Outline Of Instrument; Simplified Explanation; Sampling Flow - Nippon AQUA Instruction Manual

Table of Contents

Advertisement

Instruction Manual: AQUA kit for MA-3 Solo

2. OUTLINE OF INSTRUMENT

SIMPLIFIED EXPLANATION

This unit is the option for MA-3 Solo for the measurement of extremely low concentration
liquid samples like environmental water. The mercury contained in the liquid samples is
reduced and vaporized, and then collected to the mercury collector tube. The amount of
mercury is determined by MA-3 Solo.

SAMPLING FLOW

When sulfuric acid is added to sample and then stannous chloride solution is added, bivalent
mercury ions (Hg
by bubbling. The collection concentration is carried out selectively for the mercury gas as
gold amalgam.
*Acid mist and steam generated by bubbling are removed using a KOH solution and a mist
catcher because they obstruct a measurement.
* Although sample with small quantities of interfering substances, such as raw water for tap
water, can be measured only by adding sulfuric acid, it is impossible to obtain a precise and
accurate measurement result for sample containing interfering components.
Exhaust Port
Nippon Instruments Corporation
2+
) are reduced to zerovalent metal mercury and changed to mercury gas
Hg
2+
→ Hg
0
+SnCl
2
Sample
Reagent (sulfuric acid,
stannous chloride solution)
2
NIC-TD-0000187-01
Mist catcher
Mercury
Collector Tube
KOH solution