Concurrent Measurement Command; Concurrent Measurement Command With Crc - Xylem YSI WaterLOG H-3301 Owner's Manual

Incremental shaft encoder with sdi-12 and quadrature outputs
Table of Contents

Advertisement

Example of an 'aM!' Command:
Command
Response
'aMC!'
"a0012<cr><lf>"
Subsequent Command
"aD0"
Where:
X.XX
= Current position (stage), in user programmable units
NNNN
= Current position, in raw position counts
YYY
= CRC (ASCII Characters)

Concurrent Measurement Command

A "Concurrent Measurement" is one which occurs while other SDI-12 sensors on the bus are also
taking measurements. This command is similar to the "aM!" command, however, the nn field has
an extra digit and the sensor does not issue a service request when it has completed the measurement.
Communicating with other sensors will NOT abort a concurrent measurement. Data values generated
in response to this command are stored in the sensor's buffer for subsequent collection using "D"
commands. The data will be retained in the sensor until another "M", "C" or "V" command is executed.
Command
"aC!"
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
C
is an upper-case ASCII character
ttt
is a three digit integer (000-999) specifying the maximum time, in seconds, the sensor
will take to complete the command and have measurement data available in its buffer.
nn
is a two digit integer (00-99) specifying the number of values that will be placed in the
data buffer. If "n" is zero (0), no data will be available using subsequent "D" commands.
The data recorder may wake the sensor with a break and collect the data anytime after the specified
processing time has elapsed.

Concurrent Measurement Command with CRC

The "Concurrent Measure with CRC" command causes a measurement sequence to be performed
with a CRC appended to the data. Data values generated in response to this command are stored
in the sensor's buffer for subsequent collection using "D" commands. The data will be retained in
the sensor until another "M", "C" or "V" command is executed.
Command
"aCC!"
Time
1 sec
Response
a+X.XX+NNNN<cr><lf>
Response
"atttnn<cr><lf>"
Response
"atttnn<cr><lf>"
SDI-12 Command & Response Protocol
Values
Description
2
Return encoder position
Description
Initiate measurement
Description
Initiate measurement