Model W393B04/ Seismic, miniature (50 gm), ceramic flexural ICP® accel., 1 V/g, 0.06 to 450 Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Repair and Maintenance Returning Equipment If factory repair is required, our representatives will PCB guarantees Total Customer Satisfaction through its provide you with a Return Material Authorization (RMA) “Lifetime Warranty Plus” on all Platinum Stock Products number, which we use to reference any information you sold by PCB and through its limited warranties on all other have already provided and expedite the repair process.
Page 3
CAUTION Refers to hazards that could damage the instrument. NOTE Indicates tips, recommendations and important information. The notes simplify processes and contain additional information on particular operating steps. The following symbols may be found on the equipment described in this manual: This symbol on the unit indicates that high voltage may be present.
Introduction Model 394C06 Handheld Shaker is a small, handy, completely self-contained vibration reference source. It is intended for rapid checking of vibration measurement, monitoring and recording systems using piezoelectric accelerometers, as well as other types of vibration transducers having a maximum weight of 210 grams. Accelerometer mounting hole Low Power...
A section view of the shaker’s vibration head is shown in Figure 2. The shaker consists of an electromagnetic exciter, driven by an oscillator at a frequency of 159.2 Hz (1,000 rads ). A small ICP® accelerometer provides a servo feedback to maintain a constant vibration level of 1 g. This enables accelerometers with masses up to 7.4 oz (210 grams) to be exited without their mass influencing the reference level.
Page 10
The unit has several unique features of interest: Continuous Use Option Switch: In this mode, once the shaker is turned on, it does not shut off until the On/Off button is depressed a second time. The default position of this switch is Auto; the unit turns off automatically after 90 seconds.
Warnings & Cautions Warning 1 – Non-Serviceable The power supply/signal conditioner should not be opened by anyone other than qualified service personnel. This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid injury.
Warranty PCB instrumentation is warranted against defective material and workmanship for 1 year unless otherwise expressly specified. Damage to instruments caused by incorrect power or misapplication, is not covered by warranty. If there are any questions regarding power, intended application, or general usage, please consult with your local sales contact or distributor.
Total Customer Satisfaction PCB, a division of PCB Piezotronics, guarantees Total Customer Satisfaction. If, at any time, for any reason, you are not completely satisfied with any PCB product, PCB will repair, replace, or exchange it at no charge. You may also choose, within the warranty period, to have your purchase price refunded.
Page 14
Model Number Revision: G SEISMIC ICP® ACCELEROMETER ECN #: 47130 393B04 Performance ENGLISH OPTIONAL VERSIONS Sensitivity(± 10 %) 1000 mV/g 102 mV/(m/s²) Optional versions have identical specifications and accessories as listed for the standard model except where noted below. More than one option may be used. Measurement Range ±...
Page 15
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. STANDARD STUD MOUNT UPDATE DRAWING 25686 "A" "B" "D"...