Service Manual ® QUICK 100 M Docking Station Made in Switzerland ARGUS Medical AG, CH-3627 Heimberg / Switzerland (A member of the CODAN group) 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 1 / 22...
1. INTRODUCTION 1.1. General IMPORTANT! This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ® QUICK 100 M docking station. The service manual is meant to be used together with the user manual.
2. BARCODE READER AND LABEL DATA 2.1. General The barcode reader "Scan Plus 1800" (Intermec) including a gender changer may be ordered from ARGUS Medical AG (Art. no. 90.150). It should be ® connected to the appropriate plug of the QUICK 100 M docking station marked with the sign.
CONFIGURATIONS 3. CONFIGURATIONS 3.1. Configuration of the docking station A special software tool for the configuration of the docking station will be released soon. Please contact your local distributor or ARGUS Medical AG. 3.2. Pump configuration CAUTION! ® Every pump used with the QUICK...
4.1.2. Requirements for a software update If you want to update an ARGUS Medical device please be sure you have the following items available: 1. Standard RS232 null modem serial interface cable, connected to the barcode interface.
Page 7
4.1.5. How to enter the flash mode of the docking station To enter the boot loader mode of the device please follow the procedure listed below: 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 7 / 22...
Page 8
7. Follow the steps described in the next chapters based on the kind of update you want to perform. ARGUS Medical AG CH-3627 Heimberg Bootloader V0.21 20030328 (C) AM Loader for flashing ARGUS Medical devices. 1. Erase user program. 2. Update user program. 3. Erase bootloader. 4. Update bootloader. 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ...
This part will take some time until the entire data is flashed. ® The QUICK 100 M docking station will flash the red LED as long as it is re- ceiving data from the PC. 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 9 / 22...
If an error occurred during the update process the device will en- ter the fail safe state (fault codes please refer to chapter 7.2). Please refer to your local distributor or ARGUS Medical AG to determine the latest software release able to run on your device hardware.
In case the docking station detects a technical failure, the last event line registered includes the failure number (fault code) which generated that technical failure. Please refer to chapter 7.2 for the complete list. 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 11 / 22...
= 00 Place 3 = No pump Status = 00 Place 4 = No pump Status = 00 Place 5 = No pump Status = 00 Fail = None 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 12 / 22...
Invalid barcode programming at pump 4 Pump in place 3 has failure Pump 5 programmed by barcode Pump in place 4 has failure Invalid barcode programming at pump 5 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 13 / 22...
Displaying of the rele- vant data. Select next address in the docking station The protocol syntax of the status transmitted is part of the next chapters. 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 14 / 22...
Page 15
P1...P5 : Word boundaries for each pump place (16 Bit) : Docking Station status (8 bit) : Docking Station software release (24Bit) : Checksum CRC (16 Bit, not implemented yet) : End of text (0x03) 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 15 / 22...
Page 16
6.2.3. Pump status The transmitted pump status will be equal to the one described in the pump ser- vice manual. An example of an ARGUS 600 D syringe pump status is shown below: 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ...
MAINTENANCE AND FAULT FINDING 7. MAINTENANCE AND FAULT FINDING 7.1. Maintenance CAUTION! Only authorized persons who have been instructed by ARGUS Medical AG or by ® the trained local distributor are allowed to service the QUICK 100 M docking sta- tion.
8.2. Replacement parts CAUTION! ® The QUICK 100 M docking station may only be used with accessories and spare parts which have been approved by ARGUS Medical AG for safe technical use. 10.114 Cover 11.215 Casing 10.096 Voltage regulator 11.220 Centring fixation 14.147.B_QUICK100M en.SM.V1.00.doc...
Page 19
10.097 Transformer 230V 10.115 Main board Quick 100M 12.043 Gender Charger 12.041 Interface cable 10.130 Barcode holder with clamp 10.087 Combination clamp 90.041 Rail set (25x10mm) 90.042 Rail set (35x 8mm) 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 19 / 22...
(30.022.A) Docking place 4 Voltage regulator Docking place 3 Operation LED Docking place 2 Main-Board_1 (30.020.A) Trafo Docking place 1 AC Main Barcode Reader Interface Nurse Call RS 232 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 20 / 22...
- Barcode reader interface ? - Staff alerting system ? 6 Electrical test according to EN60601-1: - Measurements attached? Do not install any pump! All measurements made with a 2.5 m power cord 14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 22 / 22...