Tequipment 9840 User Manual

Ac voltage/current/power calibrator

Advertisement

Model 9840 and 9841
AC Voltage/Current/Power
Calibrator
User's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 9840 and is the answer not in the manual?

Questions and answers

Summary of Contents for Tequipment 9840

  • Page 1 Model 9840 and 9841 AC Voltage/Current/Power Calibrator User's Manual...
  • Page 2: Table Of Contents

    Page 1 CONTENTS 1. General Description……………………………………………… 2. Specifications…………………………………………………….. 3. Precautions……………………………………………………….. 4. Operation………………………………………………………….. 5. 3 Phase configuration…...………………………….…………….. 8 6. Re-Calibration, Guarantee and Servicing…………..…………… 9...
  • Page 3: General Description

    Page 2 1. GENERAL DESCRIPTION Calibrators 9840 and 9841 are single phase sources of alternating current and voltage. Providing voltages in four ranges, 57-110-220-380 -500V Current up to 20A (100A for the 9841) in ranges 1-5-20A and 100A(9841). Frequency in the range 45 to 70Hz.
  • Page 4: Specifications

    2. SPECIFICATIONS Page 3 Table 3.1. Accuracy Parameter Range Settings Span Resolution Accuracy Maximum Load 0.5 - 60V 0.01V 250mA@60V 110V 1.00 - 125V 0.01V ±0.05% of set 140mA@130V value ±3 digits Voltage 220V 2.0 - 250V 0.1V 70mA@250V 500V 3.0 - 520V 0,1V 40mA@420V...
  • Page 5: Precautions

    3. PRECAUTIONS Page 4 The calibrator should be used in an environment according to table 3.2 and where apparatus will be handled carefully. The calibrator should not be used near aggressive gases and strong magnetic and electric fields. Operation is allowable according to IEC359 for instruments group 1. It can be used in temperatures from +5 to +40 deg C and stored/transported at temperatures from -25 to +70 deg C.
  • Page 6: Operation

    Page 5 4. OPERATION “power” Supply switch, connects the input power to the calibrator “output voltage” Output voltage terminals: high potential “HI” and low potential “LO”. There is a pushbutton and the indicator for standby/operate function next to the terminals. Indicator flashing is “standby” mode and the output voltage is disconnected.
  • Page 7 Page 6 “phase” Phase angle between current and voltage signals is set up by the control knob. Pressing the pushbutton below the display selects the range of phase. When the indicator “L” light is on, the range is 0 to +90 and when the indicator “C”...
  • Page 8 Page 7 Example load connection - for calibrating electricity meters M eter Under Test Calibrator output voltage output current Fig.5.2. Connection of wattmeter to calibrator terminals When connecting or disconnecting the loads please comply with the guidelines shown in Section 4 - PRECAUTIONS. When the calibrator goes into overload the “ovl”...
  • Page 9: Phase Configuration

    Page 8 5. THREE PHASE CALIBRATOR SET Calibrator 9840/41 can work as a main controller for a three phase set. Two additional slave units are linked by rear panel connections. Three phase calibrator configuration Shift angle between voltage signal 120.0° with error ±1.0° ** Difference between output voltage in any case ±1% setting value...
  • Page 10: Re-Calibration, Guarantee And Servicing

    9 6. RE-CALIBRATION, GUARANTEE, AND SERVICING Re-Calibration: The 9840/41 should be returned to Time Electronics or their authorised dealer for repair or recalibration. Guarantee Period This unit is guaranteed against defects in materials and workmanship for a period of one year from its delivery to the customer.

This manual is also suitable for:

98419845984698479848