Page 1
Gas Mass Flow Meter VE.2.01 Model MF5000 http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
Page 2
Gas Mass Flow Meter with MEMS calorimetric sensing technology MF5000 Series User Manual Document No. 07-2022-M7 EN Issue date: 2022.07 Revision: VE.2.01 久德電子 TEL:+886-4-23729418 FAX:+886-4-23724011 (40349)台中市西區福人街11號 Email: sales@jetec.com.tw http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual P a g e...
Page 3
• The gas to be measured should be clean and free of particles. Do not apply this meter for liquid medium. • Do not apply for any unknown or non-specified gases that may damage the product. • For remote data, please be sure the meter is properly configured. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
5.4.12 Reset the accumulated or totalized flow rate ..............20 5.4.13 Reset the 2nd-correction factor ..................21 5.4.14 Exit the MENU ......................21 5.4.15 MENU key sequence for the settings................22 5.5 RS485 Modbus communication protocol ..............22 http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
Page 5
8.2.2 Particle contamination and fluidic cleanness ..............35 8.2.3 Apply to a different gas medium................... 35 9. Troubleshooting ..................36 10. Service contact ..................37 Appendix I: Product evaluation kit ..............38 Appendix II: Document history ............... 39 http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
0, Group A, T4) as well. It also boosts the protection class to IP66 and offers additional mechanical connection flexibility with both male and female connectors which can be further adapted with any other threaded ones. The electrical interface is changed from DB9 to M12 for better protection. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
In most cases, this manual shall be made available to the customer before the actual order. The cable is a standard one with an M12 connector. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011...
The LCD provides all information that the product measures. Some symbols are reserved for future upgrades, and will not be lighted during the operation. The following table details the meaning of each of the symbols. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
12 ~ 24 Vdc White RS485 B (-) Blue GND, common ground Socket Cable Black RS485 A (+) Gray 4 ~ 20mA output Figure 3.3: MF5000 socket and cable http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual P a g e...
3.4. Mechanical dimensions Figure 3.4: MF5000 meter dimensions Model DN (mm) D (NPTM) MF5003 1/8” MF5006 1/4” MF5008 3/8” MF5012 1/2” MF5019 3/4” 182.5 http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual P a g e...
To ensure the measurement accuracy, an upstream straight pipe of length no less than 10DN and a downstream straight pipe of length no less than 5DN should be in place. Please refer to the following recommended installation configuration. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011...
Page 12
(e) Pipe size expansion (f) Control valve at upstream or downstream Figure 4.3: MF5000 meter installation (iii) During installation, please make sure no foreign materials (such as water, oil, dirt, particles, etc.) enter the installation pipeline. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
Page 13
Make sure no mechanical stresses in the connections. d) The strong electromagnetic interference sources close by or any mechanical shocks at the pipeline may also create malfunctioning of the product. e) Slowly open/close valves to prevent abrupt pulse flow impact. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
The manufacturer-supplied power and data cable have a locking fixture. Lock the cable and make sure it is properly engaging and will not be accidentally got unplugged. Half-duplex RS485 Modbus is used for digital data communication. Make sure the wires are properly connected to the receiver side. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
LCD. The middle line is the accumulated or totalized flow rate, and the lower line is the instant flow rate. The upper line will light up when the pressure or temperature option is selected. The display characters are limited by the LCD capability, the following table is the illustration. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011...
F2 - Addr, which indicates the task is completed. Press the “Up” or “Down” key to select F99 - qUIT and the “M” key to exit the MENU and return to the Main Display screen. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
MENU asking you to confirm. Use the “Up” or “Down” key to select the desired one and press the “M” key to confirm. The display will then return to the F11 - oFFST screen, which indicates the task is http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011...
The display will then return to the F12 - GCF screen, which indicates the task is completed. Use the “Up” or “Down” key to select F99 - qUIT and the “M” key to exit the MENU and return to the Main Display screen. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
“M” key to confirm. The display will then return to the F16 - rESPS screen, which indicates the task is completed. Use the “Up” or “Down” key to select F99 - qUIT and the “M” key to exit the MENU and return to the Main Display screen. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
F32 - UnT-F screen, which indicates the task is completed. Use the “Up” or “Down” key to select F99 - qUIT and the “M” key to exit the MENU and return to the Main Display screen. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011...
- CLr-A screen, which indicates the task is completed. Use the “Up” or “Down” key to select F99 - qUIT and the “M” key to exit the MENU and return to the Main Display screen. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
At the MENU settings, use the “Up” or “Down” key to select the F99 - qUIT option and press the “M” confirming key to exit the MENU settings and return to the Main Display screen. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
5.4.15 MENU key sequence for the settings Figure 5.2: MF5000 MENU key sequence http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual 22 | P a g e...
9600 bps Start bits Data bits Stop bits Even/Odd parity None Bits period 104.2 µsec Bytes period 1.1458 msec Maximum data length Maximum nodes http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual 23 | P a g e...
Note: At the time of shipping, the write protection function is enabled except for address and baud rate. Once the user completes the register value change, the write protection will be automatically enabled once again to prevent incidental data loss. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011...
Page 26
Flow rate = [Value (0x003A)*65536 + value (ox003B)]/1000 e.g.: For a flow rate of 20.340 L/min, the user will read “0 (0x0000)” from Notes register 0x003A and “20340 (0x4F74)” from register 0x003B, therefore Current flow rate = (0*65536+20340)/1000 = 20.340 http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
Page 27
Notes: please disable the write protection before executing this function. Write Units 0x0090 Read Description Set the instant flow rate unit, and accumulated or totalized flow rate unit Value type UINT 16 http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual 26 | P a g e...
Page 28
GCF, the user needs to send 0xAA55 Notes to the register 0x00FF, and then the write function will be enabled (write protection is disabled). After the write execution is completed, the firmware will automatically re-enable the write protection. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
The loop resistor connection is illustrated below. The current output load depends on the power supply (the yellow area in the graph). The maximum load resistor, R , with a 24Vdc supply, will be 850Ohm. Figure 5.4: Analog output. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec...
MF5006 MF5008 MF5012 MF5019 Maximum overflow 1200 SLPM Maximum flow change SLPM/sec Note: For other features or specifications not listed, please contact the manufacturer. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual 30 | P a g e...
Another key point to comparing the different flow meters is that as long as the fluidic flow is a continuous flow without pulsation, then the fluidic dynamic will have the system following the Bernoulli equation: http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
Therefore, measurement by the meter for a gas medium other than the calibration gas would bear larger measurement errors, particularly at the low Reynold number range where the laminar flow has a sensitive flow profile. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
Circuitry instability Check system, power off No digital interface Wrong address, software Check commands, connection No wireless, BT cannot pair Wrong model, data jam Check model, power off/on http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual 36 |...
Direct customer service request(s) should be addressed to 久德電子 TEL:+886-4-23729418 FAX:+886-4-23724011 (40349)台中市西區福人街11號 Email: sales@jetec.com.tw For orders, please provide an accurate and full postal address. http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw...
However, for multiple sensors in serial, connect to PC the power via the USB cable may not be enough, an external power adapter with 8 ~ 24Vdc will External power be required. adapter (optional) http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual...
➢ Add the RS485 Modbus protocol (Mode P1). Revision C.5 (November 2017): ➢ Add the maximum overflow and maximum flow changes; ➢ Add the revision history (Appendix). …… Revision A.0 (October 2007): ➢ First release http://www.jetec.com w om.tw/ 電話:04-23729418 傳真:04-23724011 Email:sales@jetec.com.tw LINE:@jetec MF5000 User Manual...