Magnetic-Inductive Flow Velocity Sensor UMF2 Operating Manual Please read the instructions carefully and store them in a save space Heinrichs Messtechnik GmbH OPERATING MANUAL PIT / UMF2B January 2021...
11 “Returning the Meter”. before returning the device to Heinrichs Messtechnik. The operator is liable for any substance removal or personal damage costs arising from inadequate cleaning of a device sent for repair.
The device is to be installed and serviced by a qualified technician only. The UMF2B transmitter is to be used exclusively to measure volume flow or flow velocity, in conjunction with a Heinrichs Messtechnik EPS, PIT or PITE sensor.
Heinrichs Messtechnik GmbH 3 Application The magnetic-inductive flow velocity sensor PIT-UMF2 is used to measure or monitor the volume flow of liquids with and without solids concentration, slurries, pastes and other electrically conductive media while minimizing pressure drops. The medium must possess a conductivity of at least 20 µS/cm.
Heinrichs Messtechnik GmbH 4 Mode of Operation and System Design 4.1 Mode of operation It was back in 1832 that Faraday suggested utilizing the principle of electromagnetic induction for measuring flow velocities. His experiments in the Thames, though unsuccessful due to superimposed polarization effects, are nonetheless regarded as the first ones in the field of magnetic-inductive flow measurement.
The PIT-*** sensor can be operated with all transmitters for magnetic-inductive flow meters manufactured by Heinrichs Messtechnik. The sensor is installed (inserted) into the pipe while the transmitter is mounted directly on top of the sensor or separate remote, depending on the equipment design.
4.2.2 Remote installation of the transmitter Heinrichs Messtechnik recommends this type of installation when there is little space or the medium temperatures are high. The sensor and the transmitter are connected by a field coil and an electrode cable. The electrode cable must be shielded and protected against disturbing interferences.
Heinrichs Messtechnik GmbH 5 Characteristics Accuracy 1.5 % of measured value plus 0.5 % of URV Conductivity of fluid 20 µS/cm Influence of ambient temperature Pulse output 0, 05% pro 10K Current output 0, 1% pro 10K...
Heinrichs Messtechnik GmbH 6 Conditions of use 6.1 Installation conditions Disturbing elements (e.g. shut-off and control valves) are to be arranged downstream from the sensor. If this is not possible, flow conditioners must be installed so that no vortexes can reach the pipe section of the sensor.
Heinrichs Messtechnik GmbH 6.1.3 Grounding For safety reasons and to ensure faultless operation of the magnetic-inductive flow meter, earthing of the flow sensor is important. In accordance with VDE 0100, Part 540, the earthing connections must be connected to the grounding conductor.
Heinrichs Messtechnik GmbH 6.4 Medium temperature and pressure 6.4.1 Direct mounted transmitter Version Medium temperature Pressure PIT-S ( SS / PTFE) -20°C to 80°C 16 bar PIT-A ( PFA) -20°C to 80°C 40 bar 6.4.2 Remote mounted transmitter Version Medium temperature...
Heinrichs Messtechnik GmbH 7 Dimensions/weight 7.1 PIT-UMF2 dimensional drawing for remote transmitter Transmitter with terminal connection box (all dimensions in mm ) Weight: 3, 6 kg Nominal Diameter Transmitter with directly connected cable ( up to 10 m) Length of sensor Ød:...
Heinrichs Messtechnik GmbH PIT-** dimensional drawing with direct mounted transmitter UMF2 (all dimensions in mm ) Weight: 5, 5 kg Nominal Diameter Length of sensor Ød: Diameter of sensor Length of socket weld fitting Ød1 Version PIT-A (PFA) 150-600 163mm...
Heinrichs Messtechnik GmbH 8 Power supply / electrical connection See name plate or Operating Instructions of the corresponding transmitter. 9 Safety information 9.1 Intended use The PIT-UMF2 flow meter may be used only for flow measurements of fluids whose conductivity exceeds 20µS/cm.
Heinrichs Messtechnik GmbH 10 Options 10.1 Removal and installation under process pressure In some cases it might be necessary to remove and reinstall the PIT sensor head for cleaning when the pipe is under process pressure. A special retraction device can be used for this purpose.
Heinrichs Messtechnik GmbH 10.2.1 Wiring diagram for remote installation The outer shield hast o be connected to the metalized cable glands at both ends. The inner shields are connected to each other and are plugged into the terminal with the label “Schirm / shield”.
Heinrichs Messtechnik GmbH 11 Returning the Meter If all attempts to return the device to an operational condition have failed, contact our service department to arrange the return of your device for repair. Before sending the device back for repair or servicing, please ensure the following steps have been performed: Always enclose a fully completed declaration of decontamination.
Heinrichs Messtechnik GmbH 11.1 Declaration of Decontamination Company name: ..............Address: ................Department: ................ Contact person: ..............Phone: .................. The enclosed Coriolis flow meter Model: ..........................was operated using the following fluid: ........................... In as much as this fluid is water-hazardous / toxic / corrosive / combustible / a health hazard /...