ClearView ™ Smoke Evacuation System Operator’s Manual...
Page 2
For a period of one (1) year following the date of delivery, CONMED warrants the ClearView™ against any defects in material or workmanship. CONMED will repair or replace (at CONMED’S option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by CONMED.
The third stage filtration uses the highest grade virgin activated carbon, especially designed for CONMED for the removal and adsorption of odors and toxic gases produced by burning tissues. These harmful gases may constitute a health hazard to healthcare professionals who are subjected to prolonged exposure.
10. The fuses on the motor curcuit are to be serviced by an authorized CONMED technician as follows: 220/240 VAC, 50/60 Hz use 3,15 Amp 250 Volt Fuse (Fast-Acting), (F3) 11. This equipment needs special precautions regarding ElectroMagnetic Compatibility and needs to be installed according to EMC information found in this manual.
Page 7
ClearView™ or shielding the location. • The use of ACCESSORIES other than those specified by CONMED, or sold by CONMED as replacement parts for internal components, may result in increased emissions or decreased immunity of the ClearView™...
Page 8
• Do not block either the tubing or the filter. If either becomes occluded or significantly restricted, the motor/blower may overheat and cause the unit to fail. • Using any other filter or accessory not supplied by CONMED may cause damage and/ or cause the system to be inoperable and may void the warranty.
A footswitch or a Automatic Activation Device may be added to any system by simply plugging in a CONMED activation accessory into the appropriate jack on the front or back of the unit. When the footswitch is plugged in, the unit may be turned on or off by depressing the footswitch pedal once for each operation.
Page 12
If the system is overheated or if there is an electrical surge in the electrical system, fuses will break and the system will not operate. When the Service light illuminates, please contact CONMED Customer Service for system service instructions.
Note: Before installing or removing any filter, be sure that the system is turned off. Filter Installation Instructions: The installation of the Filter into the CONMED ClearView™ Smoke Evacuation System(s) is quick and simple. 1. Remove the Filter from the shipping box and discard any protective wrapping. Examine all filters for damage during shipping and storage.
Section 3.0 General Maintenance Information This section contains information for ordinary upkeep of the CONMED ClearView™ Smoke Evacuation System. While the system has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed by a qualified Biomedical Technician to ensure continued safe and effective operation.