Summary of Contents for Calex PyroMiniBus Series

  • Page 1 PyroMiniBus Series Operator’ s Guide...
  • Page 2: Sensor Specifications

    PyroMiniBus non-contact infrared sensors measure temperatures from -20°C to 1000°C and provide an RS485 Modbus RTU interface. The sensors are capable of being used in ambient temperatures up to 120°C without cooling (XCF models: 70°C). They may be connected directly to third-party Modbus Masters, and they are compatible with the optional PM180 6-channel touch screen interface, which provides temperature display, data logging, alarms and analogue outputs.
  • Page 3 DIMENSIONS Removable spigot Mounting Nut Air Purge Collar 1/8" BSP fitted to type APSN (included) air fitting Sensing Head for use with -201 Thread M16 and -XCF models. Ø 18 Ø 18.6 x 1 mm Not fitted to type APSW for -21 models.
  • Page 4: Installation

    INSTALLATION The installation process consists of the following stages: • Preparation • Mechanical installation • Electrical installation Please read the following sections thoroughly before proceeding with the installation. PREPARATION Ensure that the sensor is positioned so that it is focused on the target only. DISTANCE AND SPOT SIZE The size of the area (spot size) to be measured determines the distance between the sensor and the target.
  • Page 5: Power Supply

    WIRING Check the distance between the sensor and the Modbus Master. If necessary, the sensor can be ordered with a longer cable attached. POWER SUPPLY Be sure to use a 24 V DC power supply. The maximum current draw is 50 mA per sensor. MECHANICAL INSTALLATION All sensors come with a 1 m cable and a mounting nut.
  • Page 6: Electrical Installation

    This diagram shows how to connect the sensors and output modules to a Modbus Master, such as a Modbus PLC, SCADA system, Calex Panel PC or PM180 touch screen terminal. Up to 247 slave devices, such as PyroMiniBus sensors and output modules, can be connected to the same Modbus network.
  • Page 7: List Of Modbus Registers

    LIST OF MODBUS REGISTERS The list below includes all available addresses on the PyroMiniBus sensor. R = Read, W = Write Address Length (words) Description 0x00 Sensor Type (17 for PyroMiniBus) 0x01 Field of view 0 for 2:1; 1 for 20:1 0x02 Serial number 0x04...
  • Page 8: Operation

    OPERATION Once the sensor is in position and the power, air, water, and cable connections are secure, the system is ready for continuous operation by completing the following simple steps: Turn on the power supply Turn on the Modbus Master Read / monitor the temperature IMPORTANT BE AWARE OF THE FOLLOWING WHEN USING THE SENSOR:...
  • Page 9: Lens Cleaning

    example, it is in the same room). In this case, Reflected Energy Compensation should remain Disabled for an accurate measurement. However, if the sensor is positioned outside an oven or furnace, with the target object inside, the reflection of the hot furnace interior can affect the measurement. In this case, Reflected Energy Compensation should be turned ON and Reflected Temperature should be set to the temperature inside the oven or furnace.
  • Page 10 Calex infrared temperature sensors. It is compatible with all models in the PyroBus and PyroMiniBus series, as well as all PyroMini and FibreMini models with Modbus output. The PM180 functions as the Modbus Master on an RS485...
  • Page 11: Data Logging

    DATA LOGGING With an optional MicroSD Card installed in the slot inside the PM180, data logging may be manually started and stopped via a button on the temperature display screen, or scheduled to begin at a pre-determined time via the Settings menu. If the PM180 is connected via the slave interface to another Modbus network, logging may also be started remotely by the Modbus Master on that network.
  • Page 12: Electrical Connection

    ELECTRICAL CONNECTION...
  • Page 13: Using The Touch Screen Interface

    USING THE TOUCH SCREEN INTERFACE Visit www.calex.co.uk/PM180 for help on how to use the touch screen interface, including a diagram showing how to navigate to each setting and set up the optional output modules.
  • Page 14 MODBUS OVER SERIAL LINE (RS485) INTERFACE SUPPORTED FUNCTIONS Baud rate 9600 Read register 0x03, 0x04 Format 8 data, No parity, 1 stop Write single register 0x06 Reply delay (ms) Write multiple register 0x10 The list below includes all available addresses (R = Read, W = Write): Address Length Description...
  • Page 15: Maintenance

    Calex for further advice. GUARANTEE Calex guarantees each instrument it manufactures to be free from defect in material and workmanship under normal use and service for the period of two years from the date of purchase. This guarantee extends only to the original buyer according to Calex Terms and...
  • Page 16 Issue E - Jul 2018 Calex Electronics Limited PO Box 2, Leighton Buzzard, Bedfordshire, England LU7 4AZ Tel: +44 (0)1525 373178 Fax: +44 (0)1525 851319 Email: mail@calex.co.uk Online: www.calex.co.uk...