Alarms and Safety Features
Set Insertion
The set is not inserted correctly.
The wrong set used.
Remedy
Check the positioning.
Clean the clamp.
Use an Ambix activ giving set.
Pump Mechanism Blocked
Contact Calea Patient Services or the helpline.
Downstream Occlusion
The infusion will stop.
Check below the pump for the following:
Kink/blockage/clamps/air particles.
Remedy
Remove the kink/blockages/clamps/air particles.
Upstream Occlusion
The infusion will stop.
Check above the pump for the following:
Kink/blockage/clamps/air particles.
Remedy
Remove the kink/blockages/clamps/air particles.
Ambix Activ Ambulatory Pump and Accessories -
Quick Reference Guide
Line Empty
The bag is empty.
The set is not fully primed.
The air sensor area is contaminated.
Remedy
Stop the infusion, change, and check the bag and set.
Prime the set.
Clean.
Batteries are not charging properly
The pump is plugged in at the mains.
Remedy
Contact Calea Patient Services or the helpline.
Batteries are low
The battery power will last for 30 minutes.
Remedy
Plug the pump in to the mains.
Battery is depleted
There is no external power available.
Remedy
Check the installation in the holder.
Check the contact pins and the pump.
Check the cable and connect to the mains.
Plug into the mains to complete.
Door Open
The pump door or lever is open.
The pump door is installed incorrectly.
The pump door mechanism is defective.
Remedy
Close the pump door and lever.
Remove the door and re-hang.
Contact Calea Patient Services or the helpline.
Freephone 24 hour Advice Line
0800 121 8300
Quick Reference Guide
Ambix Activ
Ambulatory Pump & Accessories
Freephone 24 hour Advice Line
CAL099
0800 121 8300

Advertisement

loading

Summary of Contents for Calea Ambix Activ

  • Page 1 Ambix Activ Ambulatory Pump and Accessories - Quick Reference Guide Line Empty Alarms and Safety Features The bag is empty. The set is not fully primed. The air sensor area is contaminated. Remedy Set Insertion The set is not inserted correctly.
  • Page 2 Set target volume activ pump. Make sure the foot of the clamp is facing outwards. Position the top of the blue section on the Ambix activ giving set into the blue section on the Ambix activ pump. Close the pump door.