ConMed ClearView 60-8585-120 Operator's Manual

ConMed ClearView 60-8585-120 Operator's Manual

Smoke evacuation system

Advertisement

ClearView
Smoke Evacuation System
Operator's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ClearView 60-8585-120 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ConMed ClearView 60-8585-120

  • Page 1 ClearView ™ Smoke Evacuation System Operator’s Manual...
  • Page 2 For a period of one (1) year following the date of delivery, CONMED warrants the ClearView™ against any defects in material or workmanship. CONMED will repair or replace (at CONMED’S option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by CONMED.
  • Page 3: Table Of Contents

    Specifications MAINTENANCE General Maintenance Information Cleaning Periodic Inspection Troubleshooting CUSTOMER SERVICE Equipment Return Ordering Information TERMS & WARRANTY Figure Title Page Control Panel © 2015 All rights reserved. Page 3 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 4: System Description

    The third stage filtration uses the highest grade virgin activated carbon, especially designed for CONMED for the removal and adsorption of odors and toxic gases produced by burning tissues. These harmful gases may constitute a health hazard to healthcare professionals who are subjected to prolonged exposure.
  • Page 5: Operational Information

    8. Upon request, CONMED will provide the following: Service and Repair Instructions, including Circuit Diagrams and Parts List 9. The fuses on the circuit board are to be serviced by an authorized CONMED techni- cian as follows: 100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo) 220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo) ©...
  • Page 6: Cautions And Warnings

    10. The fuses on the motor curcuit are to be serviced by an authorized CONMED technician as follows: 220/240 VAC, 50/60 Hz use 3,15 Amp 250 Volt Fuse (Fast-Acting), (F3) 11. This equipment needs special precautions regarding ElectroMagnetic Compatibility and needs to be installed according to EMC information found in this manual.
  • Page 7 ClearView™ or shielding the location. • The use of ACCESSORIES other than those specified by CONMED, or sold by CONMED as replacement parts for internal components, may result in increased emissions or decreased immunity of the ClearView™...
  • Page 8 • Do not block either the tubing or the filter. If either becomes occluded or significantly restricted, the motor/blower may overheat and cause the unit to fail. • Using any other filter or accessory not supplied by CONMED may cause damage and/ or cause the system to be inoperable and may void the warranty.
  • Page 9 There are no user serviceable components in the ClearView™ Smoke Evacuation System(s). Refer service to qualified service personnel. Use only with the power cord provided and always plug into a grounded outlet. © 2015 All rights reserved. Page 9 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 10 PROTECTIVE EARTH, (GROUND) EQUIPOTENTIALITY DENOTES THE DATE THE EQUIPMENT WAS MANUFACTURED DENOTES THE MANUFACTURER OF THE DEVICE NON-IONIZING RADIATION CONSULT INSTRUCTIONS REMOTE ACTIVATOR © 2015 All rights reserved. Page 10 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 11: Operating Instructions

    A footswitch or a Automatic Activation Device may be added to any system by simply plugging in a CONMED activation accessory into the appropriate jack on the front or back of the unit. When the footswitch is plugged in, the unit may be turned on or off by depressing the footswitch pedal once for each operation.
  • Page 12 If the system is overheated or if there is an electrical surge in the electrical system, fuses will break and the system will not operate. When the Service light illuminates, please contact CONMED Customer Service for system service instructions.
  • Page 13: Filter Instructions

    Note: Before installing or removing any filter, be sure that the system is turned off. Filter Installation Instructions: The installation of the Filter into the CONMED ClearView™ Smoke Evacuation System(s) is quick and simple. 1. Remove the Filter from the shipping box and discard any protective wrapping. Examine all filters for damage during shipping and storage.
  • Page 14 Figure 1 Unit Control Panel © 2015 All rights reserved. Page 14 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 15 Motor Watts 1000 ± 10% Motor Static Suction kPa (6.5 mm orifice) 21.04 *For reference purposes only **Using a new 7/8 in tubing. © 2015 All rights reserved. Page 15 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 16 Harmonic Class A Not applicable. Emissions IEC 61000-3-2 © 2015 All rights reserved. Page 16 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 17 Power frequency magnetic fields should (50/60 Hz) be at levels characteristic of a typical magnetic field location in a typical commercial or hospital environment. IEC 61000-4-8 © 2015 All rights reserved. Page 17 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 18 If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model ClearView™. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. © 2015 All rights reserved. Page 18 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 19 NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structure, objects, and people. © 2015 All rights reserved. Page 19 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 20: Maintenance

    Section 3.0 General Maintenance Information This section contains information for ordinary upkeep of the CONMED ClearView™ Smoke Evacuation System. While the system has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed by a qualified Biomedical Technician to ensure continued safe and effective operation.
  • Page 21: Customer Service

    • Automatic Activation Device • Fluid Trap • Hoses & Tubing • Reducer Fittings • Electrosurgical Pencil Adapters • Electrosurgical Smoke Plume Pencils © 2015 All rights reserved. Page 21 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 22: Terms & Warranty

    2. Incorrectly shipped merchandise is kind relating to the system and its use. exempt from restocking fees. Please contact CONMED’s Customer Service before shipping back merchandise. © 2015 All rights reserved. Page 22 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 23 Purchaser hereby consents to the jurisdiction of the New York Courts with respect to any controversy or dispute arising out of this agreement or the merchandise sold hereunder. © 2015 All rights reserved. Page 23 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...
  • Page 24 525 French Road, Utica, NY 13502 USA 1-800-448-6506 (U.S. Only) or +1 (315) 797-8375 (International) www.conmed.com Manufactured by: Buffalo Filter LLC 5900 Genesee Street Lancaster, NY 14086 © 2015 All rights reserved. Page 24 904568 Rev A CONMED ClearView™ Smoke Evacuation System Operator’s Manual...