Our Solderability Test Systems are designed to provide fast, accurate and
objective measurement of solderability for components, including SMD and
printed circuit boards The system's output results are expressed in quantitative
units – capillary wetting forces (nM) and wetting meniscus angle (degrees). The
results directly correspond to the solder wettability of the specimen. The
Metronelec solderability systems advance the state of the art of solderability
testing by eliminating the need to use indirect comparisons or subjective
operator interpretations.
Meets requirement of all international standards:
IPC/EIA J-STD 002 1 003
JEDEC MIL 883 L
IEC 68-2-69
IEC 68-2-58
NFC 89-400
User's standard definition
Specifications
Dimensions:
Immersion depth:
Globule size:
Languages:
Force measurement range:
Force measurement accuracy: 0.001mN
Version 2.0
USER MANUAL
SOLDERABILITY TESTER
w310 x l500 x h720
0.01-15mm by step of 0.01mm
1, 2, 3.2, 4 mm
English, French, German, Chinese
0.1 – 40 mN
01/10/2021
V2.0 – 10/2021
Page 1 sur 83

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MENISCO ST88 NEO and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Metronelec MENISCO ST88 NEO

  • Page 1 – capillary wetting forces (nM) and wetting meniscus angle (degrees). The results directly correspond to the solder wettability of the specimen. The Metronelec solderability systems advance the state of the art of solderability testing by eliminating the need to use indirect comparisons or subjective operator interpretations.
  • Page 2: Table Of Contents

    IEC 68-2-69......................14 IPC/ANSI/J-STD-002/3 ................... 14 User’s specification ....................14 Utilization ........................15 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 2 sur 83...
  • Page 3 Validate / Invalidate / Open / Print / Delete / Average / Merge ........36 Options ........................37 Configuration......................37 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 3 sur 83...
  • Page 4 TYPE 28 : - SAMPLE (PBRC/AVX-4M) ................73 TYPE 29 : - CHIPS FILM (LCC) ................... 74 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 4 sur 83...
  • Page 5 Solder globule ....................... 81 Annex A ........................81 PCB wetting test ....................... 81 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 5 sur 83...
  • Page 6: Introduction

    The angle 8 formed by the surface of the solid, gaseous (vapour) and the liquid at their extreme point of contact is the wetting angle. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 6 sur 83...
  • Page 7: Balance Measurement

    Before, when the meniscus is moving (climbing) the values of the force are uncertain. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 7 sur 83...
  • Page 8: Wetting Force

    In the general case (component fully wettable, regular shape on “z” axis), METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 8 sur 83...
  • Page 9: Wetting Force Divided By Perimeter

    • Wave soldering with liquid flux • Solder cream reflow with pasty flux METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 9 sur 83...
  • Page 10: Solderability Testing

    Chemicals suppliers for pcb or component fabrication and assembly. • Pcb and component manufacturers. • Assembly companies • Laboratories and schools METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 10 sur 83...
  • Page 11: Dip & Look Test Method

    If the shape is simple, regular in the “z” axis or known in the library, we can apply a model to define the quality of wetting following the force measured. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 11 sur 83...
  • Page 12: Substrates

    The sanction is the value of the wetting angle obtained with the following test conditions : Ageing : not mandatory METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 12 sur 83...
  • Page 13: Mil - Std - 883C Method 2022

    This standard is a US military specification, designed for traditional components like DIL cases. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 13 sur 83...
  • Page 14: Iec 68-2-69

    The user can define is own specification pass / fail criteria in time, level of force, time to buoyancy, etc all criteria defined in standards. See section Utilization. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 14 sur 83...
  • Page 15: Utilization

    Menis.exe file and change the last line from 0 to 1 as shown on the image below METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 15 sur 83...
  • Page 16 With the mouse click on the Menisco icon and you must obtain the following screen : METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 16 sur 83...
  • Page 17: User Level

    Use the arrow keys or the mouse to select the items wanted corresponding to the type of test to do. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 17 sur 83...
  • Page 18: Definition Of Parameters

    METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 18 sur 83...
  • Page 19: New File

    Select the directory and file to open on your hard disk or on a accessible network drive. Confirm by enter or Open button. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 19 sur 83...
  • Page 20: Parameters

    Before to start test, check the parameters definition by opening the Parameters Menu and Next Measure Parameters. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 20 sur 83...
  • Page 21 In this case you have to select in the main software the action : Measure / relancer le module MC. Please see the two following windows : METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 21 sur 83...
  • Page 22 If you still have the problem of starting MC, try to shut down the machine and turn it back on after a while. Solder globule receptacle METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 22 sur 83...
  • Page 23 Midle activated (RMA type 415 mN/m Activated (RA type) 370 mN/m No clean 425 mN/m METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 23 sur 83...
  • Page 24 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 24 sur 83...
  • Page 25 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 25 sur 83...
  • Page 26: Measuring Solderability Of Component / Pcb

    For “classical” components, use the list of type in section 4 and 5, operator will find an answer to the majority of needs. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 26 sur 83...
  • Page 27: Specific Components

    The video recording is accessible from the tool bar or from the video menu and the hot key F11 to start a video recording. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 27 sur 83...
  • Page 28: Video Capture

    240 the default value up to 640 x 480, and vertical frequency. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 28 sur 83...
  • Page 29: Play Video

    The software can use the different compression tools available on your system, operator can adjust some settings following the compression chosen. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 29 sur 83...
  • Page 30: Measuring The Gamma Lv

    The results are expressed in mN; Mil, IEC and IPC standard are not available. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 30 sur 83...
  • Page 31: Dip & Look Test

    Parameters menu : METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 31 sur 83...
  • Page 32 50.8 The immersion time is from 1 to 30 s (step 1 s) METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 32 sur 83...
  • Page 33: Start Measure

    : METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 33 sur 83...
  • Page 34: Analysis Of Results

    (if opened!). METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 34 sur 83...
  • Page 35: Apply Standard

    In case of average of curves, the table of values include also the "standard deviation". METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 35 sur 83...
  • Page 36: Zoom / Scale (Force, Angle, Time)

    Average : Make an average curve of selected curves, calculate the average value and the standard deviation in the table of values. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 36 sur 83...
  • Page 37: Options

    Also if the printer is color printer, select "Print on color" in the window. If the printer is black and white don’t select the "Print on color" in the window. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 37 sur 83...
  • Page 38 Print on color : Appear to be “has been” today but still necessary for old systems non compatible ! METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 38 sur 83...
  • Page 39 Crossing time is 0.26 s between the curve and the buoyancy level (discontinuous pink line) METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 39 sur 83...
  • Page 40 (minimum known after this maximum), so the dewetting is equal to (Fmax – METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 40 sur 83...
  • Page 41: Other Spec (Specification)

    Time : T1 and T2 times for wetting angle Theta1 and theta 2 and for force. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 41 sur 83...
  • Page 42: Cpk

    This software supports different languages, we can switch easily by the menu Options Language. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 42 sur 83...
  • Page 43: Change Password

    Print Menu: Definition of the printer to use, must be defined under the operating system. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 43 sur 83...
  • Page 44: Printing Result In Table

    Fr max, TF1, TF2, F1, F2, Dew, and Comment. Click on « OK ». Then click on « File/Print ». METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 44 sur 83...
  • Page 45: Component Types For Bath Module

    28 - SAMPLE (PBRC/AVX - 4 M) N° 28 29 - CHIPS FILM (LCC) N° 29 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 45 sur 83...
  • Page 46: Type 1 : - Rectangular Sample

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 46 sur 83...
  • Page 47: Type 2 : - Cylindric Sample

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 47 sur 83...
  • Page 48: Type 3 : - Multi Rect. (Angular) Leads

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 48 sur 83...
  • Page 49: Type 4 : - Multi Cylindrical Leads

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 49 sur 83...
  • Page 50: Type 5 : - Plated Substrate Single Side

    Dip the extremity (epoxy and metal) of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 50 sur 83...
  • Page 51: Type 6 : - Plated Substrate Double Side

    Dip the extremity (epoxy and metal) of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 51 sur 83...
  • Page 52: Type 7 : - Chips Capacitor

    Dip the extremity of chip in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 52 sur 83...
  • Page 53: Type 8 : - Chips Resistor

    Dip the extremity of chip in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 53 sur 83...
  • Page 54: Type 9 : - Particular Lc04/04 Case

    Dip the extremity (package and metal) of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 54 sur 83...
  • Page 55: Type 10 : - Sot 23

    Dip the extremity of lead in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 55 sur 83...
  • Page 56: Type 11 : - Sot 143

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 56 sur 83...
  • Page 57: Type 12 : - Flat Pack, Mqfp100, Pqfp, So

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 57 sur 83...
  • Page 58: Type 13 : - Sot 89

    Dip the extremity of lead in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 58 sur 83...
  • Page 59: Type 14 : - Dpack Package

    Dip the extremity of lead in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 59 sur 83...
  • Page 60 Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 60 sur 83...
  • Page 61: Type 16 : - Jlcc, Plcc, Cqpj, Soj

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 61 sur 83...
  • Page 62 Dip the extremity of chip (ceramic and metal) in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 62 sur 83...
  • Page 63: Type 18 : - Dupont Connector Solder Resist

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 63 sur 83...
  • Page 64: Type 19 : - Pc Board Or Component Partially Wettable

    Dip the extremity of sample (epoxy and metal) in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 64 sur 83...
  • Page 65: Type 20 : - Diode Do4

    Dip the flat extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 65 sur 83...
  • Page 66: Type 21 : - Diode Do5

    Dip the flat extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 66 sur 83...
  • Page 67: Type 22 : - Chips Solid Tantale A

    Dip the extremity (package and metal) of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 67 sur 83...
  • Page 68: Type 23 : - Chips Solid Tantale B (Obsolete)

    Dip the extremity (package and metal) of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 68 sur 83...
  • Page 69: Type 24 : - Chips Solid Tantale C

    Dip the extremity (package and metal) of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 69 sur 83...
  • Page 70: Type 25 : -Solder Paste

    USER MANUAL SOLDERABILITY TESTER TYPE 25 : -SOLDER PASTE (Measure not available) METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 70 sur 83...
  • Page 71: Type 26 : - Cqpl, Mqfp>100

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper, . METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 71 sur 83...
  • Page 72: Type 27 : - Pcb (Test Pad & Plated Hole)

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 72 sur 83...
  • Page 73: Type 28 : - Sample (Pbrc/Avx-4M)

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 73 sur 83...
  • Page 74: Type 29 : - Chips Film (Lcc)

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 74 sur 83...
  • Page 75: Component Types For Globule

    02 - cylindrical sample N° 02 03 – complex shape sample N° 03 METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 75 sur 83...
  • Page 76: Type 1 : - Rectangular Sample

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 76 sur 83...
  • Page 77: Type 2 : - Cylindric Sample

    Dip the extremity of sample in the flux, and remove the drop on the absorbent paper. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 77 sur 83...
  • Page 78: Type 3 : - Complex Shape Sample

    Dip the extremity of leads in the flux, and remove the drop on the absorbent paper METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 78 sur 83...
  • Page 79: Servicing

    • Video cam (optional) Installation The startup is done under the METRONELEC's responsibility or one of the authorized agent's. Work station environment To avoid the disturbing of the signal during the measurement, the apparatus shall be on a stable working bench top, free of vibrations, away from air draughts and noises.
  • Page 80: Hardware Requirements

    Pentium2 with Windows 2000 mini VGA Screen mini Serial port 256 Mb RAM memory METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 80 sur 83...
  • Page 81: Installation Of The Software

    USER MANUAL SOLDERABILITY TESTER Installation of the software The first installation is made by METRONELEC or by the official sales representatives. The updating of the software is made by the users, according to the following procedure : 1. Start WINDOWS software 2.
  • Page 82 As the red area is greater than the green one, it is fully clear than in this case the result will be negative ! but the test result is Ok. METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 82 sur 83...
  • Page 83 “complex” you need to input by yourself the total perimeter wettable “p” called in this type “sigma pcb str width” METRONELEC V2.0 – 10/2021 6 Avenue Eiffel - 78420 Carrières-sur-Seine Tél : 01 30 15 20 00 Web : http://www.metronelec.com| Mail : contact@metronelec.com Page 83 sur 83...