www.infiuxbrasil.com.br
INFLUX BRASIL – Av Paraná, 3059 – Cajuru do Sul — Sorocaba
– São Paulo - CEP:18105-002 Tel.: (15) 3359 9739 (15) 99621
1260 - e-mail: vendas@influxbrasil.com

Advertisement

Table of Contents
loading

Summary of Contents for Influx Technology INU300

  • Page 1 INU300 Touch Screen Ultrasonic Level Meter User Manual www.infiuxbrasil.com.br INFLUX BRASIL – Av Paraná, 3059 – Cajuru do Sul — Sorocaba – São Paulo - CEP:18105-002 Tel.: (15) 3359 9739 (15) 99621 1260 - e-mail: vendas@influxbrasil.com...
  • Page 2: Table Of Contents

    Catalogue 1Summary...……………………………………………………………………………………….….…3 2 Technical indicators and selection code………………….….……………….………………….….…4 3 Installation guide………………………………………….……………….……………………….….5 3.1 Bracket installation and flange dimensions ……………………………...……..………..….5 3.2 Installation guide ………………………………………………………………….….…..……6 3.3 Principles of installation…………………………………………………………..….………...6 3.4 Installation notes …………………………………………………………….………………..6 3.5 Instrument connection………………………………………………………………………..7 4 Instrument debugging…………………………………………………………………...…………..9 4.1 Instrument display …………………………………………………………….…….………..9 4.2 Keyboard……………………………………………….……………………….….…..….….10 4.3 Menu…………………………………………………………………………….………..…11 4.4 Parameter setting ……………………………………………………………...…………...…12...
  • Page 3 Notes Do not use and transportation process in the strong shaking or impact device. Instrument in the transport and storage period, the environment temperature does not allow lower than -40 ℃ and above +80 ℃, relative humidity is not more than 90%, and the surrounding does not contain any corrosive gas, no strong electromagnetic field, using the original distribution box during transportation.
  • Page 4 The instrument can be connected to a variety of DCS system through 4~20mA or RS485 (Modbus/protocol), provide real time level data for industrial automation operation. INU300 ultrasonic level meter has following advantages:  Stable and reliable  We select high quality module starting from the power part of circuit design, procurement of key components of high stability and reliability of the device selected.
  • Page 5: Technical Indicators And Selection Code

    2、Technical indicators and selection code Type: TL300E Measuring range: 0~10m Blind spot: 0.45m~0.60m Precision: 0.3% Resolution: Pressure: Under 4 atmospheres With LCD display level or space Instrument display: distance Analog output: 4~20mA Digital output: RS485, Modbus protocol or custom protocol AC110/220V DC24V, build-in lightning Power supply: protection...
  • Page 6: Installation Guide

    3.Installation guide 3.1 Bracket installation and flange dimensions Note: the flange size will change based on different transducers and will be forewarned. 3.2 Installation guide 3.2.1 Bracket installation: fixed L-type bracket on the top of level, then fixed the level meter on the bracket with screws. 3.2.1 Flange installation: fixed the flange on plate or board, then fixed the level meter on the flange.
  • Page 7: Installation Notes

    3.4 Installation notes 3.4.1 The instrument should have connected with the earth reliably. 3.4.2 Wire and cable protection pipes to prevent excessive water. 3.4.3 Although the instrument already with a lightning protection device, preferably plus a specific lightning protection device on the inlet and outlet side of instrument when using in the area full of thunders.
  • Page 8: Instrument Debugging

    1-2 is DC24V power input,3-4 is RS485 output,5-6 is 4~20mA output. Note: instrument use AC220V power, do not use XS4’s 1-2, instrument has AC200V lane leads to the outer shell, with a AC220V power plug. Note:  CA-5 specific cable length should be (split type) ≤50m ...
  • Page 9: Keyboard

    4.2 Keyboard Keyboard formed by the four keys, use “inductive-pen” point button, enter the parameter setting state Keyboard described as below: 【Mode】:Mode key.In normal working(power on 30s),use the”inductive-pen”click the button to enter the parameter setting menu; In the menu, use the”inductive- pen”click the button, can withdraw from the class menu, return to the previous menu.
  • Page 10 INFLUX BRASIL – Av Paraná, 3059 – Cajuru do Sul — Sorocaba – São Paulo - CEP:18105-002 Tel.: (15) 3359 9739 (15) 99621 1260 - e-mail: vendas@influxbrasil.com...
  • Page 11: Parameter Setting

    4.4 Parameter setting 4.4.1 Calibration steps 1)Liquid crystal display the level value after instrument plug in, shown as the right figure, the actual level is 3.320m, calibration method described as following 2)Using the “inductive-pen” to click Mode key to enter the parameter setting menu, first entry is the P01(level)calibration menu.
  • Page 12: Display Mode

    3)Click ▼ can choose other relays, click OK button to enter relay 1 setting interface, shown as right figure, the formula indicated relay 1 The relay pick-up when the level is greater than 4m and less than 8m. < and > symbol location in the interface can choose “<”...
  • Page 13 3.5.2.2 AirH option display the space distance and temperature. 3.5.2.3 LevelT option display the level and temperature. Click OK to choose. 4.4.5 Address ID number setting Adrr menu for communication, especially the multi-machine communication, the default is 01 4.4.6 P—Multi Manu P—Multi Menu had been setted before out of factory, and generally do not change。...
  • Page 14 temperature exceeds the range IN300E Explosion-proof Type Ultrasonic Level Meter Warranty Card Name Address Contact Product Model Product No. Installing the person Acceptance Date in charge …………………………………………………………………… YIN3000E Explosion-proof Type Ultrasonic Level Meter Warranty Card Description Product Model Product No. Installing the person Acceptance Date in charge...
  • Page 15 reissued.  This warranty card explains permissions owned by the Company, the Company reserves the right to modify the contents of the card, without prior notice. INFLUX BRASIL – Av Paraná, 3059 – Cajuru do Sul — Sorocaba – São Paulo - CEP:18105-002 Tel.: (15) 3359 9739 (15) 99621 1260 - e-mail: vendas@influxbrasil.com...