. As HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 14
Channel status: Display sensor channel status (open circuit, short circuit & etc.). Sensor channel value: Display the original value of sensor channel. Cold end temperature: Display the current cold end HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 15
⚫ Current Calibration The current calibration steps are as follow: HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 16
When the main variable is higher than the upper limit, the HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 17
Currently, the refreshed variables are PV value, current value, percentage and cold end temperature. HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 18
Figure 5.2 Variable monitoring HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Check the cable is away from the strong electromagnetic Reading Error interference Software Configuration Check sensor type configuration Check function block parameter configuration Software Configuration Check sensor type configuration Check function block parameter configuration HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Terminal IP00, housing IP20 Voltage effect ±0.005%/V Alarm output can be set as 3.5…23 mA Alarm signal Upper/lower limit current 20.8mA / 3.8mA Current 0.03% accuracy HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Sensor Type B,E,J,N,K,R,S,T;-100mV~+100mV Common Mode ≥70dB(50Hz and 60Hz) Rejection Series Mode ≥70dB(50Hz and 60Hz) Rejection Temperature <50ppm/℃ Effect 1.4 Physical Parameter Dimension 22.5*110*99mm Housing Nylon Material HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Page 23
YOUR FIELDBUS EXPERT CONTACT INFORMATION Address: 17-8 Wensu Street, Hunnan New District, Shenyang, China Website: www.microcyber-fieldbus.com Phone: +86-24-31217278/+86-24-31217280 Fax: +86-24-31217338 Email: fang.siqi@microcybe.cn HART DIN Rail Temperature Transmitter | NCS TT106H R1 | Page Email | fang.siqi@microcyber.cn...
Need help?
Do you have a question about the NCS-TT106H-R1 and is the answer not in the manual?