FDI P2131 Technical Manual

FDI P2131 Technical Manual

Automated field steam sterilizer
Table of Contents

Advertisement

Quick Links

P2131 
AUTOMATED FIELD STEAM STERILIZER 
NSN: 6530‐01‐641‐4641 
TECHNICAL MANUAL  
Fort Defiance Industries LLC (FDI) 
2411 Maremont Pkwy ∙ Loudon, TN ∙ 37774 
Phone: (865) 408‐0100 ∙ techsupport@fortdefianceind.com 
P2131‐1‐8001 Rev 17 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P2131 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for FDI P2131

  • Page 1 P2131  AUTOMATED FIELD STEAM STERILIZER  NSN: 6530‐01‐641‐4641  TECHNICAL MANUAL   Fort Defiance Industries LLC (FDI)  2411 Maremont Pkwy ∙ Loudon, TN ∙ 37774  Phone: (865) 408‐0100 ∙ techsupport@fortdefianceind.com  P2131‐1‐8001 Rev 17 ...
  • Page 2 The P2131 Automated Field Steam Sterilizer complies with the  applicable requirements of the following standards:   ANSI/AAMI ST8:2013, Hospital Steam Sterilizers  UL 61010‐1 2nd Edition, Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use – Part 1 General Requirements  IEC 61010‐2‐040 1st Edition, Safety requirements for electrical equipment for measurement, control and laboratory use ‐ Part 2‐040: Particular requirements for sterilizers and washer‐disinfectors used to treat medical materials  ASME Boiler and Pressure Vessel Code, Section VIII, Division I, Current Edition, Rules for Construction of Pressure Vessels U  Fort Defiance Industries (FDI) is ISO 13485:2016 certified   and FDA compliant for the design and manufacture of medical devices  and supportive accessories for the medical device industry.   Fort Defiance Industries LLC          Certified to ISO 13485:2016       MDSAP: United States, Canada         Certificate #: US19/81841407 ...
  • Page 3 Continual Improvement  Fort  Defiance  Industries  LLC  (FDI)  is  committed  to  providing  our  customers  with  a  safe  and  effective product that is simple and easy to operate and maintain.  As an ISO 13485:2016 certified  company, we have a strong desire to continually improve.  If you have any feedback regarding  the  P2131  sterilizer  or  this  technical  manual,  please  contact  us.  If  any  critical  updates  to  the  P2131 Sterilizer become necessary, the updates will be communicated through a Medical Device  Advisory Notice. It is imperative that these notices be read in full and followed.   FEEDBACK FOR IMPROVEMENT AND GENERAL QUESTIONS Fort Defiance Industries LLC  2411 Maremont Pkwy  Loudon, TN  37774 ...
  • Page 4: Table Of Contents

    TABLE OF CONTENTS  SECTION  DESCRIPTION  PAGE  1  INTRODUCTION  1  1.1  Safety Precautions  1  2  GENERAL INFORMATION  3  2.1  Intended Use and System Design  3  2.2  Specifications and Physical Data  4  2.3  Sterilization Cycles – FDA Cleared  5  2.4  Description of Cycle Phases  6  2.5  Water Supply Quality  7  2.6  Sterility Assurance Products  8  3  OPERATION  9  3.1  Installation, Setup, and Qualification (IQ, OQ, PQ)  11  3.2  Controls and Indicators  29  3.2.1 ...
  • Page 5: Section Description Page

    SECTION  DESCRIPTION  PAGE  4.1.5  Post Maintenance Checklist 81  4.2  Special Tools and Materials Required for Maintenance  82  4.3  Maintenance Procedures  83            Replacing the HTS‐1 RTD Sensors  4.3.1  84            Disinfecting Clean for WRS  4.3.2  85            Cleaning the Sight Glass  4.3.3  85            Replacing the SWS Cartridge  4.3.4  86            Replacing the HEPA Filter  4.3.5  86            Replacing the Steam Trap  4.3.6  86            Replacing the RTD Sensor  4.3.7  87            Replacing the Chamber Door Gasket  4.3.8  87            Replacing the Chamber Doorpost O‐Ring  4.3.9  88            Replacing Electric Immersion Heating Elements  4.3.10    ...
  • Page 6 COMPACT DISC TABLE OF CONTENTS  The following files can be found on the CD that accompanies this technical manual. ITEM  FILE NAME  1.  Technical Manual – P2131 Automated Field Steam Sterilizer  2.  An Installer’s Pocket Guide for Swagelok Tube Fittings  3.  Programmable Logic Controller (PLC) Manual  LIST OF ACRONYMS  ACRONYM DEFINITION     AAMI  Association for the Advancement of Medical Instrumentation  AFS  Automated Field Steam Sterilizer   ASME  American Society of Mechanical Engineers  BI  Biological indicator  B&PV  Boiler and Pressure Vessel  CI  Chemical integrator  DI  Deionized  FDA  Food and Drug Administration  FDI  Fort Defiance Industries LLC  IFU  Instructions for Use  HTS  High Temperature Limit Switch  IQ/OQ/PQ  Installation, Operational, and Performance Qualification  IUSS ...
  • Page 7: Introduction

        WARNINGS      This  manual  contains  critical  information  on  the  proper  use  of  the  Automated  Field  Steam  Sterilizer  (P2131  sterilizer).  All  operators  and  maintainers  of  this  sterilizer  must  be  properly  trained prior to operation or service and are urged to carefully read this manual in its entirety to  become familiar with the warnings, cautions, and instructions for use.     This sterilizer is equipped with a High Temperature Limit Switch. Should the heaters be  energized in a low or no water condition, this switch will protect the chamber and other  components  by  switching  off  power  to  the  PLC  and  the  Heater  Contactor.  This  alarm ...
  • Page 8    Refer to the instrument device manufacturer’s Instructions for Use (IFU) before selecting  the appropriate cycle, including exposure and dry times.     The  P2131  sterilizer  is  designed  to  sterilize  loads  using  only  the  FDA‐cleared  cycles  as  specified in this manual.  If you have any question or concern about a specific material,  instrument, or device, contact the manufacturer of the device to obtain the specific IFU.     The P2131 sterilizer does not include a sterilization cycle for liquids. DO NOT ATTEMPT  TO STERILIZE ANY LIQUIDS IN THIS STERILIZER.   After  transport  and  relocation,  the  installation  must  include  proper  Installation, ...
  • Page 9: General Information

    SECTION 2 – GENERAL INFORMATION    Intended Use and System Design      The P2131 sterilizer is designed for sterilization of porous and nonporous and heat‐ and moisture‐ stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131  sterilizer  is  a  rugged,  transportable  device  designed  for  field  use  in  a  variety  of  austere  environments.  Ambient operating conditions include sea level to 8,000 feet altitude and 40°F to  130°F.    The P2131 sterilizer is a pre‐vacuum and post‐vacuum sterilizer that has a conditioning stage with  vacuum air removal before the start of the exposure stage, as well as a post‐exposure drying  stage that is based on the combined operation of heat and vacuum.  The sterilization agent is  steam that is electrically generated in a jacketed boiler.    The P2131 sterilizer consists of three components that work together as one integrated system.  These components are the (1) Sterilizer Main Unit (SMU), (2) Water Recovery System (WRS), and  (3) Sterilizer Water Softener (SWS).  The WRS reclaims hot condensate and exhaust steam for re‐ use by the SMU.  The WRS also includes a water eductor, which provides the vacuum capability  for the pre and post‐vac phases of sterilization.  The SWS deionizes the incoming water to prevent  scale buildup and reduce the effects of water on the metal components.  A manual hand pump  is also provided with the SWS to pull water from a Jerry can or bucket if a pressurized water  system is not available.      The P2131 sterilizer is an automated, microprocessor‐controlled, pre‐vacuum autoclave. A visual ...
  • Page 10: Specifications And Physical Data

    secondary safeguard, the unit is equipped with a High Temperature Limit Switch to turn off power  to  the  heaters  and  PLC.  Hot  condensate  exits  the  SMU  to  the  WRS  through  a  chamber  drain  screen, drain manifold, and hose.  Air and steam can exhaust rapidly to the WRS through the  same drain manifold via hoses and the vacuum/exhaust solenoid valve.  The WRS sits underneath  the  SMU  and  has  three  main  purposes:  (1)  to  capture  exhaust  steam  and  condensate,  thus  creating a closed‐loop system to reclaim nearly all the water used during sterilization, (2) to cool  and then pump the recovered water back to the boiler for reuse, and (3) to generate vacuum  through a water eductor for the pre‐ and post‐vac phases of operation.    Specifications and Physical Data   ...
  • Page 11: Sterilization Cycles - Fda Cleared

    Wrapped  4  3  36 lbs.  270/132  Instruments  instruments/utensils  Bowie‐Dick  Test  NA  273/134  3.5  0  Test  Vacuum Leak  Test time  Test  NA  NA  0  Test  20     The AAMI minimum temperature is 270°F (132°C) for all cycles except Bowie‐Dick which is 273°F (134°C). The P2131    sterilizer setpoint temperature is 273°F (134°C) for all cycles except Bowie‐Dick which is 274°F.  2   AAMI standard 16‐towel pack (9”x9”x6”).  3   Dry times are pre‐set to established standard conditions, but can be manually increased.  4   The FDA‐cleared Process Challenge Device (PCD) for this 10‐minute cycle can be purchased from FDI at   (865) 408‐0100 or via email at sales@fortdefianceind.com.  Reference Appendix A.3 for additional information  on this PCD.    WARNING:  The P2131 sterilizer does not include a sterilization cycle for liquids. DO NOT  ATTEMPT TO STERILIZE ANY LIQUIDS.       ...
  • Page 12: Description Of Cycle Phases

    Description of Cycle Phases          Schematic of a Dynamic Air Removal (Pre‐Vac) Cycle    Below is a description of the various phases of a P2131 sterilizer textile or wrapped instrument  cycle.  NOTE: The IUSS cycle has no dry phase.    1. Preconditioning (pre‐vacuum)    a. Steam  Flush – An initial pulse of steam is released into the chamber and flows  directly out of the chamber into the WRS.  This steam flush removes most of the  air from the chamber.    b. Vacuum Pulse #1 – A medium vacuum is pulled on the chamber to remove more  air.    c. Steam  Pulse #1 – Steam is released into the chamber to increase pressure and  temperature to heat up the load.    d. Vacuum Pulse #2 – A slightly deeper vacuum is pulled on the chamber to remove  even more air.    e. Steam Pulse #2 – Steam is released into the chamber again to increase pressure  and to continue heating up the load.    f. Vacuum Pulse #3 – This is the final and deepest vacuum pulse during which nearly  all of the air is removed from the chamber.   ...
  • Page 13: Water Supply Quality

      3. Exposure – Steam flowing into the chamber slows and the temperature sensor, PLC, and  heaters begin controlling the temperature inside the chamber so that it stays within a  specified range during the entire exposure time selected.    NOTE: The P2131 is controlled by temperature, not by pressure, during exposure.    4. Drying  –  A  deep  vacuum  is  drawn  on  the  chamber  to  remove  the  steam  and  residual  moisture from the sterilized items.    5. Venting – The vacuum is released by bringing in outside air through the HEPA filter so the  chamber door can be opened by the operator.      Water Supply Quality      It is critical to the long‐term reliability and performance of the P2131 sterilizer that only potable  water (i.e., water safe for human consumption) should be used as the feedwater to the SWS. The  feedwater should meet FDA bottled water criteria for total coliform (i.e., a disinfectant such as  chlorine  should  be  present).  Reference  Appendix  A.6  for  suggested  maximum  values  of ...
  • Page 14: Sterility Assurance Products

    Sterility Assurance Products      FDI recommends either of the two options listed below to ensure successful sterilization with  each load processed. Option 1 is a combination of VERIFY™ and 3M®products, and Option 2 is  the 3M® line of products only.       VERIFY™ and 3M® System    3M® System         Reusable Process Challenge Device    P/N:  P2131‐1‐9003     1296 ‐ Biological Process Challenge Device      P/N:  P2131‐1‐9010 4‐Minute      OR  Exposure      1243 ‐ Class 5 Chemical Integrator  Class 5 Chemical Integrator    P/N:  P2131‐1‐9011 P/N:  P2131‐1‐9004          1292 ‐ Biological Indicator ...
  • Page 15: Operation

    SECTION 3 – OPERATION    The P2131 sterilizer system includes the following:  1. One SMU    (P/N: P2131‐1) 2. One WRS    (P/N: P2131‐2) 3. One SWS    (P/N: P2131‐3) 4. One SAB    (P/N: P2131‐4)                 Sterilizer Main  Unit (SMU)  Sterilizer Accessory   Bag (SAB)  Water Recovery  Sterilizer Water  System (WRS)  Softener (SWS) ...
  • Page 16 Sterilizer Accessory Bag (SAB):       The SAB is stored in the SMU chamber during transport but contains items necessary for both  setup and maintenance.  The items found in this bag include:  1. One Sterilizer Accessory Bag    (P/N: P2131‐1‐7000) 2. Three SMU to WRS connection hoses    Recovered Water Hose    (P/N: P2131‐1‐3101)  Vacuum/Exhaust Hose    (P/N: P2131‐1‐3102)  Condensate Hose    (P/N: P2131‐1‐3103) 3. One high‐temperature drain hose    (P/N: P2131‐1‐3104) 4. Six shelves    (P/N: P2131‐1‐2002) 5. Twelve leveling shims    (P/N: P2131‐1‐2016) 6. Handheld TDS (Total Dissolved Solids) Meter    (P/N: P2131‐3‐9010) 7. Twenty‐five spare hose fitting O‐rings ...
  • Page 17: Installation, Setup, And Qualification (Iq, Oq, Pq)

    INSTALLATION, SETUP, AND QUALIFICATION (IQ, OQ, PQ)        WARNING:  To  prevent  potential  injuries  during  transport,  installation  and  setup,  FDI  recommends a six‐person team lift for the SMU and a four‐person team lift for the  WRS.          1. The P2131 sterilizer is designed to be carried by 6 people.            2. Place the SMU on one of its ends.           ...
  • Page 18   3. Unlock the two latches holding the endcap against the main frame.            4. Holding securely onto the SIDES of the endcap, slowly pivot the endcap to the fully open  position (do not allow endcap to drop).      WARNING:  A potential pinch point exists near the hinge area of the endcap.  It is very      important to keep hands positioned on the SIDES of the endcap and away      from the pinch point.            Placing your hand  Proper hand  near the hinge can result  placement is on the side  in injury.  Keep hands  of the end cap, away  positioned on the sides  from the pinch point.   of the endcap and away  This will help avoid injury  from the pinch point.  when opening the  endcap.        5. On each side, screw the black endcap knobs down securely. Ensure the endcap knobs are  fully seated by checking to see if the white washer in between the knob and the spring does  not spin. Do not overtighten.     ...
  • Page 19 6. Tilt the SMU down onto the endcap that was just opened.          7.  Position  one  person  on  each  side  of  the  SMU.    Using  the  side‐carry  handles,  lift  the  unopened end of the unit, pivoting on the endcap that is already opened and secured. As  the SMU is raised into position, rest the weight of the unit on your thigh, allowing you to  unlock the two latches holding the endcap against the main frame.  If a third person is  available  to  help,  they  should  be  positioned  at  the  end  of  the  SMU  to  help  unlock  the  latches.   ...
  • Page 20 WARNING:  A potential pinch point exists near the hinge area of the endcap.  It is       very important to keep hands positioned on the SIDES of the endcap.               9.  Once  the  endcap  is  lowered  into  position,  screw  the  two  black  endcap  knobs  down  securely. Ensure the endcap knobs are fully seated by checking to see if the white washer  in between the knob and the spring does not spin. Do not overtighten.        10. Remove the SAB from the chamber of the SMU. The SAB contains items necessary to  complete the assembly of the unit. Refer to Section 3.2.1 – Chamber Door Operation for  instructions on operating the chamber door. The multi‐tool is stored in the rear endcap  and will also be used (for S/N’s AFS‐0001 through AFS‐0215, the multi‐tool is in the SAB).     Multi‐tool    (P/N: P2131‐1‐2020) ...
  • Page 21 11. Remove left side panel of SMU with the multi‐tool supplied in the rear endcap leg and  perform visual inspection. On S/N’s AFS‐0001 through AFS‐0215, the multi‐tool is stored  in the SAB. Ensure that:  a. HEPA filter cartridge is not damaged.  b. All wiring, hoses, and fittings are secure.  c. Insulation is not damaged.  d. There are no other visible issues or concerns.    CAUTION: Hold panel firmly in place when loosening the panel latches.    12. Remove right side panel of SMU (side with inlet air filter to the electrical enclosure) and  perform visual inspection to ensure that:  a. There are no loose or disconnected wires.  b. There are no damaged or loose components.  c. The fuses are in place and not damaged.    CAUTION: Hold panel firmly in place when loosening the panel latches.    13. Remove top panel and front/back panels of the WRS to ensure that:  a. There are no loose or disconnected wires.  b. There are no damaged or loose components.    14. Clean and wipe down the unit as needed and reinstall all panels.    15. Place the WRS next to the SMU, orienting the unit such that the side of the WRS with the  power switch and sight glass are on the right side of the SMU when facing the front of the  SMU.  Then, slide the WRS under the SMU. ...
  • Page 22   16. Remove the lower front panel of the SMU to access the connection for the condensate  hose.  Use the multi‐tool provided in the SAB to loosen the front panel latches, and then  gently remove the panel.        17. On the bottom left side of the SMU is an access hole for the condensate hose (red‐tagged).   Push the hose through this hole and unscrew the cap on the left side of the drain manifold.   Ensure the O‐ring on the manifold fitting is in place, and then connect the hose to the  condensate outlet on the chamber drain manifold.  Hand‐tighten the nut and then tighten  45° (max) using the multi‐tool.    NOTE: If the fitting O‐ring is missing, spare O‐rings are provided in the SAB.                        ...
  • Page 23   18. Replace the lower front panel.    19. On the WRS, unscrew the “Condensate from Sterilizer (IN)” bulkhead cap, and verify that  the O‐ring on the bulkhead fitting is in place.  Attach the red‐tagged condensate hose.   The  connection  has  a  red  bulkhead  retainer  to  match  the  red  tag  on  the  hose.  Hand‐ tighten the nut and then tighten 45° (max) using the multi‐tool.    NOTE: If the fitting O‐ring is missing, spare O‐rings are provided in the SAB.                            20.  On the back of the SMU, unscrew the “Vacuum/Exhaust to WRS (OUT)” bulkhead cap,  and verify that the O‐ring on the bulkhead fitting is in place.  Attach the white‐tagged  vacuum/exhaust hose. The connection has a white bulkhead retainer to match the white  tag on the hose. Hand‐tighten the nut and then tighten 45° (max) using the multi‐tool.    NOTE: If the fitting O‐ring is missing, spare O‐rings are provided in the SAB.              ...
  • Page 24   21. On the WRS, unscrew the “Vacuum/Exhaust from Sterilizer (IN)” bulkhead cap, and verify  that  the  bulkhead  fitting  O‐ring  is  in  place.    Attach  the  white‐tagged  vacuum/exhaust  hose.  The connection has a white bulkhead retainer to match the white tag on the hose.  Hand tighten the nut and then tighten 45° (max) using the multi‐tool.    NOTE: If the fitting O‐ring is missing, spare O‐rings are provided in the SAB.        22. On the back of the SMU, unscrew the “Recovered Water from WRS (IN)” bulkhead cap,  and verify that the bulkhead fitting O‐ring is in place.  Attach the blue‐tagged recovered  water hose. On the WRS, unscrew the “Recovered Water to Sterilizer (OUT)” bulkhead  cap,  and  verify  that  the  bulkhead  fitting  O‐ring  is  in  place.  Attach  the  blue‐tagged  recovered water hose.  These connections both have blue bulkhead retainers to match ...
  • Page 25   23. The final hose configuration is pictured below.      Red‐tagged  Condensate Hose  (P/N: P2131‐1‐3103)  Blue‐tagged  White‐tagged  Recovered Water Hose  Vacuum/Exhaust Hose  (P/N: P2131‐1‐3101)  (P/N: P2131‐1‐3102)      24. Unscrew the SMU power cord receptacle from its holder and set to the side.     CAUTION: Do not connect this receptacle to a power source yet.    SMU Electrical  Power Cord  Connecting  Receptacle  (P/N: P2131‐1‐4027)   ...
  • Page 26   25. Remove the WRS Power Cord Receptacle  from its holder and connect the  (P/N: P2131‐1‐0014)  receptacle into the WRS Main Power Plug  on the WRS.  (P/N: P2131‐1‐4010)    WRS Power Cord  WRS Main Power Plug  Receptacle  (P/N: P2131‐2‐4010)  (P/N: P2131‐1‐0014)                      26. Ensure all three manual valves on the back of the SMU are closed firmly.      Jacket Vent  Chamber Vent      Jacket Drain     ...
  • Page 27 27. For S/N’s AFS‐0001 through AFS‐0203, open the chamber door, and insert the chamber  drain screen in the front of the chamber. The screen is in the large plastic bag found in  the  SAB.  For  S/N’s  after  AFS‐0203,  verify  the  drain  screen  is  already  installed  in  the  chamber.  Do NOT operate the P2131 sterilizer without the chamber drain screen.    NOTE:  Remove the screen by rotating and pulling the screen upwards. The screen is not  designed to be tightened. For S/N’s AFS‐0001 through AFS‐0203, the screen has  an outward notch on the collar. For S/N’s after AFS‐0203, the drain screen has an  O‐ring that locks the screen into the coupling in the bottom of the chamber.         28. For S/N’s after AFS‐0236, the following step can be skipped. For S/N’s AFS‐0001 through  AFS‐0236, first remove Back Lower Panel and black Heater Assembly Cover. Reinstall the  two Heater Flange Drain Caps at the rear of the unit by removing the caps from Tee Cap  Retainers and tightening them to the Heater Flange Tees finger tight, then only one  quarter‐turn with wrench. Reinstall Back Lower Panel and Heater Assembly Cover.    Tee Cap Retainers (S/N’s AFS‐0001 – 0236)       Heater Flange Tees  ...
  • Page 28   30. Using the shims provided in the side zipper section of the SAB, elevate the rear of the  SMU to ensure there is a slight decline from the back of the unit toward the front to allow  the condensate in the chamber to drain properly. From a level starting point, this will be  approximately 4 shims or 1” under the rear endcap. The target angle is 1.5°.                       31. To test for proper angle, pour a cup of water in the back of the chamber, and ensure that  it flows to the chamber drain.      32. Clean the chamber, and then install the trays found in the SAB.      CAUTION:  Do not allow sand, chemicals, or other debris to flow down the chamber drain  into the WRS.  This could contaminate the water or damage components in the  WRS.           33. Ensure that the power switches for the SMU and WRS are in the OFF position.     ...
  • Page 29 See  Section  2.5  –  Water  Supply  Quality  to  ensure  the  correct  quality  of  feedwater is used.    CAUTION: Water in SWS can foul and become discolored if not used regularly.   Outlet Hose (To WRS)  Water Supply Hose     (P/N: P2131‐3‐3006)   36. For  S/N’s  AFS‐0001  through  AFS‐0236,  with  the  yellow  valve  handle  as  shown  below,  connect the clear Outlet Hose from the SWS to the Fill Connection on the bottom of the  WRS.  This is a cam‐style connection.  Ensure WRS drain valve is closed, then open the  yellow valve on the end of the SWS outlet hose (pictured below) to allow water to flow  into the tank. Once the HIGH mark is reached, close the yellow valve at the end of the  SWS outlet. ...
  • Page 30 NOTE: If a pressurized water source is not available, the manual hand pump mounted on the  SWS can be used to pump potable water from a bucket or Jerry can. Before each use, flush SWS  until water is clear (2 minutes).     Release the two Handle Latches and fold the handle down to expose the manual hand  pump and hoses. Connect the Suction Line and drop strainer into a Jerry can or bucket.  Handle Latches  Suction Strainer  Manual Hand Pump  (P/N: P2131‐3‐3009)  (P/N: P2131‐3‐0003)  Suction Line to Manual Hand Pump    (P/N: P2131‐3‐3007)      Hook hand pump Discharge Line to Inlet of SWS.  Discharge Line to Inlet of SWS    (P/N: P2131‐3‐3008)      Push/pull the white handle back and forth to pump water into the WRS. Once the HIGH  mark is reached, close the yellow valve at the end of the SWS outlet hose and the blue  valve on the Water Fill Connection.       ...
  • Page 31 37. Flip the power switch on the WRS to the ON position (UP position).          38. Close and tighten the chamber door.  Flip the control power switch on the SMU to the ON  position (UP position).         NOTE:  When the SMU is turned ON with no water in the jacket, a “Low Water” alarm will  occur and the WRS will turn ON. This alarm confirms that the liquid level switch is  functioning  correctly  in  the  “Low  Water”  position.  The  alarm  will  prevent  the  heater elements from energizing, but will allow operation of the WRS to fill the  jacket with water. (The “Low Water” alarm will be reset in a later step).   WARNING:  If  this  “Low  Water”  alarm  does  not  occur,  turn  the  sterilizer  OFF  immediately  to  prevent  damage  to  the  heater  elements.    Contact  a  qualified ...
  • Page 32 39. Ensure the jacket vent valve on the back of the SMU is fully open and close the chamber  vent valve.  This will allow the jacket to vent the air while it is filled with water from the  WRS.  Minimizing air in the jacket improves sterilizer performance since air acts as an  insulator during sterilization.    40. Press and hold the “Water to Sterilizer” button on the WRS and fill the SMU to the HIGH  mark on the level sight glass. This should pump about 4.25 gallons of water into the jacket  of the sterilizer. Add water to the WRS with the SWS as needed to fill the SMU to the HIGH  mark.      CAUTION:  Verify  the  SMU  sight  glass  fills  from  the  bottom.  If  not,  contact  a  qualified  maintenance technician to inspect the sight glass plumbing for debris.  CAUTION:  Never add water to the jacket during a cycle.  This will cause the jacket to cool,  lose pressure, and alarm.  41. Firmly close the jacket vent valve on the back of the SMU.    42. Refill the WRS back to its HIGH mark and disconnect the SWS hose.  NOTE: The correct water levels are HIGH/HIGH ‐ when the water is at the HIGH mark  in the jacket, the water level should be at or near the HIGH mark in the WRS.    43. Reset the “02 – LOW WATER LEVEL” alarm. If the alarm clears, the liquid level switch is  functioning correctly in the “Water OK” position.  ...
  • Page 33 “Log”  screen  after  each  cycle  for  recording  results.    This  can  be  especially  helpful  if  there  is  more  than  one  sterilizer  in  operation.    The  factory  default for this UNIT ID is 2131.  This can be changed as desired by the user.    45. Barometric Pressure Setup.  Since the P2131 sterilizer is designed for altitudes ranging  from sea level to 8,000 feet, the “BAROMETRIC SETUP” screen is used to set the gauge  pressure in the PLC to zero.     NOTE:  The  sterilizer  control  system  uses  absolute  pressure  transmitters  for  cycle ...
  • Page 34   This  test  will  ensure  that  the  air  removal  process  during the pre‐vac phase is working correctly. (Reference Section 3.3.4 for procedure.)    51. Perform  Performance  Qualification  using  PCDs  per  AAMI  ST79  standards.    Reference  Section 2.6 for information on recommended sterility assurance products.    52. After all of these tests are successful, the P2131 sterilizer is ready for OPERATION.  If any  of these tests fail, the sterilizer cannot be placed into service.  Call a qualified maintenance  technician  for  repair.  It  is  critical  to  reference  Section  3.2  to  become  familiar  with  all  Controls & Indicators before following the Operating Procedure in Section 3.3.     ...
  • Page 35: Controls And Indicators

    Controls and Indicators      Sterilizer Main Unit (SMU)        2. Control Power  4. Indicator Light (LAMP‐1)  3. Pressure Gauges    Switch (SW‐1)  (P/N: P2131‐1‐4018)  (P/N: P2131‐1‐3406)  (P/N: P2131‐1‐4017)    5.  Controller  (P/N: P2131‐1‐4000)  1. Warnings  6. Audible Alarm  (ALM‐1)   (behind panel)    (P/N: P2131‐1‐4008)  7. Door Handles  (P/N: P2131‐1‐1020)  8. Locking Arm Lever  (P/N: P2131‐1‐1023)  9. Radial Arms  (P/N: P2131‐1‐1021)      Front View of SMU   ...
  • Page 36: Chamber Door Operation

    Function(s) of Control or Indicator (Front View of SMU)    ITEM  CONTROL/INDICATOR  FUNCTION  Critical warnings for the safe operation and maintenance of the  1.  Warnings  P2131.  Be sure to reference all warnings listed in Section 1.1.  A two‐position toggle switch that energizes the control system and  2.  Control Power Switch  heater elements.   Up position is ON, and down is OFF.  3.  Pressure Gauges  Indicates gauge pressure (psig) in jacket and chamber.  Green = Successful cycle   = Jacket warming up  4.  Indicator Light  Red = Alarm  No light = Idling or in a cycle.  Touch‐screen controller for operating the P2131 sterilizer. For  5.  Controller  details on the menu, screens, and all functions, reference Section  3.3 ‐ Operating Procedure.  Buzzer/alarm to audibly notify the operator of conditions.  Single tone = successful cycle  6.  Audible Alarm  Three tones in rapid sequence = alarm notification  Sustained Tone = Low water & Liquid Level Switch malfunction or  damaged HTS‐1 RTD  Turning these handles tightens and seals the door into place.   7.   Door Handles  These handles may become warm during operation.  Pivots the radial arms to lock the door into place as well as  8. ...
  • Page 37     12b.  Tee Cap Retainer  1.  Water Level Sight  Gauge  6.  Vacuum/Exhaust  (P/N: P2131‐1‐3405)  Hose Connection  (White‐Tagged)  (P/N: P2131‐1‐9007)  9.  Jacket Drain Valve  (P/N: P2131‐1‐0039)  5.  Recovered Water  Hose Connection   (Blue‐Tagged)  12a.  Tee Cap Retainer  (P/N: P2131‐1‐9007)  10b.  Heater Flange Drain  7.  Jacket and Chamber  Cap, Water Level  Manual Vent Valve Outlets  (P/N: P2131‐1‐3016)  (P/N: P2131‐1‐9008)  8.  Jacket Drain Hose  10a.  Heater Flange  Connection  Drain Cap, Fill  (P/N: P2131‐1‐9009)  (P/N: P2131‐1‐3015)  *Note: Back Lower Panel (P/N: P2131‐1‐0139) and  Heater  Assembly  Cover  (P/N:  P2131‐1‐9041)  are  required but not shown. They are depicted on the  11.  Multi‐tool  next page.  (P/N: P2131‐1‐2020) ...
  • Page 38 Rear View of SMU for S/N’S after AFS‐0236    4.  Chamber Vent  Valve  (P/N: P2131‐1‐0034)  2.  Pressure Safety Valve   (P/N: P2131‐1‐3403)  3.  Jacket Vent Valve  (P/N: P2131‐1‐0035)  1.  Water Level Sight  Gauge  6.  Vacuum/Exhaust  (P/N: P2131‐1‐3405)  Hose Connection  (White‐Tagged)  (P/N: P2131‐1‐9007)  13.  Back Lower Panel    (P/N: P2131‐1‐0139) 5.  Recovered Water  Hose Connection   (Blue‐Tagged)  (P/N: P2131‐1‐9007)  14.  Heater Assembly  Cover  7.  Jacket and Chamber    (P/N: P2131‐1‐9041) Manual Vent Valve Outlets  (P/N: P2131‐1‐9008)  9.  Jacket Drain Valve  (P/N: P2131‐1‐0039)  8.  Jacket Drain Hose  Connection  (P/N: P2131‐1‐9009)  11.  Multi‐tool  (P/N: P2131‐1‐2020)   ...
  • Page 39   Function(s) of Control or Indicator (Rear View of SMU)  ITEM  CONTROL/INDICATOR  FUNCTION  Indicates water level inside jacket.   Fill to HIGH mark.  A low water level  1.  Water Level Sight Gauge  switch (separate from sight gauge and not shown in picture) will create  an alarm, abort the cycle, and de‐energize the heaters.  Pressure Safety Valve  Relieves overpressure @ 40 psig to prevent potential rupture of  2.  (PSV)  jacket/chamber  Manual valve that opens the jacket to atmosphere. This valve must be  closed during normal operation.  Best practice includes NOT opening  3.  Jacket Vent Valve  this valve after shutdown to vent steam, but allow SMU to cool and  steam to condense (this will pull a vacuum in jacket, which is OK).  Manual valve that opens the chamber to atmosphere.  This valve must  be closed during normal operation.  This valve is used to relieve  4.  Chamber Vent Valve  pressure or vacuum (e.g., if vacuum develops in chamber from  accidently leaving door closed during cool down).   Recovered Water Hose  5.  Connection (Blue‐ This hose supplies water to the jacket from the WRS.   Tagged)  Vacuum/Exhaust Hose  This hose exhausts steam/air and allows the WRS to create the vacuum  6.  Connection (White‐ necessary for sterilizer operation.  Tagged)  Outlet points for the Jacket and Chamber vent valves.  Jacket and Chamber  WARNING:  A small amount of hot condensate will collect above these ...
  • Page 40   1.  Inlet Air Filter    (behind panel)    (P/N: P2131‐1‐5001)    2.  Endcap Knob  (P/N: P2131‐1‐0020)      Right Side of SMU  3.  Carry Handle    (P/N: P2131‐1‐5011)          Function(s) of Control or Indicator (Right Side of SMU)    ITEM  CONTROL/INDICATOR  FUNCTION  Filters air for cooling electrical components.  Filter is  Inlet Air Filter  1.  washable and should be inspected and cleaned.   (behind panel)  Reference Section 4, Post‐Maintenance Checklist.  Located on each corner of the SMU, these are used to  secure endcap in down position during setup.  When  breaking down the sterilizer for storage or transport,  2.  Endcap Knob  the endcap knobs should be turned counter‐clockwise.   When the knobs disengage from the endcap, a spring  will pop them upwards, indicating they are free.  Recessed spring‐loaded handle used for transporting  3. ...
  • Page 41 Water Recovery System (WRS)    The WRS reclaims steam and condensate from the SMU for re‐use.  A circulation pump and heat  exchanger allow for hot steam to be cooled and condensed into water. A water eductor creates  a vacuum during the preconditioning phases to remove air and provides a deep vacuum to assist  in drying a load during the dry phase.  The WRS also provides an easy, push‐button method to  refill the SMU jacket.    NOTE:  The Power Switch on the WRS should always be in the ON position during operation.    3.  Water Return Switch (SW‐2)    (P/N: P2131‐2‐4009)        2. HIGH Water Line  4.  Power Switch (SW‐3)    (P/N: P2131‐2‐4008)                      5. Power Connection  1.  LOW Water Line    (P/N: P2131‐2‐4010)               ...
  • Page 42 Function(s) of Control or Indicator (Right Side of WRS)    ITEM  CONTROL/INDICATOR  FUNCTION  Water level in the tank is indicated by the position of a  floating ball inside the Sight Gauge tubing. On S/N’s after  AFS‐0236, a water line in the Sight Gauge tubing indicates  the water level. During operation, the water level should  be at or near the HIGH water line (when the jacket is at  1.  LOW Water Line  the HIGH mark). If the water level becomes too low, the  Low Level Switch (located underneath the tank near the  motor) will activate and prevent the pump from running  dry. If the pump starts and stops, or will not operate,  water likely needs to be added to the WRS.  Water level in the tank is indicated by the position of a  floating ball inside the Level Gauge tubing. On S/N’s after  AFS‐0236, a water line in the Sight Gauge tubing indicates  the water level. During operation, the water level should  2.  HIGH Water Line  be maintained at or near the HIGH water line. Overfilling  and running the WRS with water above the HIGH mark  (when the jacket is at the HIGH mark) may result in water  “burping” out of the relief vent or breather on top of the  WRS water tank.  A push‐button that turns ON the pump to flow water into  Water Return Switch  3.  the sterilizer when pressed.  Releasing the switch will stop  (SW‐2)  the flow of water.  A two‐position switch that turns electricity ON and OFF to  4.  Power Switch (SW‐3)  the WRS.  This switch also has a 20‐amp circuit breaker  built into the switch.  Where the power cord from the SMU is plugged into the  5. ...
  • Page 43     3.  Condensate  Hose Connection  (Red‐Tagged)    (P/N: P2131‐2‐0009) 2. Vacuum/Exhaust  Hose Connection  (White‐Tagged)    (P/N: P2131‐2‐0009) HOT CAUTION Label  (S/N’s after AFS‐0236)  1.  Recovered Water  Hose Connection   (Blue‐Tagged)     (P/N: P2131‐2‐0009)     Left Side of WRS    Function(s) of Control or Indicator (Left Side of WRS)    ITEM  CONTROL/INDICATOR  FUNCTION  Recovered Water  This hose carries water from the WRS into the jacket in  1.  Hose Connection   the SMU when the water return switch is depressed.    (Blue‐Tagged)  This hose exhausts steam/air from the SMU chamber  Vacuum/Exhaust  and connects to a water eductor inside the WRS, which  2.  Hose Connection  generates the vacuum needed for the pre‐vac and post‐...
  • Page 44 Sterilizer Water Softener (SWS)     The  SWS  is  critical  to  the  long‐term  reliability  and  performance  of  the  P2131  sterilizer.    The  purpose of the SWS is to remove ions (e.g., calcium, magnesium, and chlorides) from the water    to prevent scale buildup on the heating elements, jacket walls, and other components. 3.  Outlet Hose  1.  Inlet Hose Connection     Connection  (P/N: P2131‐3‐3004)     (P/N: P2131‐3‐3005)   2.  Pentek Big Blue, 20”x3/4”, Housing    (P/N: P2131‐3‐9013)   4.  Outlet Hose      (to WRS)    (P/N: P2131‐3‐3006)  ...
  • Page 45 Panel Fasteners    5. Spanner Wrench    (P/N: P2131‐3‐1012)   3.  Discharge Line to  Inlet of SWS    (P/N: P2131‐3‐3008)   2. Manual Hand Pump    (P/N: P2131‐3‐0003)   4. Suction Strainer    (P/N: P2131‐3‐3009)     1.  Suction Line to    Manual Hand Pump      (P/N: P2131‐3‐3007)   Rear View of SWS    Function(s) of Control or Indicator (Rear View of SWS)    ITEM  CONTROL/INDICATOR  FUNCTION  Used when manual hand pump is in service.  Water flows  Suction Line to  1.  from bucket or Jerry can through hand pump and into  Manual Hand Pump  SWS. ...
  • Page 46: Operating Procedure

    Operating Procedure      Overview of Touch‐Screen Operator Control Panel    The P2131 sterilizer is controlled with an advanced programmable logic controller (PLC)  with  a  touch‐screen  display  that  performs  pre‐programmed  sterilization  cycles  and  performance tests.  NOTE:  The  following  instructions  assume  that  the  P2131  sterilizer  has  been  set  up,  checked out, and potable water is already in the SMU and WRS.  Reference Section  3.1.    NOTE:  The “System” and “F1” through “F3” buttons are disabled. The “F4” button is only    used as part of maintenance of the PLC.      3.3.1 TURNING THE STERILIZER ON:  1.  Flip up the Control Power Switch on the front panel to turn the sterilizer ON. ...
  • Page 47: Zeroing Gauge Pressure (Barometric Pressure Setup)

      3.3.2 ZEROING GAUGE PRESSURE (BAROMETRIC PRESSURE SETUP):  Since the P2131 sterilizer is designed for altitudes ranging from sea level to 8,000 feet,  the “BAROMETRIC SETUP” screen is used to set the gauge pressure in the PLC to zero.   NOTE:  The  sterilizer  control  system  uses  absolute  pressure  transmitters  for  cycle  operation.    By  pressing  ZERO,  the  controller  knows  the  exact  atmospheric  pressure so that the chamber can be vented consistently to 0 psig (equal  to  atmospheric pressure).  This allows the chamber door to be opened easily, and  not be locked with a vacuum.    1.  From  the  “CONFIRM  SETUP”  screen,  press  the  “BAROMETRIC  PRESSURE  SETUP” ...
  • Page 48 2.  Open  the  door.    Insert items  to  be  sterilized.    Close  and  tighten  the  door.    On  the  “Main Menu” screen, choose “STERILIZER CYCLES”.    3.  On  the  “STERILIZER  CYCLES”  screen,  choose  the  type  of  sterilization  cycle  that  you  need to run.        4.  For the TEXTILES or WRAPPED INSTRUMENTS cycles, use the arrow buttons to select  the desired dry time.  For the WRAPPED INSTRUMENTS cycle, select the 4‐minute or  10‐minute exposure.  Press the “ACCEPT” button.  CAUTION: ...
  • Page 49   5.  Verify that the jacket water level is sufficient, the chamber vent valve is closed, and  the chamber door is closed and tightened. Then press “START” to initiate the cycle.  NOTE:  When  water  is  added  to  the  jacket,  there  will  be  a  slight  cooldown,  and  therefore a reduction in pressure.  The pressure in the jacket will need a few  minutes to recover when water is added.        If the jacket is not at setpoint pressure (see NOTE above), the “START” screen will  display “WARMING UP” instead of “Ready”.  If you choose “START” before the jacket  is ready, the “START” button will disappear and the display will show “Cycle Pending.”  Once the jacket is up to setpoint pressure, the cycle will begin automatically.  During the cycle, the run screen will display the total cycle time, type of cycle, phase  within the cycle, the time remaining (for exposure, drying, or vacuum testing only),  the chamber temperature, and the chamber pressure.  A cycle can be aborted at any  time by choosing the “ABORT” button.  This will stop the cycle and vent the chamber.    Total Cycle Time    Cycle Type    Phase within Cycle    Time Remaining   ...
  • Page 50: Running A Performance Test (Vacuum Leak Or Bowie-Dick Test)

    The “Log” screen will be displayed at the end of the cycle showing the results of the cycle.   A total of 40 previous “Log” entries can be viewed using the LEFT and RIGHT arrows.  The  “Log” screen can also be accessed from the “Main Menu.”        3.3.4 RUNNING A PERFORMANCE TEST:  1.  On the “Main Menu” screen, choose “TESTS”.    2.  On the “Performance Tests” screen, choose either “BOWIE‐DICK TEST” or “VACUUM  LEAK TEST.”    For a BOWIE‐DICK TEST:  Place a U.S. Food and Drug Administration (FDA)‐approved BOWIE‐DICK test pack  on  the  bottom  rack  near  the  drain  in  the  chamber.    Follow  all  test  pack  manufacturer’s  recommended  procedures.    For  example, a  warmup  cycle  (e.g.,  IUSS 4/0) should always be run prior to the BOWIE‐DICK test.    ...
  • Page 51 For a VACUUM LEAK TEST:  The sterilizer should be warmed up prior to running the VACUUM LEAK TEST by  running any cycle (e.g., Textile 4/5) before the test.    The VACUUM LEAK TEST consists of two phases.  The first phase is a settling phase  (5  minutes)  and  the  second  phase  is  the  test  phase  (15  minutes).  The  P2131  sterilizer  will  conduct  both  phases  automatically,  and  a  countdown  timer  will  indicate the time remaining for each phase.  At the end of a successful test, the  “Log” screen will appear.  An “ALARM” screen will appear if unsuccessful.     WARNING:  A VACUUM LEAK TEST should be performed when the sterilizer is set up at a    new location or after any major service or maintenance.  In a hot environment    the test may take longer because the WRS water may be hot, which affects the  eductor’s ability to pull a deep vacuum.  Perform the test when the WRS water   ...
  • Page 52: Alarms

    During  the  cycle,  the  “RUNNING”  screen  will  display  the  total  cycle  time,  type  of  test,  where it is in the cycle, the time remaining (during exposure or vacuum testing only), the  chamber temperature, and the chamber pressure.  A test can be aborted at any time by  choosing the “ABORT” button.  This will stop the test and vent the chamber.           7.  At the end of the test, the buzzer will sound and the chamber will vent.  Once the  chamber is completely vented, the door can be opened.    NOTE:  Unlocking  the  door  (which  actuates  the  door  switch)  before  the  chamber  reaches 0 psig will stop the venting and may create a situation in which the  door may not open due to vacuum in the chamber.  If this happens, lock the  door and press “VENT” from the “Log” screen. ...
  • Page 53: Maintenance Reminder

    Choosing to “RESET” will either (1) switch the display to the “Log” screen or (2) repeat the  same alarm if the condition is still present.  The controller will not allow the sterilizer to  continue until the alarm has been reset.  For example, if the alarm was “02‐Low Water  Level,” water will need to be added to the jacket before the alarm will reset.  The  controller  will  then  display  the  “Log”  screen  with  the  details  of  the  alarm  for  recordkeeping.       3.3.6 MAINTENANCE REMINDER:  Per FDA regulations, every 100 cycles or 14 calendar days, the water in the entire system  must be drained and refilled. After every 100 cycles, a “Maintenance Reminder” screen is  shown to remind the operator that the water in the SMU jacket, the WRS, and the SWS  must be drained and replenished with fresh potable water. If needed, flush the SWS until  clear (2‐minute flush).   This reminder also serves to help encourage maintenance technicians to check and follow  the recommended maintenance schedule (Reference Section 4).      3.3.7 STATUS:    The “STATUS” screen can be accessed (except when a cycle is running) in order to look at  data such as jacket pressure, chamber pressure, chamber condensate temperature (both  RTDs), whether the door switch is showing open or closed, if the water level is OK or low,  and how many total cycles the unit has run.      ...
  • Page 54: Programmable Cycles

        3.3.8 PROGRAMMABLE CYCLES:    WARNING:  Only the standard AAMI cycles listed in Section 2.3 have been cleared by the  FDA.  Programmable  cycles  have  not  been  validated.  The  FDA  highly  recommends  that  programmable  cycles  be  used  for  research  purposes.  The  user accepts all risks associated with sterilizing any load using a programmable  cycle.      Programmable  cycles  allow  the  user  to  run  cycles  with  exposure  times  that  can  range  from 15 minutes to 40 minutes in increments of 5 minutes. Dry times can range from 0 to ...
  • Page 55     4.  After  checking  the  water  level,  ensuring  the  chamber  vent  valve  is  closed,  and  tightening the door, press “START” to begin the cycle.          5.  At the end of the cycle, the buzzer will sound and the chamber will vent. Once the  chamber is completely vented, the door can be opened.     NOTE:  Unlocking  the  door  (which  actuates  the  door  switch)  before  the  chamber  reaches 0 psig will stop the venting and may create a situation in which the  door may not open due to vacuum in the chamber.  If this happens, lock the  door and press “VENT” from the “Log” screen.   ...
  • Page 56: Operational Checklists

    3. Test  the  TDS  (Total  Dissolved  Solids)  of  the  water  and  replace  the  deionized  water  cartridge as needed. (Reference Section 3.4.1 Testing Total Dissolved Solids and Section  4.3.4 SWS Cartridge Replacement)    4. Follow  every  step  of  Section  3.7  ‐  Checklist  for  Transport  and/or  Storage  to  prevent  storage‐related issues upon startup.    Why the SWS is Important:    Preventing scale buildup (buildup of mineral deposits) is critical for ensuring that the P2131  sterilizer continues to operate reliably for many  years.  Preventing scale buildup will keep  heat‐up times short. In addition, scale buildup can cause small pieces of scale to plug valve  orifices and fittings.  Lastly, if deionized water from the SWS is used, the sterilizer jacket and  heating elements will not require costly labor and downtime for cleaning.  ALWAYS ensure  that the SWS is used for every fill‐up to the WRS to keep minerals out of the entire system.         ...
  • Page 57 ROUTINE STARTUP CHECKLIST  CHECK  OPERATION  1. Remove and clean the chamber drain screen. (Inspect O‐ring for    potential damage on S/N’s after AFS‐0203. Replace with spare  from SAB if necessary).    2. Ensure chamber and jacket vent valves are closed.    3. Flip the control power switch to the “ON” position.  4. Fill the jacket with water to the HIGH mark using the WRS.    NOTE:  Never add water to the sterilizer during a cycle.  This will    cause the jacket to cool off, lose pressure, and alarm.  5. Ensure the water level in the WRS is at, or near, the HIGH mark.     If not, add water using the SWS.  6. Run an IUSS cycle to warm up the chamber.      NOTE:  It is critical to run this warmup cycle to ensure successful    completion of the Bowie‐Dick test.  7. Run a Bowie‐Dick test per AAMI standards to ensure that air    removal during the preconditioning phase is functioning  properly.    8. Proceed with a normal sterilization cycle as needed.      WARNING:  Run a VACUUM LEAK TEST on a periodic basis per AAMI standards or  at least weekly and always after any transport, relocation, or ...
  • Page 58 ROUTINE SHUTDOWN CHECKLIST  CHECK  OPERATION  1. Add water to the jacket to the HIGH mark.  This will help (1) cool  down the jacket and (2) ensure that the jacket is full of water and    ready to run again when needed.  Also, “top off” the WRS water  level to the HIGH mark using the SWS if needed.  2. Leave the chamber door latched but not tightened to prevent a  vacuum from being created in the chamber during cooldown. To    break a vacuum in the chamber and open the door, press the  “VENT” button on the control panel touch screen.  3. Turn OFF SMU control power switch.  DO NOT open the jacket  vent valve.  NOTE:  As the SMU cools down, the steam in the jacket will  condense, resulting in a deep vacuum.  This condition is not a    problem and will not cause any damage to the SMU.  In fact, this  is a recommended practice since relieving the vacuum by  opening the jacket vent valve will allow air to enter the jacket.    Keeping air out of the jacket will enable the next cycle to operate  more efficiently.  Air is the enemy of effective sterilization.          WARNING: If freezing temperatures are possible around the sterilizer, all water from  the SMU, WRS, and SWS must be drained to prevent possible freezing  and  damage  to  the  pressure  vessel,  pump,  SWS  housing,  or  other  components.  Reference Section 3.7 for information on draining water. ...
  • Page 59 DAILY OPERATOR CHECKLIST    CHECK  TASK  1. Remove and clean the chamber drain screen (Inspect O‐ring for    potential damage on S/N’s after AFS‐0203. Replace with spare  from SAB if necessary).  2. Wipe the chamber interior surfaces and shelves.       WARNING: Do not allow any contaminants to go down the  chamber drain and into the WRS.    3. Look and listen for possible water and steam leaks.    4. Verify that the cooling fans in the SMU (qty: 1) and the WRS   (qty: 2*) are running.       NOTE:  On S/N’s after AFS‐0236, there is only 1 cooling fan  mounted to the Back Panel of the WRS.  5. Monitor water levels in the SMU jacket and WRS.      NOTE:   When the SMU jacket is at the HIGH mark, the WRS    water level should be at or near the HIGH mark also.    REMEMBER:  Always use the SWS when adding water.  6. Listen for any abnormal sounds during operation.  Notify a    qualified maintenance technician if there is a concern.     ...
  • Page 60 FDA‐REQUIRED WATER CHANGEOUT  EVERY 14 DAYS OR 100 CYCLES  (WHICHEVER OCCURS FIRST)    CHECK  TASK  1. Clean the SMU chamber thoroughly using a mild detergent:  a. Disconnect the condensate hose (red‐tagged) at the WRS  and place the end into a bucket.   b. Drain any cleaning water / mild detergent into the bucket.     c. Flush the hose with fresh water and then reconnect the  hose to the WRS.    WARNING: Do not allow any water or detergent to drain into  the WRS.    2. Drain all the water from the SMU jacket by following the Jacket  Blowdown Procedure in Section 4.1.2.       WARNING:  To avoid a potential burn hazard, ensure the high‐ temperature drain hose is used.     3. Drain the WRS by attaching the drain hose and opening the drain    valve.  4. SWS maintenance:  a. Test the TDS level of the water coming out of SWS and  replace the deionized water cartridge as needed.   Reference Section 3.4.1 Testing Total Dissolved Solids and   ...
  • Page 61: Testing Total Dissolved Solids (Tds)

    (e.g., every 14 days or 100 cycles, whichever occurs first).  Follow the instructions below.    Cartridge Replacement Criteria: If SWS outlet water TDS is ≥ 15 ppm, replace the SWS DI Water  Cartridge. If the TDS reading is < 15 ppm, the SWS can be used as normal.    NOTE: If the feedwater is of poor quality and has a high TDS level, a brand‐new cartridge  may result in a TDS reading of greater than 15 ppm. To improve the TDS reading:  1. Improve incoming water quality.  2. Use SWS’s in series to lower the TDS level before filling WRS.  3. Using a collection bucket or Jerry can, reprocess the water twice through the SWS  before filling the WRS    NOTE: The meter is not water‐tight. Do not dip the meter beyond the water level limit.    Handheld TDS Meter  P/N: P2131‐3‐9010      TDS Handheld Meter Calibration Check (every 6 months):   Immerse the meter into distilled water and verify it reads ≤ 5 ppm. If it reads > 5 ppm, either the  distilled water could be contaminated, or TDS Meter is out of calibration and should be replaced.    Taking a TDS Measurement:  1. Remove the protective cap.  2. Turn the TDS Meter ON.  3. Collect a sample of SWS outlet water in a clean container.  4. Immerse the Meter into the sample up to the Water Level Limit.  5. Lightly stir the Meter in the water to dislodge any air bubbles.  6. Wait until the reading stabilizes and then press the HOLD button. The value displayed is  the TDS Level of the water.    ...
  • Page 62: Testing Free Chlorine Level

    If  no  free  chlorine  is  present,  water  can  foul.  Free  chlorine, in water, will completely evaporate within 24 hours and the evaporation rate increases  when the water is exposed to the atmosphere. Trace amounts of free chlorine will be filtered out  through the SWS Deionizing (DI) cartridge.    Using the free chlorine test strips provided in the SAB, test the supply water and SWS outlet water  stream every time the SMU jacket and WRS are drained and refilled (e.g., every 14 days or 100  cycles, whichever occurs first). Follow the instructions listed on the chlorine test strip container  to test the Free Chlorine Level.     Chlorine Test Strip  P/N: P2131‐3‐9007      Test the supply water and outlet water stream as shown below:    3. Connect the  Outlet Hose  1. Locate and test the  Connection to  supply water with  the SWS  the test strip ...
  • Page 63: Operational Situations And Responses

    Section 3.5   Operational Situations and Responses      SMU – Testing Related Situations  What do you do if …  Response  A biological indicator fails but  Reprocess per latest edition of ANSI/AAMI ST79.  the cycle ran successfully with  no alarms?  The Bowie‐Dick Test fails?  Reprocess per latest edition of ANSI/AAMI ST79.    Was a warm‐up cycle (IUSS) run to ensure that the chamber  was heated up?  If NO, then run the Bowie‐Dick test again.  If  YES, then call a qualified maintenance technician to  troubleshoot.  The Vacuum Leak Test fails?  Take the sterilizer out of service and contact a qualified  maintenance technician who will use the Vacuum Leak Test  troubleshooting guide in Section 4.3.14.    SMU – Cycle and PLC Related Situations  What do you do if …  Response  The unit shuts off and emits a  Turn the SMU OFF and disconnect power. Contact a qualified  sustained alarm tone?  maintenance technician to troubleshoot the system.   You experience a power  Flip the control power switch to the OFF position.  When the  failure during operation?  power returns, flip the control power switch ON.  If the power  outage was short, there will still be pressure in the jacket and  pressure or vacuum in the chamber. Press “CONFIRM” on the  “SETUP” screen, then press “VENT.”  This will vent the pressure  or vacuum in the chamber.  After venting, open the door and  reprocess the load per AAMI ST79 standards.  You want to stop a cycle?  Press “ABORT.” ...
  • Page 64 SMU – Water and Steam Related Situations  What do you do if …  Response  You hear or see a steam leak  If it is safe to do so, turn the control power switch OFF and  somewhere other than the  allow the sterilizer to cool and depressurize.  Contact a  door?  qualified maintenance technician.  You hear or see a steam leak  If it is safe to do so, wait for the cycle to finish and then inspect  at the door?  the door frame and door gasket for any debris that could  prevent a good seal.  Closely inspect the gasket for any damage  such as nicks or cuts.  Replace it if needed.  If there is no  damage, then tighten the door a little tighter on the next cycle.  You get a low water alarm?  Add water to the SMU from the WRS, close and tighten the  chamber door, press “RESET,” and restart the cycle.  Also,  recheck level in WRS after pumping and refill using the SWS if  needed. It is best to run the WRS at the HIGH water mark  (when the jacket is at the HIGH mark).  The jacket is overfilled with  Turn the sterilizer OFF and let it cool down and depressurize.   water?  With zero pressure on the jacket, attach the high‐temperature  hose and drain enough water out of the jacket to bring the level  down to the HIGH mark on the sight gauge.  Close the drain  valve and turn the sterilizer back ON.    The jacket fills with water, but  Contact a qualified maintenance technician for further  VERY slowly?  diagnosis and repair on the WRS pump.  The jacket does not refill  Verify the water level in the WRS, adding water through the  when switch on WRS  SWS if necessary.  Ensure that the power cord is plugged into  depressed?  the WRS and that the power switch is in the ON position.  If the  problem continues, contact a qualified maintenance technician  for further diagnosis and repair. ...
  • Page 65 What do you do if …  Response  Water is seen entering the  Contact a qualified maintenance technician to inspect and clean  SMU sight glass through the  the Liquid Level Switch and Sight Glass plumbing. Address any  top during jacket fill‐up  blockages or issues found.     SMU – General Situations  What do you do if …  Response  There is a vacuum in the  No response needed.  This is normal.  The sterilizer can be  jacket at startup?  turned on as usual.  There is algae in the sterilizer  No response needed.  Algae forms due to heat and light.  It is  sight glass?  normal and will not affect sterilization.  To clean the sight glass,  follow the procedure found in Section 4.3.3.  The chamber door will not  Press “VENT” on the “Main Menu” or the “Log” screen.  If you  open due to vacuum on the  still cannot open the door, open the chamber vent valve in the  chamber?  back of the SMU to vent the chamber.  You hear an abnormal sound?  Take the sterilizer out of service and contact a qualified  maintenance technician.  During exposure, the two  One of the gauges is likely malfunctioning.  Contact a qualified  pressure gauges are  maintenance technician.  different?  The chamber is failing to pull  Contact a qualified maintenance technician.  vacuum during pre‐vac or  drying?    WRS and SWS – General Situations ...
  • Page 66: Alarm Code Troubleshooting Guide

    Alarm Code Troubleshooting Guide    WARNING:  Service and maintenance work must be performed by qualified personnel only.      ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Liquid Level Switch  Turn the SMU OFF and disconnect power. Inspect Liquid  Malfunction & Unit  Level Switch. Does float move freely? Electrical continuity  run with low or no  of Liquid Level Switch with water? No continuity without  water  water? Replace liquid level switch or fix root cause of the    malfunction. Re‐qualify the switch per Operational  Qualification (OQ) of the Liquid Level Switch before placing      the unit back into service. See Section 3.1.    Sustained      Alarm Tone   If Liquid Level Switch is functional, inspect plumbing that  Plug / debris in  &  leads to switch, including the heater flange tees, the hoses  Liquid Level Switch  PLC Off  and any other fittings. Remove blockage and flush hoses.  Plumbing & Unit    Re‐qualify the switch per Operational Qualification (OQ) of  run with low or no    the Liquid Level Switch before placing the unit back into ...
  • Page 67 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Jacket manual vent  Close jacket vent valve.  valve open        03  Warm up ambient conditions to 40°F, if possible.  Low ambient  temperatures  Jacket      Warmup    Heater element  Contact a qualified maintenance technician to inspect and  Slow  failure  replace heater element.          Contact a qualified maintenance technician to inspect  Scale buildup on  heaters and jacket walls and clean as needed.  Ensure SWS  heater elements  is being used on every water refill.  and jacket  Steam leak  Inspect door gasket and chamber fittings for steam leaks.      Faulty steam trap  Ensure trap is working properly. If steam trap is failed  closed, condensate will pool in chamber. If steam trap fails ...
  • Page 68 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Steam leak  Inspect door gasket and chamber fittings for steam leaks.      Faulty steam trap  Ensure trap is working properly. If steam trap is failed    closed, condensate will pool in chamber. If steam trap fails    open, steam will flow to WRS water tank and make a loud    popping noise or use more water than normal.    Replace as needed.      RTD or RTD  Contact a qualified maintenance technician to inspect RTD  transmitter failure  and RTD transmitter for physical damage, loose wires,    broken communication cable to PLC.  05    Compare temperature versus pressure on STATUS screen.   Long    Replace as needed.  Exposure      Time  Condensate drain  Remove and clean screen.  screen plugged     ...
  • Page 69 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Loose  Inspect O‐rings at each end and re‐tighten the hose.  vacuum/exhaust    hose fittings (white‐   tagged hose)        Pump not operating  Check WRS water level, power cord, and power switch.  in WRS        Hot water in WRS  Ensure both WRS fans are running when pump is ON.     Ensure water level is at least 1/2 full.  Too little water will  06    tend to overheat and cause low eductor performance.  Insufficient      Vacuum  Eductor problem in  Contact a qualified maintenance technician to inspect  WRS  eductor for leak or worn/broken nozzle inside eductor.     Replace if needed.      Vacuum valve not  Contact a qualified maintenance technician to inspect ...
  • Page 70 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Gasket leak at door  Inspect for leak with pressure and identify location.     Inspect gasket for debris or cut.  Replace as needed.      Manual chamber  Contact a qualified maintenance technician to inspect vent  vent valve leaking  valve and replace as needed.      Faulty steam trap  Ensure trap is working properly. If steam trap is failed    closed, condensate will pool in chamber. If steam trap fails    open, steam will flow to WRS water tank and make loud    popping noise or use more water than normal.  08    Replace as needed.  Chamber      Pressure  Vacuum/Exhaust  Contact a qualified maintenance technician to inspect  Low  valve leaking  outlet of valve for leaking.      Sterilize valve  Contact a qualified maintenance technician to inspect  failure – not ...
  • Page 71 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Air compressed in  Turn OFF sterilizer and allow jacket to cool and  jacket during water  depressurize.  Open jacket vent valve to bring pressure to  fill  zero.  Close jacket vent valve.  Turn sterilizer back ON.      RTD or RTD  Contact a qualified maintenance technician to inspect RTD  transmitter failure   and RTD transmitter for physical damage, loose wires,    broken communication cable to PLC. Compare    temperature versus pressure on STATUS screen.  Replace    as needed.      09  SMU tilt is not steep  Shim the rear of the sterilizer higher to improve  Jacket  enough  condensate flow to the chamber drain.  Pressure      High  Condensate drain  Remove and clean screen. Replace after cleaning.  screen plugged        Solid state heater ...
  • Page 72 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Leak in chamber or  Contact a qualified maintenance technician to inspect  11  associated fittings  solenoid valves and manual valves for leaks. Inspect fittings  and valves,  for leaks.  Inspect vacuum hose (white‐tagged hose) to  Long Dry  including vacuum  WRS for leaks.  Repair as needed.  Time  hose to the WRS  Leak in chamber or  Contact a qualified maintenance technician to inspect  associated fittings  solenoid valves and manual valves for leaks. Inspect fittings  and valves,  for leaks.  Inspect vacuum hose (white‐tagged hose) to  12  including vacuum  WRS for leaks.  Repair as needed.  Vacuum Test  hose to the WRS    Failed  NOTE: Reference Appendix A.5 Vacuum Leak Test  troubleshooting guide for more help on troubleshooting a  potential leak in the system.  Overcurrent  Contact a qualified maintenance technician to correct  cause of overcurrent and replace fuse F3.  13      Fuse F3 ‐ Inputs  Faulty fuse ...
  • Page 73 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  RTD or RTD  Contact a qualified maintenance technician to inspect RTD  transmitter failure  and RTD transmitter for physical damage, loose wires,    broken communication cable to PLC.    Compare temperature versus pressure on STATUS screen.     Replace as needed.      Faulty steam trap  Ensure trap is working properly. If steam trap is failed    closed, condensate will pool in chamber. If steam trap fails    open, steam will flow to WRS water tank and make loud    popping noise or use more water than normal.  16    Replace as needed.  Chamber      High  Air in jacket  Reset and rerun cycle.  Temperature      Condensate drain  Remove and clean screen. Replace after cleaning.  screen plugged     ...
  • Page 74 ALARM  POSSIBLE CAUSES  CORRECTIVE ACTION  CODE  Large RTD  Contact a qualified maintenance technician to inspect RTD  deviation,  and RTD transmitter for physical damage, loose wires,  21  broken communication cable to PLC.    Temperature  Compare RTD‐1 versus RTD‐2 on “STATUS” screen.  Sensor  RTD or RTD  transmitter failure  Replace as needed.  RTD or RTD  Contact a qualified maintenance technician to inspect RTD  transmitter failure  and RTD transmitter for physical damage, loose wires,    broken communication cable to PLC.    Compare temperature versus pressure on STATUS SCREEN.     Replace as needed.      Faulty steam trap  Ensure trap is working properly. If steam trap is failed    closed, condensate will pool in chamber. If steam trap fails    open, steam will flow to WRS water tank and make loud  22    popping noise or use more water than normal.  High    Replace as needed. ...
  • Page 75: Checklist For Transportation And Storage

    Section Checklist for Transport and/or Storage          CHECK  OPERATION  Ensure that the water in the WRS is at the HIGH mark (add some    water if needed using the SWS).    Remove the top panel from the WRS.  Pour in 1/2 gallon of household chlorine bleach (3‐6% sodium  hypochlorite) to the bleach port of the WRS water tank. Reference    Section 4.5.   CAUTION: Do not pump water into the SMU until bleach has been  rinsed from the WRS.  Run a Textile 4/5 cycle. The 4/5 cycle has a total cycle time of over  20 minutes, providing continuous water circulation in the WRS to    ensure complete contact and effective disinfection. Drain the WRS  and refill with deionized water from the SWS. Run a Textile 4/5  cycle. This will help rinse any remaining bleach solution.  Perform the Jacket Blowdown Procedure in Section 4.1 to remove  any contaminants that may have accumulated in the jacket. Allow    the SMU to cool to ambient temperature. Remove the drain hose.  WARNING: Do not blow out with compressed air.  Confirm the SMU power switch and the WRS power switch are in    the OFF position.    Disconnect the WRS power cord and store in SMU endcap.    Disconnect the SMU power cord and store in endcap.  Ensure the jacket and chamber are at 0 psig. The jacket and    chamber vent valves should be left open for storage.    10.
  • Page 76 Checklist for Transport and/or Storage (continued) CHECK  OPERATION    14. Pull WRS out from under SMU.  15. Connect high‐temperature drain hose to WRS Tank Drain  Connection.  Open valve and allow WRS to drain fully. Lift and tilt    WRS repeatedly to ensure all water is drained.  The Tank Drain Valve  should be left open for storage. Remove the drain hose.  WARNING: Do not blow out with compressed air.  16. For S/N’s AFS‐0001 through AFS‐0236, loosen the four panel latches  and remove WRS front panel.  Open the petcock valve (do not  remove completely) on bottom of pump and allow to drain fully. Tilt  the WRS back and forth repeatedly to ensure all water is drained.  Close petcock valve and reinstall panel.      For S/N’s after AFS‐0236, open the Pump Drain Valve. Tilt the WRS  back and forth repeatedly to ensure all the water is drained. The  Pump Drain Valve must be left open for storage. Open the Water Fill  Valve for storage.  WARNING:  Water remaining in pump could freeze and crack the  pump or cause it to seize.    17. For S/N’s AFS‐0001 through AFS‐0236, remove Back Lower Panel and  the black Heater Assembly Cover (reference Section 4.3.17) and then  drain the Heater Flange Tees by removing the two Tee Caps on the  back of the SMU (reference Section 3.2 ‐ Controls and Indicators –  Rear View of SMU for location). Store the caps by screwing them    onto the Tee Cap Retainers on the back of the unit finger tight.  Reinstall the Heater Assembly Cover and Back Lower Panel    For S/N’s after AFS‐0236, the Heater Flange Tees no longer exist, and  the unit will naturally drain through the Jacket Drain Valve.    18.
  • Page 77   Checklist for Transport and/or Storage (continued)   CHECK  OPERATION  21. Verify that all items are in the SAB and then place the SAB inside the    chamber of the SMU.  Close and tighten the chamber door. Leave the  door snug but not fully tightened.    22. Reference steps 1 through 9 from Section 3.1 ‐ Installation and Setup    to fold up endcaps on SMU.  Using light oil, lubricate endcap hinges  and threaded portion of endcap knobs.  23. Perform TDS water quality test to ensure SWS is correctly deionizing    the water. Replace cartridge as needed. See Section 3.4.1.  24. Remove the inlet and outlet hoses on SWS.  Remove the blue  housing and turn upside down to allow the water to drain completely  from the inlet/outlet.  Place back into frame, connect outlet hose,  and ensure manual valve is closed.  Connect manual pump outlet  hose to inlet of SWS. Wrap and secure all hoses for transport.      NOTE: On S/N’s after AFS‐0255, all hoses can be stored in the SWS  frame. Release the Handle Latches to fold the handle down and store  all hoses inside the frame. Close the Handle and secure the two  Handle Latches.     ...
  • Page 78: Consumable Items List

    Section 3.8   Consumable Items       The following operator supplies are typically consumed during standard use of the P2131  system. For replacement parts, see Section 4.4.    ITEM  DESCRIPTION  PART NUMBER  1.  VERIFY™ REUSABLE PROCESS CHALLENGE DEVICE  P2131‐1‐9003  2.  VERIFY™ CLASS 5 CHEMICAL INTEGRATOR (2,000/BOX)  P2131‐1‐9004  VERIFY™ EXTENDED CYCLE TUBE (50/BOX)  P2131‐1‐9005  3.  4.  VERIFY™ BOWIE DICK TEST STRIP (100/BAG)  P2131‐1‐9006  5.  3M 1296 ‐ BIOLOGICAL PROCESS CHALLENGE DEVICE (25/BOX)  P2131‐1‐9010  6.  3M 1243 ‐ CLASS 5 CHEMICAL INTEGRATOR (100/BAG)  P2131‐1‐9011  3M 1292 ‐ BIOLOGICAL INDICATOR (50/BOX)  P2131‐1‐9012  7.  8.  3M 1233L ‐ BOWIE DICK TEST PACK (30/BOX)  P2131‐1‐9013  9.  TEST STRIPS, WATER CHLORINE LEVEL (PACKS OF 50)  P2131‐3‐9007  10.  CARTRIDGE, SWS DEIONIZED WATER (PACK OF 6)  P2131‐3‐9008 ...
  • Page 79: Maintenance

    WARNING:  When not in use, the P2131 SMU and the WRS should be drained completely  to prevent potential growth of biological matter.  To drain water, lift and gently  rock the SMU or WRS back and forth.  Do not use compressed air to blow out  water.      WARNING:    The  P2131  sterilizer  has  a  pressure  vessel  that  has  been  ASME  B&PV  code  stamped.  It is a violation of federal code to replace the main door plate or make  any repairs to the pressure vessel.  All repairs to the pressure vessel, including  the door plate, must be made BY FDI.      WARNING:  Replacement  parts  should  be  sourced  from  FDI.    Use  of  unauthorized  parts ...
  • Page 80: Preventive Maintenance Summary

     4.1.1   Preventive Maintenance Summary    Frequency  Task  Section  2 Weeks, or 100 Cycles  1. FDA‐Required Water Changeout  3.4  (whichever comes first)  2 Months  1. Pressure Safety Valve Test  4.1.2    2. Jacket Blowdown   6 Months, or 1300 Cycles  1. Replace the HEPA Filter  4.1.3     2. Lubricate Chamber Doorpost Threads      3. Inspect Electrical Box Inlet Air Filter     4. Inspect Electrical Box Cooling Fan Air Filter    5. Inspect Heat Exchanger in WRS    6. Inspect Heat Exchanger Fan in WRS    7. Inspect WRS and SMU Electrical Wiring   ...
  • Page 81: 2-Month Preventive Maintenance

    4.1.2  2‐Month Preventive Maintenance    Pressure Safety Valve Test    (NOTE: this procedure also conducted on 1  startup after storage) CHECK  TASK  WARNING: Do not pull the pin on the PSV while the unit is under  pressure. Hot Steam will exit the valve and can cause injury.    1. Pull the pressure safety valve (PSV) pin on the rear of the SMU to ensure    it is not stuck. This should only be done when the jacket pressure is at  atmospheric pressure (e.g., 0 psig).  2. Then turn sterilizer on and allow to fully heat up. Inspect the PSV for no  leaks.    Jacket Blowdown Procedure   CHECK  TASK  1. With the sterilizer completely setup and the jacket at normal operating  pressure, connect the high‐temperature drain hose to the jacket drain  valve.     CAUTION: The jacket drain valve is hot. To avoid a potential burn hazard,    use heat resistant gloves to operate the valve and attach/remove the  drain hose.    CAUTION: To avoid a potential burn hazard, ensure the high‐temperature  drain hose is used to perform the jacket blowdown.  2. Relocate the shims to the front of the sterilizer to tilt the unit towards    the jacket drain valve. ...
  • Page 82: 6-Month Preventive Maintenance

    4.1.3  6‐Month Preventive Maintenance    EVERY 6 MONTHS OR 1300 CYCLES   (WHICHEVER OCCURS FIRST)  WARNING:  To avoid a potential electrical shock or burn hazard, be sure the sterilizer has  been de‐energized and depressurized prior to performing these tasks.     CHECK  TASK  1. Replace the HEPA filter located under the SMU top panel.  The HEPA    filter is a safeguard against contamination during vacuum and should  always be in good working condition.  2. Lightly lubricate the chamber doorpost threads using high temperature    (275°F) NLGI #2 grease.  3. Inspect inlet air filter on right‐hand side of SMU.  If dirty, wash and    reinstall.  4. Inspect air filter on outlet of SMU electrical box cooling fan.  If dirty, wash    and reinstall.  Remove the SMU top and right‐hand side panel to access  this filter.    5. Inspect the heat exchanger in the WRS and clean if dirty.  6. Apply power (120V/60Hz) to heat exchanger fan in WRS and ensure that  it is functional. NOTE: Operation of this fan is normally hidden.    NOTE: On S/N’s after AFS‐0236, there is a single fan mounted on the  heat exchanger mounted on the Back Panel. Operation of this fan can be  observed by checking for air flow during operation of the Sterilizer.  7. Inspect wiring in SMU and WRS electrical boxes for loose wires,   ...
  • Page 83: 12-Month Preventive Maintenance

    4.1.4  12‐Month Preventive Maintenance    Disinfecting Clean of the WRS  (FDA‐Required)    CHECK  TASK  1. Ensure that the water in the SMU and WRS are at the HIGH mark    (add some water if needed using the SWS).  2. Remove the top panel from the WRS.    3. Pour  in  1/2  gallon  of  household  chlorine  bleach  (3‐6%  sodium  hypochlorite) to the bleach port of the WRS water tank. Reference  Section 4.5.     NOTE: On S/N’s after AFS‐0236, the breather also functions as the    bleach port and can be removed and replaced for this procedure.    CAUTION: Do not pump water into the SMU until bleach has been  rinsed from the WRS.  4. Run a Textile 4/5 cycle. The 4/5 cycle has a total cycle time of over  20  minutes,  providing  continuous  turbulent  water  circulation  in   ...
  • Page 84 Heater Flange Tee Inspection and Cleanout  WARNING:  To avoid a potential electrical shock or burn hazard, be sure the sterilizer  has been de‐energized, depressurized, and drained prior to performing these tasks.       *NOTE: This procedure is no longer necessary for S/N’s after AFS‐0236.      CHECK  TASK  1. Remove Back Lower Panel and Heater Assembly Cover (Reference    Section 4.3.17)  2. Remove the two Heater Flange Drain Caps. (Reference Section 3.2    Controls and Indicators – Rear View of SMU for location).  3. Inspect the heater flange tees to ensure there are no blockages.    CAUTION: Use eye protection when examining heater flange tees.  4. If  any  blockages  exist,  use  a  small  tool  to  break  debris  free  and  clean out the heater flange tees. Take care to prevent damage to    the fittings.  5. Replace heater flange drain caps. Tighten finger‐tight and 1/4 turn.   ...
  • Page 85 WRS K4 Relay and Liquid Level Switch Check    CAUTION: High Voltage contacts exposed with the WRS Top Panel removed. Electrical shock    hazard when using a multi‐meter to check the K4 Relay and Liquid Level Switch.     CHECK  TASK  1. Remove the Top Panel from the WRS.     2. Connect power to the WRS from the SMU, but do not slide the    WRS underneath the SMU or connect the white hoses.   3. Turn the SMU and WRS on. With the WRS empty, use a multi‐ meter to check terminal #10 on S/N’s AFS‐0001 through AFS‐0236    or terminal #14 on S/N’s after AFS‐0236 for 115 VAC.  4. With the WRS empty, use a multi‐meter to check for no    continuity across terminal #1 and #5 on K4 Relay.   5. Fill the WRS. Use a multi‐meter to check terminal #10 on S/N’s  AFS‐0001 – 0236 and terminal #14 on S/N’s after AFS‐0236 for        0 to 40 VAC.  6. With the WRS filled, use a multimeter to check for continuity    across terminal #1 and #5 on K4 Relay.    Inspect SMU Fill Check Valve & WRS Fill Valve    CAUTION: Burn Hazard.  Hot water and/or steam can remain in the hose that will be removed    below when it is disconnected.    CHECK ...
  • Page 86 Retorque WRS Pump Head Screws    CHECK  TASK  1. Ensure the WRS is disconnected from the SMU and is accessible    from all sides.   2. Remove the Front Panel (panel closest to bulkhead end) to expose    the motor.  3. Using a 10 mm socket and torque wrench set to 108 in‐lbs., torque  the  4  pump  head  bolts  (See  Section  4.5  Interior  Components    Diagrams for location).  4. Replace  the  Front  Panel,  reconnect  the  WRS,  and  verify  proper    operation of the WRS by pumping water into the SMU.     WRS Breather & PSV Inspection ...
  • Page 87: Post-Maintenance Checklist

    4.1.5  Post‐Maintenance Checklist     (After any intrusive service work)   CHECK  TASK  1. The unit is tilted for correct condensate drainage. Ensure the    back of the unit is 1” higher than the front (approx. 4 shims).    2.  All hoses are connected.    3.  All electrical cables are connected.    4.  No loose parts or tools in the electrical panels.    5.  All panels are reinstalled.  6.  The water levels in the SMU and the WRS are both at the HIGH    marks.  7.  The jacket vent valve, chamber vent valve, and both drain valves    are closed firmly.    8.  The WRS is turned ON.  9.  All accessories are accounted for in the SAB and the SAB is    properly stored.    10. All shelves are in place.  11. All alarms, repairs, and cycle count (found in “STATUS” screen) ...
  • Page 88: Special Tools And Materials Required For Maintenance

    Special Tools and Materials Required for Maintenance      The P2131 is specifically designed using components with standard sizing, allowing the system  to be maintained and repaired using tools that are common to most maintenance technicians.   These items are assumed to be readily available.    Only a few items would be considered necessary for maintaining this equipment that may be  less readily available.  These unique items are listed below.    1. Wrench, Torque (3/8” Square Drive, capable up to 150 inch‐lbs.),     (P/N: P2131‐1‐9015) 2. Grease, High Temperature (275°F), NLGI #2,     (P/N: P2131‐1‐9016) 3. Bleach, Chlorine, 3‐6% Solution,   (P/N: P2131‐2‐9004)   4. Mallet, Soft,     (P/N: P2131‐1‐9020) 5. Gloves, Heat Resistant,      (P/N: P2131‐1‐9017) 6. Brush. Long, Thin Bottle‐type,      (P/N: P2131‐1‐9018) 7. Wrench, 11/16” Flare Nut,      (P/N: P2131‐1‐9021) 8. Pigtail, WRS Power Cord,   (P/N: P2131‐2‐4016) NOTE: This is optional and allows the WRS to be setup to the side of    the SMU for troubleshooting.    ...
  • Page 89: Maintenance Procedures

    Maintenance Procedures    WARNING:  Maintenance work must be performed by qualified maintenance technicians.  Maintenance Videos    Please visit the FDI video channel for helpful and informative videos on the maintenance of the  P2131 sterilizer.   Scan the QR code below, click on the hyperlink (if you are using the electronic  version of the Technical Manual), or enter the Web address listed to access these videos.     NOTE: This channel is updated frequently, so check back periodically to view new content.      https://tinyurl.com/2kraxrf9      Procedures Summary List:    Section  Procedure  4.3.1  Replacing the HTS‐1 RTD Sensors  4.3.2  Disinfecting Clean for WRS  4.3.3  Cleaning the Sight Glass  4.3.4     Replacing the SWS Cartridge  4.3.5  Replacing the HEPA Filter  4.3.6  Replacing the Steam Trap  4.3.7  Replacing the RTD Sensor  4.3.8  Replacing the Chamber Door Gasket  4.3.9  Replacing the Chamber Doorpost O‐Ring  4.3.10     Replacing Electric Immersion Heating Elements  4.3.11 ...
  • Page 90: Replacing The Hts-1 Rtd Sensors

      4.3.1  Replacing the HTS‐1 RTD Sensors   1. Turn OFF the SMU, disconnect power, and ensure pressure in the chamber and jacket is  at 0 psig.  2. Remove the Back Lower Panel, Heater Assembly Cover, and Heater Block Enclosure by  removing 10 pan head screws. (Reference Section 4.3.17)  3. Disconnect the 3 incoming heater wires by using 3/8” wrench to counter‐torque and 3/8”  socket to loosen nuts. Heater element nuts are torqued to 35 in‐lbs.      S/N’s AFS‐0001 through 0286  S/N’s after AFS‐0286    Wiring Configuration  Wiring Configuration        NOTE: Always reuse the blue isolation washers against heater plate when reinstalling  heater plate.    4. Disconnect the white RTD Connector in bottom of heater box.   5. Remove 8 white heater stud caps (After S/N’s AFS‐0340).   6. Remove the 8 heater plate nuts with a 9/16 socket. Remove the lock, flat, and isolation  washers. The heater plate nuts are torqued to 120 in‐lbs.   7. Use a rubber mallet to pry the heater plate away from the heater flange.   8. Slide the Heater Element Assembly out of the jacket.   9. Compress the springs and slide the springs off the end of both RTD’s.  Isolation Bushings  RTD’s     ...
  • Page 91: Disinfecting Clean For Wrs

    4.3.2  Disinfecting Clean for WRS  1. Ensure that the water in the WRS is at the HIGH mark (add some water if needed using  the SWS).  2. Remove the top panel from the WRS.  3. Pour in 1/2 gallon of household chlorine bleach (3‐6% sodium hypochlorite) to the bleach  port of the WRS water tank. Reference Section 4.5. CAUTION: Do not pump water into  the SMU until bleach has been rinsed from the WRS. NOTE: On S/N’s after AFS‐0236, the  breather  also  functions  as  the  bleach  port  and  can  be  removed  and  replaced  for  this  procedure.  4. Run a Textile 4/5 cycle. The 4/5 cycle has a total cycle time of over 20 minutes providing  continuous  turbulent  water  circulation  in  the  WRS  to  ensure  complete  contact  and  effective disinfection.  ...
  • Page 92: Replacing The Sws Cartridge

    4.3.4   Replacing the SWS Cartridge  1. Remove inlet and outlet hoses from the blue housing.  2. For S/N’s AFS‐0001 through AFS‐0255, remove the bungee cords and slide the blue  housing out from the SWS frame. For S/N’s after AFS‐0255, remove the bracket that  secures the blue housing by loosening the two panel fasteners and slide the blue  housing out from the SWS frame.  3. Use the supplied spanner wrench stored inside the SWS frame to remove the blue  housing from the black top.   4. Slide the old cartridge out and discard.  5. Slide the new cartridge in place, ensuring the gasketed side is up towards the opening of  the housing.  6. Verify large Blue Housing O‐ring is in place and is lubricated with silicone grease.  7. Follow steps 1‐3 in reverse order to return the SWS to operation.  Gasket Side Up  Blue Housing  O‐ring      4.3.5  Replacing the HEPA Filter  1. Turn OFF the SMU, disconnect power, and ensure pressure in the chamber and jacket is  at 0 psig.  2. Remove the left side panel of the SMU.  3. Using an 11/16” wrench to counter‐torque the fitting near solenoid valve, use an 11/16”  wrench to loosen the fitting nut at the filter.  4. Install the new HEPA filter by following steps 1‐3 in reverse order.  Loosen Nut  Counter Torque fitting     ...
  • Page 93: Replacing The Rtd Sensor

    6. Use a 13/16” wrench to counter‐torque the fitting while using a 9/16” wrench to loosen  the 1/4" fitting nut (located at the top/right).  Loosen Nut                                               Counter‐torque stationary fitting  7. Carefully slide the RTD sensor out.   8. Install the new RTD Sensor by following steps 1‐7 in reverse order.    4.3.8  Replacing the Chamber Door Gasket  NOTE:  The chamber door gasket should be replaced whenever a leak is detected around  the door.  The recommended door O‐ring is made of ethylene propylene diene  monomer  (EPDM)  and  has  no  seam.    Orange‐colored  silicone  door  gaskets  should  not  be  used  due  to  inferior  resistance  to  steam.    Contact  FDI  for ...
  • Page 94: Replacing The Chamber Doorpost O-Ring

      Carefully  work  the  gasket  into  the  groove  until  it  is  evenly  distributed around the door.  Working from one point only will result in too much or too  little gasket left over at the other side of the door.  Wipe the gasket and the mating face  on the chamber door frame to ensure that no debris could cause a leak path.  4.3.9  Replacing the Chamber Doorpost O‐Ring    NOTE:  The O‐ring under the doorpost should be replaced whenever a leak is detected at  the doorpost or when the doorpost is refitted.  The O‐ring material is EPDM. Do not use  Buna‐N or silicone doorpost O‐rings due to inferior resistance to steam.  Contact FDI for  replacement parts (Reference Section 4.4 ‐ Recommended Replacement Parts).  1. Turn OFF the SMU, disconnect power, and ensure pressure in the chamber and jacket is  at 0 psig.  2. Open  the  chamber  door  and  remove  the  handwheel  retainer  (large  washer)  on  the  outboard end of the doorpost.  3. Remove the six radial arm wire retainers.  4. Rotate the handwheel assembly until the entire assembly unscrews from the doorpost. ...
  • Page 95: Replacing Electric Immersion Heating Elements

    4.3.10 Replacing the Electric Immersion Heating Elements  NOTE: With wires disconnected, the resistance of the heaters should be between 16.7  and 19.4 Ohms.    1. Turn OFF the SMU, disconnect power, and ensure pressure in the chamber and jacket is at 0  psig.   2. Remove  Back  Lower  Panel,  Heater  Assembly  Cover,  and  Heater  Block  Enclosure.  Then  disconnect  the  3  incoming  heater  wires,  noting  the  exact  connections  of  each  wire.  Disconnect the white RTD Connector. (Reference Section 4.3.17 ‐ Removing Back Lower Panel    and Heater Assembly Cover)       3.
  • Page 96: Replacing The Plc

    8. Slide  new  compression  fitting  nuts  and  then  ferrules  over  the  ends  of  the  replacement  heaters.  9. Insert the ends of the heating elements into the male fittings in the heater plate.  Ensure that  the ends of the heaters do not protrude too far, which would potentially interfere with the  cover of the heater box.   Insert a wrench or other flat tool through the loop end of the heating  element  to  prevent  it  from  twisting.    Tighten  the  compression  fitting  nut  firmly  using  an  11/16” flare nut wrench.   10. Reinstall RTD’s by inserting into Swagelok fitting and tightening nut finger tight then 1/4 turn.  The long RTD is installed in the right fitting, on right leg of center heater element. The short  RTD is installed in the center fitting, on left leg of center heater element.   11.
  • Page 97 21. As  the  PLC  boots  up,  take  note  of  the  software  version  as  it  flashes  on  the  screen  for  approximately 2 seconds.  If the software version in the newly installed PLC is older than  what  was  noted  in  Step  1,  the  software  should  be  updated.    Contact  FDI  for  update  information and procedures.   22. Reset  settings  using  the  “CONFIRM  SETUP”  screen,  including  resetting  the  barometric ...
  • Page 98: Plc Software Update

    4.3.12   PLC Software Update   The following is the procedure for updating the software in the P2131 Automated Field Steam  Sterilizer using the Micro SD card that is shipped with each sterilizer.      NOTE:  This procedure should only be performed by a qualified Maintenance Technician     1. The updated software file will be delivered via email.  2. Using the Micro SD card and adapter supplied with each sterilizer, save the file onto the  Micro SD card with your computer.  3. Ensure the sterilizer is OFF.  4. Insert the Micro SD card (without the adapter) into the extender cable and insert the other  end of the cable into the top of the Programmable Logic Controller (PLC).  The exposed pins  should face toward the front of the sterilizer.                                           5. Note the firmware version number by turning the sterilizer ON and watching the first screen  that appears.  The firmware version number will display for only 2 seconds.  6. From the CONFIRM SETUP screen, press the F4 function key.  (F4 is only accessible from this  screen.)  7. The UPDATE SOFTWARE screen will display and the PLC will initialize the Micro SD card.   This may take a few minutes after which a list of files will appear.  8. Use the up and down arrow keys on the right‐hand side of the screen to select the new  software file. (Example: V1285 OR V1043)  9. Choose OK when you are asked if you are sure you want to load the application.  10. A moving clock and progress bars will appear while the software is loaded.  This may take a  few minutes.  11. Next, you will see LOAD PAGE COMPLETE and you will be asked if you would like to place ...
  • Page 99 The P2131 FIRMWARE UPDATE KIT includes the following items:  ‐ CD with Cscape programming software  ‐ CD with the new PLC firmware  ‐ Micro SD card and adapter with the sterilizer software program  ‐ Micro SD extender cable  ‐ Communication cable  ‐ Instruction sheet  WARNING:  Firmware updates typically delete the user applications to ensure compatibility.   After updating the firmware, you will need to reload the sterilizer software program.  Make  sure the sterilizer software program is available before beginning the firmware update.    Step 1 – Loading the New Firmware    1. Insert the CD with Cscape software and install the Cscape programming software to your  computer. NOTE:  Computer must be a PC with Windows 7 or newer operating system.  2. Insert the CD with the new PLC firmware, and run the XLe‐XLt‐firmware application from  the CD.  This will save several files into the Cscape/Firmware folder on your computer.   Verify that these files are in the correct folder on your computer.  3. Ensure the sterilizer is OFF, and connect the communication cable to a serial COM port on  your computer and to MJ1 on the Programmable Logic Controller (PLC).  4. Turn the sterilizer ON.  5. Open Cscape on your computer and establish communication between Cscape and the PLC.  a. At the ABOUT Cscape window, choose OK.  b. On the CONNECTION WIZARD window, select SERIAL and choose NEXT.  c. On the following window, enter the COM port number that you are using for  communication and choose FINISH. ...
  • Page 100 14. To double‐check the download, in Cscape choose CONTROLLER then STATUS.  This will show  the latest firmware revision on the PLC.  This should match what was downloaded.  The Self‐ Test may fail because there is not a program in the PLC.  Choose CLOSE to exit the window.  15. After the firmware update, the PLC is in IDLE mode.  To switch it to RUN mode in Cscape  choose CONTROLLER and then RUN/MONITOR.  16. Turn the sterilizer OFF.    Step 2 – Reloading the Sterilizer Software Program    1. Insert the Micro SD card (without the adapter) into the extender cable and insert the other  end of the cable into the top of the Programmable Logic Controller (PLC).  The exposed pins  should face toward the front of the sterilizer.                                           2. Turn the sterilizer ON and watch the first screen that appears and note the firmware version  number that appears.  NOTE: The firmware version number will display for only 2 seconds.  3. Press the SYSTEM button.  4. From the SYSTEM MENU use the up and down arrow keys on the right‐hand side of the  screen to select REMOVABLE MEDIA.  Press the RETURN button.  5. The Media Directory screen will display and the PLC will initialize the Micro SD card.  This  may take a few minutes after which a list of files will appear.  6. Use the up and down arrow keys on the right‐hand side of the screen to select the software  file that matches the firmware version number.  (For example, if the firmware version is  12.85, then choose software file 1285.pgm).  Press RETURN.  7. Choose OK when you are asked if you are sure you want to load the application.  8. A moving clock and progress bars will appear while the software is loaded.  This may take a  few minutes. ...
  • Page 101: Vacuum Leak Test Troubleshooting

    4.3.14   Vacuum Leak Test Troubleshooting Guide  Air leaks into the sterilizer can be problematic since the purpose of a dynamic air removal  sterilizer (i.e., pre‐vacuum sterilizer) is to remove the air to enable more effective  sterilization.  Vacuum leaks are often difficult to diagnose since a small vacuum leak might not  be visible with steam under pressure.  If the P2131 sterilizer fails a Vacuum Leak Test, the  following Vacuum Leak Troubleshooting Guide should be used by a qualified maintenance  technician to help find the root cause of the vacuum leak.      The source of the vacuum is the water eductor that is located inside the WRS.  Water flowing  through an internal tapered nozzle uses the Venturi effect to create a drop in pressure at the tip  of the nozzle.  This pressure drop creates the source of vacuum for the SMU chamber.   The  green highlighted areas of the Process Flow Diagram in Section 4.6 show the parts of the  system that experience vacuum during the conditioning phase (pre‐vacuum pulses) and the  drying phase.  Using this diagram in conjunction with the Vacuum Leak Troubleshooting Guide  should help the maintenance technician diagnose and correct the source of the vacuum leak.    Please see the full troubleshooting guide in Appendix A.5 – Vacuum Leak Test Troubleshooting  Guide.     ...
  • Page 102: Door Switch Reprogramming

    4.3.15   Door Switch Reprogramming  The door switch may cease to operate correctly if, (1) maintenance on the door is performed,   (2) the door becomes misaligned, or (3) the door switch is replaced. To check for correct door  switch operation, navigate to the STATUS screen and note that the indication is either DOOR  OPENED or DOOR CLOSED depending on the position of the Door Locking Arm Lever.    WARNING: This must be performed with the sterilizer turned ON. Be cautious of high voltage  within the electrical enclosure.     1. To reprogram the door switch, remove the right side panel to expose the electrical  enclosure. Locate the yellow door switch in the upper left portion of the enclosure.  2. Press the black TEACH button on the side of the door switch sensor until the green LED  turns to yellow or blinks red.  3. Present the ON condition, which is the door closed and the Locking Arm Lever rotated  clockwise until the radial arms are fully extended. Press the TEACH button and the  yellow LED will turn off.  4. Present the OFF condition, which is the door in the fully open position. Press the TEACH  button and the LED will turn green, confirming the new settings have been accepted.  5. Test for correct operation by observing the STATUS screen while operating the door.    OFF Condition: Door  ON Condition: Door  Opened  Closed and Locked         ...
  • Page 103: Replacing The Plc Battery

    Be  sure  to  consult  with  the  appropriate  regulatory  agencies before disposing batteries.   NOTE:  Do  not  make  substitutions  for  the  battery.    The  Horner  XLt  PLC  uses  a  coin  lithium battery    (P/N: P2131‐1‐9002).   1. Replacing the Battery  a. Record the CYCLE COUNT from the STATUS screen for use later.  b. Disconnect all power from sterilizer.   c. Remove the PLC from the sterilizer (see Section 4.3.11). ...
  • Page 104 2. Reinstalling the PLC  a. Install the Micro SD card into the MEMORY slot on top of the PLC with the  contacts of the card toward the front of the PLC.  It should spring into place.   (After this Battery Replacement Procedure is completely finished, the card can  be left in the PLC indefinitely or returned to the plastic case in the Sterilizer  Accessory Bag.)  b. Reinstall the PLC into the sterilizer (see Section 4.3.11).  c. Verify that the MODBUS cable is installed in the MJ2 port on the PLC.  d. Apply power to the unit.  e. As the unit powers up, you should see several error screens and finally a screen  that says, “No User Screens to Display”.  3. Updating the “Screen Update Time”  a. Press the SYSTEM button.  b. Arrow down to SET SCREEN and choose ENTER ( ).  c. Arrow down to UPDATE TIME (mS) and choose ENTER ( ).  d. Change the time to 47 and choose ENTER ( ).  e. Choose ESC.  4. Loading the Program  a. Arrow down to REMOVABLE MEDIA and choose ENTER ( ).  b. Arrow down to the program (examples – 1285.PGM, V1043.PGM) and choose  ENTER ( ).  c. Choose OK and wait for the program to load.  d. When you see “PLACE IN RUN MODE?” choose OK.  5.
  • Page 105: Removing Back Lower Panel And Heater Assembly Cover

    4.3.17   Removing Back Lower Panel and Heater Assembly Cover  CAUTION: This procedure must be performed with the sterilizer OFF and cooled to room  temperature. Be cautious of pinch points.  1. Removing/Installing the Back Lower Panel  a. Use a Multi‐tool to unfasten the latches (3x) and carefully remove the Back    Lower Panel.        b. Follow Step 1a in reverse order to install the Back Lower Panel    2. Removing the Heater Assembly Cover  a. Carefully position your hands to grab underneath the black Heater Assembly  Cover.  Bottom View  Green – Recommended areas  to grab and pull  Red – Not recommended areas  to grab and pull    b. First pull down and then away from the Heater Block Enclosure.  First, use your  fingers to pull down  Second, pull away from  the Heater Block    c. Carefully remove the Heater Assembly Cover.   ...
  • Page 106 3. Installing the Heater Assembly Cover  Note: There are two tabs located on the Top and Bottom within the Cover.  Top Tabs  Bottom Tabs    a. Position Top Tabs over the top flange of the Heater Block Enclosure as shown.    b. Slowly lower the Heater Assembly Cover over the Heater Block Enclosure.  Note: The two bottom tabs are designed to secure the black Heater Cover to the  bottom flange of the Heater Block Enclosure.   c. Tap the center of black Heater Assembly Cover with a Dead Blow Hammer or the  bottom of the fist until it clips into place.    d. Check  the  fit  of  the  black  Heater Assembly  Cover.  It  is  designed  for  a  close  fit and  should not be able to be removed easily. ...
  • Page 107: Replacement Parts List

      NOTE:  Nomenclature within parentheses is for reference on the Process Flow Diagram   and Electrical Wiring Diagrams, sections 4.6 and 4.7 respectively.    Parts with an * are FDI recommended spare parts to stock.    SMU Replacement Parts    Item  Description  Part Number  1.   ALARM, BUZZER (ALM‐1)  P2131‐1‐4008  2.   BATTERY, HORNER PLC (PKG OF 4)  P2131‐1‐9002  3.   BLOCK, FUSE, INCOMING POWER (FB1)  P2131‐1‐4023  4.   BLOCK, FUSE, SMALLER FUSES, 6 POSITION (FB2)  P2131‐1‐4025  5.   BLOCK, TERMINAL, 12 POSITION (TB1)  P2131‐1‐4022  6.   BLOCK, TERMINAL, 30A (TB2)  P2131‐1‐4026  7.   CONNECTION, JACKET DRAIN HOSE  P2131‐1‐9009 ...
  • Page 108 36.   KIT, SOLENOID VALVE REPAIR (SV‐1, SV‐2, SV‐3, SV‐4, SV‐5)    P2131‐1‐9001   37.   KNOB, ENDCAP   P2131‐1‐0020  38.   LEVER, LOCKING ARM  P2131‐1‐1023  39.   LIGHT, INDICATOR, LED (LAMP‐1)  P2131‐1‐4018  40.   MANIFOLD, CHAMBER DRAIN  P2131‐1‐0022  41.   MANUAL, P2131 TECHNICAL, WITH CD  P2131‐1‐8001  42.   MODULE, RTD (RTD‐1, RTD‐2) TRANSMITTER  P2131‐1‐4006  43.   MULTI‐TOOL  P2131‐1‐2020  44.   O‐RING, CHAMBER DOOR GASKET (PKG OF 2)    P2131‐1‐1029   45.   O‐RING, CHAMBER DOOR POST (PKG OF 5)    P2131‐1‐1121  ...
  • Page 109 Item  Description  Part Number  67.   SWITCH, MAIN POWER (SW‐1)  P2131‐1‐4017  68.   TRANSMITTER, CHAMBER PRESSURE (PT‐2)  P2131‐1‐4014  69.   TRANSMITTER, JACKET PRESSURE (PT‐1)  P2131‐1‐4014  70.   VALVE, CHAMBER VENT   P2131‐1‐0034  71.   VALVE, CHECK, RECOVERED WATER  P2131‐1‐3402  72.   VALVE, JACKET DRAIN  P2131‐1‐0039  73.   VALVE, JACKET PRESSURE SAFETY  P2131‐1‐3403  74.   VALVE, JACKET VENT   P2131‐1‐0035  75.   VALVE, SOLENOID, STERILIZE (SV‐1)  P2131‐1‐0024  76.   VALVE, SOLENOID, VACUUM BREAK (SV‐3) ...
  • Page 110 Item  Description  Part Number  101. SWITCH, LIQUID LEVEL (LS‐1)  P2131‐2‐4003  102. SWITCH, ROCKER, WATER RETURN (SW‐2)  P2131‐2‐4009  103. SWITCH, WRS POWER (SW‐3)  P2131‐2‐4008  104. TANK, STAINLESS, WATER (S/N’S AFTER AFS‐0236)  P2131‐2‐9007  105. TANK, WATER  P2131‐2‐0008  106. TRAP, STEAM    P2131‐2‐3401   107. VALVE, CHECK, EDUCTOR  P2131‐1‐3402  108. VALVE, CHECK, FILL (S/N’S AFS‐0001 THROUGH 0236)  P2131‐2‐3407  109. VALVE, CHECK, PRESSURE (S/N’S AFS‐0001 THROUGH 0236)  P2131‐2‐3406  110. VALVE, CHECK, STEAM TRAP  P2131‐1‐3402  111. VALVE, CHECK, TANK DRAIN (S/N’S AFS‐0001 THROUGH 0236)  P2131‐2‐3405  112. VALVE, CHECK, VACUUM (S/N’S AFS‐0001 THROUGH 0236)  P2131‐2‐3406 ...
  • Page 111: Interior Components Diagrams

    Interior Components Diagrams      NOTE:  Nomenclature within parentheses is for reference on the Process Flow Diagram   and Electrical Wiring Diagrams, sections 4.6 and 4.7 respectively.    Sterilizer Main Unit (SMU)      Jacket Pressure Transmitter (PT‐1)  Chamber Pressure Transmitter (PT‐2)    (P/N: P2131‐1‐4014)   (P/N: P2131‐1‐4014) NOTE:  Pressure  transmitters are  factory‐calibrated  and cannot be  adjusted.  Small fire sleeve insulation                                                  Top of SMU, from Left Side      Electrical Box Cooling Fan (F1)  (P/N: P2131‐1‐4011)  NOTE:   Opposing  orientation  Sterilize Valve (SV‐1)  of vacuum  (P/N: P2131‐1‐0024)  break  valves  ensures  Large fire sleeve insulation  reliable  Flow Direction  operation  Arrows ...
  • Page 112   NOTE:  The Stainless‐Steel braided hose used for S/N’s AFS‐0001 through AFS‐0354 has been                   changed to a clear PFA hose, starting with S/N AFS‐0355. There are no changes to                  maintenance procedures or replacements of fittings. For future hose replacements, new                 PFA hose will be sent pre‐swaged and ready for installation. Vacuum exhaust hoses from                 the chamber drain to the rear bulkhead remain the same.      Hose, Jacket Vent‐In      Hose, Vent Break Valve     HEPA Filter  (P/N: P2131‐1‐0028)  (P/N: P2131‐1‐3116)        (P/N: P2131‐1‐3111)  Hose, Level Switch  Vacuum Break Valve (SV‐3)  (P/N: P2131‐1‐3106)  (P/N: P2131‐1‐0026)  Hose, Jacket Vent‐Out  (P/N: P2131‐1‐3115)  Vacuum Break Valve (SV‐4)  (P/N: P2131‐1‐0027)  Hose, Chamber Vent‐Out  (P/N: P2131‐1‐3118)  Hose, Chamber Vent‐ In  (P/N: P2131‐1‐3117)  Hose, Vacuum Valve‐Out  (P/N: P2131‐1‐3114)  Hose, Vacuum Valve‐In  (P/N: P2131‐1‐3113)  Hose, Jacket Fill (Not Shown)  Vacuum/Exhaust Valve (SV‐2)  (P/N: P2131‐1‐3107)  (P/N: P2131‐1‐0025)    Left Side of SMU ...
  • Page 113 Dual RTD (RTD‐1 & RTD‐2)  NOTE:  A single  Large fire sleeve insulation  (Inside insulation and tubing)  sensor probe  (P/N: P2131‐1‐0023)  contains two  RTD elements  that are factory‐ calibrated and  cannot be  adjusted.  Chamber Drain Manifold   (P/N: P2131‐1‐0022)  Connections for Dual RTD,  Condensate Hose (Red‐tagged) and  Vacuum/Exhaust Hose    Front of SMU, behind Lower Front Panel  Liquid Level Switch  RTD Connector  RTD Connector  (FS‐1)  (P/N: P2131‐1‐4021)  (P/N: P2131‐1‐4021)  (P/N: P2131‐1‐4016)                Right Side of SMU for S/N’s                   Right Side of SMU for S/N’s               AFS‐0001 through AFS‐0236                              After AFS‐0236   ...
  • Page 115 Water Recovery System (WRS)    For S/N’s AFS‐0001 through AFS‐0236:    Fill Check Valve  (Behind Panel)  (P/N: P2131‐2‐3407)  Right Side of WRS        Eductor Check Valve  Eductor  (P/N: P2131‐1‐3402)  (Behind Bracket)  (P/N: P2131‐2‐3402)  Drain Check Valve  (P/N: P2131‐2‐3405)  Water Fill Valve (SV‐5)  (P/N: P2131‐2‐0004)  Water Circulation Valve  (SV‐6)  (P/N: P2131‐2‐0005)     Left Side of WRS ...
  • Page 116   Water Tank  (P/N: P2131‐2‐0008)  NOTE: Not pictured:  S/N’s AFS‐0001 – 0027:  Pressure Safety Valve    (P/N: P2131‐2‐3404) Liquid Level Switch  (LS‐1)    (P/N: P2131‐2‐4003) Water Fill Valve   (SV‐5)    (P/N: P2131‐2‐0004) Pump  Motor (M)  (P/N: P2131‐2‐4001)  (P/N: P2131‐2‐3403)  Front of WRS  Steam Trap  (P/N: P2131‐2‐3401)  Cooler, Heat Exchanger  (S/N’S AFS‐0001‐0236)  (P/N: P2131‐2‐3000)  Steam Trap Check Valve  (P/N: P2131‐1‐3402)  Heat Exchanger Fan (F2)  (P/N: P2131‐2‐4004)  Back of WRS ...
  • Page 117     Relay (K3)  Vacuum Relief  Relay (K1)  (P/N: P2131‐2‐4006)    Check Valve  (P/N: P2131‐2‐4006)  (P/N: P2131‐2‐3406)    Bleach Port Plug  Relay (K2)  Relay (K4)  (P/N: P2131‐2‐9003)  (P/N: P2131‐2‐4006)  Pressure Relief  (P/N: P2131‐2‐4007)  Check Valve  (P/N: P2131‐2‐3406)          Flow    Direction    Arrows                  Top of WRS     ...
  • Page 118 Water Recovery System (WRS)    For S/N’s after AFS‐0236:    Liquid Level Switch  (LS‐1)    (P/N: P2131‐2‐4003)   Right Side of WRS        Eductor  (Behind Bracket)  Eductor Check Valve  (P/N: P2131‐2‐3402)  (P/N: P2131‐1‐3402)  Cooler, Heat Exchanger  (S/N’S After AFS‐0236)  (P/N: P2131‐2‐9009)    Left Side of WRS ...
  • Page 119 Tank Pressure Relief Valve  (P/N: P2131‐2‐9010)  Water Fill Valve (SV‐5)  Water Tank  (P/N: P2131‐2‐0004)  (P/N: P2131‐2‐0008)  Steam Trap  (P/N: P2131‐2‐3401)  Steam Trap Check Valve  (P/N: P2131‐1‐3402)  Pump, Water  (including Motor)  (P/N: P2131‐2‐9008)  Pump Head Bolts    Front of WRS    Heat Exchanger Fan (F2)  (P/N: P2131‐2‐4004)  Water Circulation Valve  (SV‐6)  (P/N: P2131‐2‐0005)    Back of WRS ...
  • Page 120 Relay (K3)  Relay (K1)  (P/N: P2131‐2‐4006)  (P/N: P2131‐2‐4006)  Relay (K2)  Relay (K4)  (P/N: P2131‐2‐4006)  (P/N: P2131‐2‐4007)  Bleach Fill Port & Breather Plug  (P/N: P2131‐2‐3019)      Top of WRS       ...
  • Page 121                           (THIS PAGE INTENTIONALLY LEFT BLANK)     ...
  • Page 122: Theory Of Operation

    (PLC).  As  heat  is  supplied  to  the  water  in  the  Jacket,  steam  is  generated  and  the  pressure  and  temperature in the Jacket rises. The Jacket Pressure Transmitter (PT‐1) supplies pressure data to allow the PLC to  control the Jacket at a steady pressure set point when the Jacket is in pressure control (during idle and certain phases  of a cycle).     The Door Switch (DS‐1, normally open) will not allow a cycle to start (and steam to flow into the chamber) unless the  door is closed and latched. When a cycle is started and steam is required in the Chamber, the Sterilize Valve (SV‐1)  is opened by the PLC to allow steam from the Jacket to flow into the Chamber. During the conditioning phase of a  cycle, the Chamber Pressure Transmitter (PT‐2) supplies pressure data to the PLC to allow it to pressurize or pull a  vacuum in the Chamber to specific set points. When steam condenses during a cycle, it flows through the chamber  drain, past the Chamber Temperature Sensors (RTD‐1 & RTD‐2), through the red‐tagged hose, and finally through  the WRS Steam Trap back into the Water Tank. The Chamber Temperature Sensors provide temperature data to  allow the PLC to control the heater elements and maintain the Chamber at a steady temperature set point when the  P2131 is in temperature control (during the heat‐up and exposure phases of a cycle). The Steam Trap allows cool  condensate to pass through to the WRS Water Tank but prevents the flow of steam.     During the conditioning phase and the vent phase of a cycle, steam and/or air is evacuated from the Chamber. The  PLC evacuates the Chamber by opening the Vacuum/Exhaust Valve (SV‐2), allowing a venturi vacuum Eductor to pull  the steam and/or air from the Chamber into the WRS via the white‐tagged Vacuum/Exhaust hose. The steam/air is  entrained in the water that flows through the Eductor back into the WRS Water Tank.    There are several cycles on the P2131 that require a drying phase after exposure. Since a vacuum facilitates faster  drying, the PLC pulls a vacuum on the Chamber by opening the Vacuum/Exhaust Valve for the duration of the dry  phase. After the dry phase, the PLC will equalize the pressure by opening the Vacuum Break Valves (SV‐3/4) to allow  outside air to be drawn through a HEPA filter into the Chamber. Once the Chamber Pressure Transmitter detects  equalized  pressure  between  the  Chamber  and  outside  environment,  the  PLC  closes  the  Vacuum  Break  Valves, ...
  • Page 123: Process Flow Diagram

    4.6.2  P2131 Process Flow Diagram (S/N's AFS-0001 through AFS-0236)
  • Page 125: Electrical Wiring Diagram

    (S/N's AFS-0001 through AFS-0236) 4.7  P2131 Electrical Wiring Diagram SMU-C-001 Sterilizer Main Unit (SMU) 015-10 30A-TD C1-A SSR-6 L1 - BLACK L1-1 012-B10 013-B10 HE-1 Power 025-10 30A-TD FB1-A C1-B SSR-7 Connection L2 - RED L2-1 022-R10 023-R10 HE-2 208 VAC, 3 Immersion 30A-TD...
  • Page 126 (S/N's AFS-0001 through AFS-0236) P2131 Electrical Wiring Diagram Sterilizer Main Unit DC Controls VDC (+) From VDC (-) From DC-4 411-B WD-1(13) WD-1(12) 401-R TB1-12 DC-1 TB2-8 750 mA Horner HTS-1 Slow Blow 402-O PLC Power 432-G DC-3 (Relay 2 is on pg.1)
  • Page 127 (S/N's after AFS-0236) 4.7 P2131 Electrical Wiring Diagram SMU-C-001 Sterilizer Main Unit (SMU) 015-10 30A-TD C1-A C1-A SSR-6 SSR-6 L1-1 L1 - BLACK HE-1 012-B10 013-B10 Power 025-10 30A-TD FB1-A C1-B C1-B SSR-7 SSR-7 Connection HE-2 L2 - RED L2-1...
  • Page 128 (S/N's after AFS-0236) P2131 Electrical Wiring Diagram VDC (+) From VDC (-) From Sterilizer Main Unit DC Controls Sterilizer Main Unit DC Controls DC-4 411-B WD-1(13) WD-1(12) 401-R TB1-12 DC-1 TB2-8 750 mA Horner HTS-1 HTS-1 Slow Blow 402-O PLC Power...
  • Page 129 Programmable Logic Controller (PLC) RTD Module #1 (CONTROL) (EXC) (N/C) (SENSE) GRAY (ARX) (- SIG) BROWN (ATX) (+ SIG) Chamber WHITE (GND) Temperature BLUE (BRX) RTD-1 RS485 BLACK (BTX) TB1-3 TB1-12 701-O 702-W/B TB2-7 703-G RTD Module #2 (MONITOR) (EXC) (N/C) (SENSE) GRAY...
  • Page 130: Appendix

                                  SECTION A – APPENDIX     ...
  • Page 131: Swagelok Fitting Installation Instructions

                              Swagelok Fitting Installation Instructions                   ...
  • Page 132 INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Installation Instructions Swagelok Tube Fittings Swagelok tube fittings 1 in./25 mm and smaller can be Up to 1 in./25 mm installed quickly, easily, and reliably with simple hand These instructions apply both to traditional fittings and tools.
  • Page 133 INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Swagelok Tube Fittings Swagelok Tube Fittings Over 1 in./25 mm Reassembly—All Sizes 1. Preswage the ferrules onto the tube using a You may disassemble and reassemble Swagelok tube Swagelok multihead hydraulic swaging unit fittings many times. (MHSU).
  • Page 134 VCO O-Ring Face Seal Fittings Features Test ports at two Swagelok VCO O-ring face seal fittings locations for easy O-ring Female nut threads are designed for rapid assembly in leak testing contained silver plated for pipe, tube, and welded systems. completely easy assembly Unique design allows easy...
  • Page 135: Plc Data Sheet

                              PLC Data Sheet          ...
  • Page 136 XLE & XLT OCS DATASHEET MODEL 3 12 DC In, 12 DC Out, 2 – 12-bit Analog In XLE MODEL XLT MODEL 1 TECHNICAL SPECIFICATIONS 1.1 General 1.3 Connectivity 1.4 Control & Logic Typical power-back- 164mA @ 10V (1.64W) RS-232 full handshaking or Advanced Ladder Control Lang.
  • Page 137 technical specifications continued... 1.6 Digital DC Inputs 1.7 Digital DC Outputs 1.8 Analog Outputs 12 Including 4 Configu- 12 Including 2 Config- Number of Channels Inputs per Module Outputs per Module rable HSC Inputs urable PWM Outputs 0 - 10 VDC Commons per Module Commons per Module 0 –...
  • Page 138 2 WIRING & CONNECTORS 3 COMMUNICATIONS 3.1 - CAN Communications 2.1 - Port Connectors CAN Pin Assignments SIGNAL DESCRIPTION DIRECTION CAN Ground V– – – Black CAN Data CN L IN/OUT Low – Blue Solid/Stranded wire; 12-24 Shield Ground awg (2.5-0.2mm). Strip SHLD –...
  • Page 139 communications continued... 5 INSTALLATION DIMENSIONS 3.4 - Dip Switches 1.9 DIP SWITCHES NAME FUNCTION DEFAULT 3.78” (96mm) MJ1 RS-485 ON = Termination Terminated MJ2 RS-485 ON = Termination Terminated Bootload Always Off 2.26” 3.78” (57.5mm) The DIP switches are used to provide a built-in termination to both the (96mm) MJ1 port and MJ2 port if needed.
  • Page 140 6 BATTERY 9 TECHNICAL SUPPORT The XLe/t uses a replaceable non-rechargeable 3V Lithium coin-cell For assistance and manual updates, contact Technical Support at the battery (CR2450) to run the Real-Time Clock and to keep the retained following locations: register values. This battery is designed to maintain the clock and North America Europe memory for 7-10 years.
  • Page 141: Verify™ Product Information

                              VERIFY™ Product Information       ...
  • Page 142 Introduction ® The VERIFY All-In-One STEAM Reusable Test Pack is a multipurpose reusable test pack designed to challenge vacuum assisted steam sterilizers operating at 270°F for a 4 minute exposure. Each test pack can be used 200 times as a daily air removal test, routine microbial challenge pack, biological indicator test pack for load release or a chemical indicator test pack for release of loads without implantable devices.
  • Page 143 The VERIFY Reusable test pack comprises three models: (1) Standard Test Pack, (2) Test Pack with Step Counter, (3) Test Pack with Tracker Tag. COMPONENTS VERIFY All-In-One STEAM Reusable Test Pack (Standard) VERIFY All-In-One STEAM Reusable Test Pack with Step Counter VERIFY All-In-One STEAM Reusable Test Pack with Tracker...
  • Page 144 Biological Indicators: The reusable test packs have been validated with biological indicators commonly used to monitor steam sterilizers. Table 1 lists the validated biological indicators. Table 1: Validated Biological Indicators. Biological Indicator Product Number Manufacturer 3M Attest Rapid Readout Biological Indicator 1292 3M Company 3m Attest Super Rapid Readout Biological Indicator...
  • Page 145 Reusable Extended Cycle Test Pack for Biological Indicators For 132° C (270° F) 10 Minute Pre-Vacuum Steam Sterilization Cycles The only device specifically designed and approved by the FDA for testing an extended 10 minute cycle! Patented Plastic Sleeve Technology , when used with the SteriScan Reusable Test Pack, provides an accurate...
  • Page 146 Challenges Extended Pre-Vac Cycles Test Pack Resistance 270˚ F Pre-Vacuum Cycle Percentage of Positive Biological Indicators (survival) 100% 100% Sec. Min. Min. Min. Min. Time Elapsed Reusable Extended Cycle Test Pack Other BI Test Packs...
  • Page 147: Wrs Pump Information

                              WRS Pump Information       ...
  • Page 148 (S/N's AFS-0001 through AFS-0236) Shertech Operating Instructions, Performance, Bronze and Cast Iron Models Specifications and Parts Manual Close-Coupled Turbine Pumps Bronze and Cast Iron Models Operation (Continued) assembly can cause an electrical motor shaft. A screwdriver slot or two shock hazard. flats for use with an open end 7/16"...
  • Page 149 (S/N's AFS-0001 through AFS-0236) Shertech Operating Instructions, Performance, Specifications and Parts Manual Bronze and Cast Iron Models Maintenance (Continued) 7. Assemble the impeller (Ref. No. 7) over the pump shaft sleeve (Ref. No. and shaft sleeve and make certain 5). Make sure the impeller floats that they are perfectly clean.
  • Page 150 (S/N's AFS-0001 through AFS-0236) Shertech Operating Instructions, Performance, Bronze and Cast Iron Models Specifications and Parts Manual Close-Coupled Turbine Pumps Bronze and Cast Iron Models Seal Assembly Removal and Installation (Figures 4 thru 7) Figure 4 - Impeller, Sleeve and Seal Assembly Removal Figure 5 - Seal Seat Removal Figure 6 - Seal Seat Installation Figure 7 - Seal, Collar, Sleeve and Impeller Installation...
  • Page 151 (S/N's AFS-0001 through AFS-0236) Shertech Operating Instructions, Performance, Specifications and Parts Manual Contact a Shertech Distributor Distributors can be found at www.shertech.com or www.hyproindustrial.com (The factory only sells pumps and parts to distributors.) Please provide following information: -Model number -Serial number (if any) -Part description and number as shown in parts list Optional Pedestal Base...
  • Page 152 T31 SERIES Operation 3. Jog the pump momentarily 3B Inlet and Outlet Locations (Refer CLOSE COUPLED PUMPS using the on/off buttons if so to Figure 3-1) FLEXIBLE COUPLED PUMPS equipped. The pump inlet is located on the end farthest from the motor. The discharge A.
  • Page 153 Connect the vacuum line to the dis- Priming time depends on lift, volume of restarting. If the pump will be down for charge side of the pump, either in the air in the suction line, and the size of more than a few weeks it is advisable to discharge opening or the drain tap.
  • Page 154 system periodically in the same way must exceed that in the seal chamber it is possible to removes the seals as 4B. As an additional system check, during disassembly and determine measure the temperature as it leaves instruction sheets for further cooling their condition.
  • Page 155 T31 CLOSE COUPLED PUMP NAME/DESCRIPTION PART NO. QTY. Motor Bracket Cover "O" Ring/Casing "O" Ring/Drain Plug "O" Ring/Thru Bolt Snap Ring Impeller Seal Rotating Element Seal Stationary Seat Thru Bolt Pipe Plug Key/Impeller Drive 2. Reconnect the electrical connections, referring to the colored or numbered tape used to mark the wires.
  • Page 156 5B — Disassembly of the P2 Unit. supported during this operation, no should be no smaller than .002" below 5D — Reassembly of the P2 Unit. damage will be done to the bearings. the nominal fractional seal sizes. Remove any nicks or burrs which may An exploded view of the unit, Figure 5-3 have occurred during disassembly.
  • Page 157 5E Testing and Final Adjustments 1. Check to be sure that the rotating assembly turns freely. Turn the shaft by hand. If it is tight or rough spots are encountered, it is likely that at least one of the bearings was damaged during disassembly/ assembly operations and will have to be replaced.
  • Page 158: Vacuum Leak Test Troubleshooting Guide

                              Vacuum Leak Test Troubleshooting Guide       ...
  • Page 159 BREAK VALVE NEXT VACUUM LEAK TEST? TO HEPA FILTER REPLACE CONDENSATE DOES UNIT PASS CHECK VALVE IN WRS VACUUM LEAK TEST? REPLACE VACUUM/EXHAUST DOES UNIT PASS CHECK VALVE IN WRS VACUUM LEAK TEST? CONTACT FDI: techsupport@fortdefianceind.com (865) 408-0100 REV 2, DATE: 4/7/16...
  • Page 160: Suggested Maximum Values Of Contaminants In Feed Water

      A.6  Suggested Maximum Values of Contaminants in Feed Water    Contaminant  Value  Residue on Evaporation  ≤ 10 mg/L   Silicate  ≤ 1 mg/L   Iron  ≤ 0.2 mg/L  Cadmium  ≤ 0.005 mg/L  Lead  ≤ 0.05 mg/L  Rest of heavy metals except Iron, Cadmium, Lead  ≤ 0.1 mg/L  Chloride  ≤ 0.5 mg/L   Free Chlorine  ≥ 0.2 mg/L  Phosphate  ≤ 0.5 mg/L  Conductivity  ≤ 5 µS/cm  pH (20°C, 68°F) Value  5 – 7.5  Appearance  Colorless; clean without sediment  Hardness  ≤ 10 mg/L   Total Dissolved Solids (TDS)  ≤ 50 ppm    NOTE:   1. Reference Section 2.5 Water Supply Quality. It is the responsibility of the user to  supply feedwater that meets FDA bottled water criteria.   2. Reference Section 3.4 Operational Checklists, Testing Total Dissolved Solids (TDS),  FDA‐Required Water Changeout, and Testing Free Chlorine Level. It is the ...
  • Page 161: Warranty

      A.7  LIMITED WARRANTY    Fort Defiance Industries LLC (“FDI”) warrants to the original retail purchaser that it will repair or replace components of the P2131 Automated Field  Steam Sterilizer (“P2131 Sterilizer”) manufactured by FDI (except for components not warranted under “Exclusions”) that are defective in material  or workmanship under normal use and service. FDI’s obligation under this warranty is limited to the repair or replacement, at FDI’s option, of the  applicable components.     In cases where return of the P2131 Sterilizer for repair is not practical, FDI may, at its sole discretion, authorize in writing repair of the unit by  trained technicians.  In such circumstances, FDI will furnish at the place of delivery the material or parts, and the instructions required to  successfully accomplish the repair.    This limited warranty shall only apply to defects that are reported to FDI within the applicable warranty period and which, upon assessment by FDI,  prove to be defective. This warranty extends only to the first retail purchaser of a P2131 Sterilizer and is not transferable or assignable.    This warranty shall not limit the Customer’s rights under any inspection clauses in relation to latent defects, fraud, or gross mistakes that amount  to fraud. This warranty applies notwithstanding inspection and acceptance or other clauses or terms of this contract.    APPLICABLE WARRANTY PERIOD  The applicable warranty period measured from the date of delivery to the original purchaser (whether wholesale/resale or retail) shall be one (1)  year.    OBTAINING WARRANTY SERVICE  FDI may be contacted for warranty service inquiries or issues via phone at 865‐408‐0100 or by email at techsupport@fortdefianceind.com.    EXCLUSIONS  This warranty does not cover, and FDI shall not be liable for, the following:  Defects, damage or other conditions caused, in whole or in part, by misuse, abuse, negligence, tampering, alteration, accident, damage  during transport, or failure to seek and obtain repair or replacement in a timely manner;  P2131 Sterilizers which are not installed, qualified, used, and properly cleaned and maintained as required in the P2131 “Technical  Manual”;  Components considered to be of a consumable nature (e.g., O‐rings, water test strips, filters) or components that will require  replacement per the maintenance schedule;  Parts and accessories not purchased through FDI, nor product failure or damage associated with parts not purchased through FDI;  Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such  products which are not expressly authorized in writing in advance by FDI;  Costs and expenses of routine maintenance and cleaning; and  Representations and warranties made by any person or entity other than FDI.   ...

Table of Contents